Peter Huber Kältemaschinenbau TC50-F
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau TC50-F

Peter Huber Kältemaschinenbau TC50-F — Immersion coolers with Noise level 60 dB.

The TC50-F is an air-cooled immersion cooler with a flexible 900 mm condensing probe (13 mm OD) and 1150 mm connection line, covering -50 to 50 °C with an ON-OFF controller and no pump. It is a single-stage unit rated 0.2 kW at -30 °C, and serves low-load cold-trap and bath-cooling duties.

Noise level
60 dB

Peter Huber Kältemaschinenbau Eintauchkuehler Kuehlfallen Series — Model Comparison & Selection

Model Noise level, dBControllerPower supplyRefrigerantSafety classTemperature range, °C
TC50-F Current 60ON-OFF-Regler208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-50–50
CT50 Single OLÉ 59OLÉ220-240V 1~/2~ 50/60HzR-290 (A3, H220)III / FL-50–50
TC100 55ON-OFF-Regler220-240V 1~/2~ 50/60HzR-1150 (A3, H220)I / NFL-100–40
TC100E-F 55MPC110-120V 1~ 60HzR-1150 (A3, H220)I / NFL-100–40
TC100-F 58ON-OFF-Regler220-240V 1~/2~ 50/60HzR-1150 (A3, H220)I / NFL-100–40
TC45 —ON-OFF-Regler208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-45–100
TC45E —MPC208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-45–100
TC45E-F 54MPC208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-45–100
TC45-F 59ON-OFF-Regler208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-45–100
TC50 —ON-OFF-Regler208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-50–50
TC50E —MPC208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-50–50
TC50E-F 60MPC208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-50–50

Full specifications

Core Performance & Functional Specifications
Noise level 60 dB(A)
Pump speed adjustable no
Pump type no pump
Temperature range -50...50 °C
Dimensions (W x D x H) mfr 260 x 330 x 415 mm
Refrigeration machine mfr air-cooled, natural refrigerant
Dimensions, Weight & Mechanical Design
Weight 25 kg
Electrical, Power & Drive Specifications
Current consumption 3 A (max.)
Power supply 208-240V 1~/2~ 50/60Hz
Interfaces, Connections & Control
Controller ON-OFF-controller
Materials, Media & Process Compatibility
Refrigerant R-290 (A3, H220) (1st stage)
Refrigerant quantity 1st stage mfr 0, 06 kg (1st stage)
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling capacity mfr 0, 2 kW (at -30 °C)
Standards, Safety & General Data
Safety class I / NFL

Description

Construction and series positioning

The TC50-F is a single-stage, air-cooled refrigeration unit using R-290 (A3, H220) as the first-stage refrigerant (0.06 kg charge). Unlike the TC100/TC100E variants, which are larger and quieter (55 dB(A)), the TC50-F is rated at 60 dB(A) ±4 dB(A) and weighs 25 kg in a 260 × 330 × 415 mm footprint. It has no pump and no adjustable pump speed; cooling is delivered directly through a flexible 900 mm probe of 13 mm nominal diameter, connected by a 1150 mm flexible line. The controller is an ON-OFF type, so the unit cycles the refrigeration circuit rather than modulating capacity.

Thermal behavior and load limits

Cooling capacity is modest and strongly temperature-dependent: 0.3 kW at 0 °C, 0.26 kW at -20 °C and 0.2 kW at -30 °C, all rated at 20 °C room temperature. Capacity drops as ambient temperature rises, so the unit is suited to low-load immersion cooling and cold-trap duty rather than bulk heat removal. The refrigeration unit runs continuously; on/off switching is not a temperature-control strategy. The working range is -50 to 50 °C, and the unit is rated for ambient temperatures of 5 to 40 °C only.

Installation, electrical and compliance

Power supply is 208–240 V, 1~/2~, 50/60 Hz with a maximum current of 3 A and a fuse between 10 A and 16 A. Voltage tolerance is ±5 % only with a simultaneous frequency tolerance of ±2 %, and the combination of -5 % voltage with +2 % frequency is not permitted. Single/two-phase units are delivered with a power cable and a country-specific plug that must be specified at ordering. The equipment is EMC class A, group 1 (EN 55011), intended for industrial environments, and complies with US-SNAP for industrial process refrigeration. Because the refrigerant is flammable (A3), ignition sources and ventilation must be considered at the installation site, and space requirements around the air-cooled condenser must be respected.

Integration & Operating Notes

Electrical
Supply tolerance pairing is restricted

The ±5 % voltage tolerance is only valid together with a ±2 % frequency tolerance, and the specific combination of -5 % voltage with +2 % frequency is not allowed. Verify the site supply against both limits before connection, because an out-of-band combination voids the rating.

TC50-F Datasheet, Specifications
Electrical
Country-specific plug must be ordered

Single/two-phase devices are delivered with a power cable and a country-specific plug that must be specified when ordering. Confirm the connector type at the order stage to avoid a mismatch with the site socket.

TC50-F Datasheet, Specifications
Process Limits
Cooling capacity is low and ambient-sensitive

Rated capacity is 0.3 kW at 0 °C, 0.26 kW at -20 °C and 0.2 kW at -30 °C at 20 °C room temperature, and it may drop as ambient temperature rises. Size the duty for low-load immersion cooling and cold-trap service, not bulk heat removal.

TC50-F Datasheet, Specifications
Cooling & Media
Flammable A3 refrigerant requires site precautions

The unit uses R-290 (propane, A3, H220) with a 0.06 kg first-stage charge. Installation must account for ignition sources and ventilation in the vicinity, and the air-cooled condenser needs the space requirements stated in the operating conditions.

TC50-F Datasheet, Specifications
Mounting
Ambient room band is 5 to 40 °C

The unit is only rated for ambient temperatures from 5 °C to 40 °C. The installation room must stay within this band, otherwise performance and reliability are not guaranteed; plan the condenser location accordingly.

TC50-F Datasheet, Specifications
Maintenance
Refrigeration unit runs continuously

The refrigeration unit works continuously and should not be switched on/off as a temperature-control strategy. Plan energy use, noise exposure (60 dB(A) ±4 dB(A)) and compressor life around continuous operation.

TC50-F Datasheet, Specifications

Selection & Integration Q&A

When should I choose the TC50-F rather than a TC100/TC100E variant in the same series?

Pick the TC50-F for low-load immersion cooling or cold-trap duty where 0.2 kW at -30 °C is sufficient and a 900 mm probe with 1150 mm line reaches the vessel. The TC100/TC100E are larger, quieter (55 dB(A)) alternatives when more capacity or lower noise is needed.

What supply and protection must be provided for the TC50-F, and what about the plug?

Provide 208–240 V, 1~/2~, 50/60 Hz with max. 3 A and a 10–16 A fuse. Voltage tolerance ±5 % only with ±2 % frequency; -5 % voltage with +2 % frequency is not allowed. The country-specific plug must be specified at ordering.

Where can the TC50-F be installed, and what duty-cycle behavior should I expect?

Install only where ambient stays between 5 and 40 °C, with space around the air-cooled condenser and attention to the flammable R-290 (A3) refrigerant. The refrigeration unit runs continuously; do not use on/off switching as a control strategy. Noise is 60 dB(A) ±4 dB(A).

What is included in base delivery and what must be ordered or planned separately?

Base delivery includes the unit with a 900 mm probe (13 mm OD) and 1150 mm flexible connection line, plus a power cable with country-specific plug only if specified at ordering. No pump is fitted; the ON-OFF controller is integral. Confirm plug type and any site-specific accessories at order.