Peter Huber Kältemaschinenbau TC45E-F
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau TC45E-F

Peter Huber Kältemaschinenbau TC45E-F — Immersion coolers with Noise level 54 dB.

The TC45E-F is an air-cooled immersion cooler with a flexible 600 mm probe (13 mm diameter) and 1050 mm flexible connection, operating from -45 to 100 °C. It is distinguished by its low cooling output at deep temperatures (0.05 kW at -40 °C) and continuous refrigeration with solenoid-valve control. It is used to cool baths or vessels directly via the immersed probe, without a pump.

Noise level
54 dB

Peter Huber Kältemaschinenbau Eintauchkuehler Kuehlfallen Series — Model Comparison & Selection

Model Noise level, dBControllerPower supplyRefrigerantSafety classTemperature range, °C
TC45E-F Current 54MPC208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-45–100
CT50 Single OLÉ 59OLÉ220-240V 1~/2~ 50/60HzR-290 (A3, H220)III / FL-50–50
TC100 55ON-OFF-Regler220-240V 1~/2~ 50/60HzR-1150 (A3, H220)I / NFL-100–40
TC100E-F 55MPC110-120V 1~ 60HzR-1150 (A3, H220)I / NFL-100–40
TC100-F 58ON-OFF-Regler220-240V 1~/2~ 50/60HzR-1150 (A3, H220)I / NFL-100–40
TC45 —ON-OFF-Regler208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-45–100
TC45E —MPC208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-45–100
TC45-F 59ON-OFF-Regler208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-45–100
TC50 —ON-OFF-Regler208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-50–50
TC50E —MPC208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-50–50
TC50E-F 60MPC208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-50–50
TC50-F 60ON-OFF-Regler208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-50–50

Full specifications

Core Performance & Functional Specifications
Noise level 54 dB(A)
Pump speed adjustable no
Pump type no pump
External sensor connection Pt100
Temperature range -45...100 °C
Temperature stability mfr ±0, 5 K
Alarm message mfr optic, acoustic
Dimensions (W x D x H) mfr 190 x 295 x 360 mm
Refrigeration machine mfr air-cooled, natural refrigerant
-20 mfr 0, 18 kW (at -20 °C)
-30 mfr 0, 1 kW (at -30 °C)
-40 mfr 0, 05 kW (at -40 °C)
Dimensions, Weight & Mechanical Design
Weight 16 kg
Electrical, Power & Drive Specifications
Power supply 208-240V 1~/2~ 50/60Hz
Current consumption max. mfr 1, 4 A (max.)
Interfaces, Connections & Control
Controller MPC
Temperature display LED Display
Materials, Media & Process Compatibility
Refrigerant R-290 (A3, H220) (1st stage)
Refrigerant quantity 1st stage mfr 0, 041 kg (1st stage)
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling capacity mfr 0, 24 kW (at 0 °C)
Standards, Safety & General Data
Safety class I / NFL

Description

Construction and positioning

Performance and behaviour

Installation and operation

Integration & Operating Notes

Process Limits
Cooling capacity collapses at deep setpoints

At -40 °C only 0.05 kW is available, versus 0.24 kW at 0 °C. The heat load of the bath or vessel must be calculated at the coldest required setpoint, not at 0 °C, or the unit will fail to reach or hold temperature.

Datasheet, Specifications
Electrical
Voltage and frequency tolerances are interlocked

A ±5 % voltage deviation is only allowed if frequency stays within ±2 %, and the combination of -5 % voltage with +2 % frequency is explicitly not allowed. Sites with unstable mains must be assessed against this rule to avoid compressor damage.

Datasheet, Specifications
Mounting
Ambient temperature must stay between 5 and 40 °C

The air-cooled refrigeration circuit is rated at 20 °C room temperature; higher ambient reduces cooling capacity, and operation outside 5–40 °C can trip the refrigeration circuit. The installation room must be kept within this band.

Datasheet, Specifications
Safety
Flammable refrigerant requires ventilation and ignition control

R-290 is classified A3 (H220), so the installation location must meet ventilation and ignition-source requirements for flammable refrigerants. The refrigerant charge is 0.041 kg in the first stage.

Datasheet, Specifications
Control & I/O
Continuous refrigeration with solenoid-valve control

The compressor runs continuously and temperature is regulated by a solenoid valve in the cooling circuit. Wiring of the valve and any external Pt100 sensor must be correct for stable control; the controller is MPC with LED display and optic/acoustic alarm.

Datasheet, Specifications
Electrical
Country-specific plug must be ordered

Single- and two-phase devices are supplied with a power cable and country-specific plug only if specified at ordering. The supply must be fused between 10 A and 16 A; maximum current draw is 1.4 A.

Datasheet, Installation

Selection & Integration Q&A

When should I choose the TC45E-F instead of a larger sibling like the TC100E-F?

Choose the TC45E-F for small heat loads where the immersed probe can reach the bath directly and no pump is needed. Its capacity is only 0.05 kW at -40 °C; for higher loads or pumped circulation, a larger sibling is required.

What electrical supply and protection does the TC45E-F require at the site?

A 208–240 V, 1~/2~, 50/60 Hz supply fused between 10 A and 16 A. Maximum current is 1.4 A. The country-specific plug must be specified at ordering; voltage and frequency tolerances are interlocked.

What ambient and installation limits apply to the TC45E-F?

Ambient temperature must remain between 5 and 40 °C; capacity is rated at 20 °C and falls at higher ambient. R-290 refrigerant requires ventilation and ignition-source control. EMC class A group 1 means industrial environments only.

What is included in base delivery and what must be ordered separately?

The unit includes the 600 mm probe, 1050 mm flexible connection, and power cable with country-specific plug only if specified at ordering. No pump is fitted. External Pt100 sensor is an interface option, not a standard accessory.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
TC45E-F Refrigerated/Heating Circulator Technical Data
Datasheet ·63,2 КБ ·2 pp.
View PDF