Peter Huber Kältemaschinenbau TC45E-F
Peter Huber Kältemaschinenbau TC45E-F — Immersion coolers with Noise level 54 dB.
The TC45E-F is an air-cooled immersion cooler with a flexible 600 mm probe (13 mm diameter) and 1050 mm flexible connection, operating from -45 to 100 °C. It is distinguished by its low cooling output at deep temperatures (0.05 kW at -40 °C) and continuous refrigeration with solenoid-valve control. It is used to cool baths or vessels directly via the immersed probe, without a pump.
Peter Huber Kältemaschinenbau Eintauchkuehler Kuehlfallen Series — Model Comparison & Selection
| Model | Noise level, dB | Controller | Power supply | Refrigerant | Safety class | Temperature range, °C |
|---|---|---|---|---|---|---|
| TC45E-F Current | 54 | MPC | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -45–100 |
| CT50 Single OLÉ | 59 | OLÉ | 220-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | III / FL | -50–50 |
| TC100 | 55 | ON-OFF-Regler | 220-240V 1~/2~ 50/60Hz | R-1150 (A3, H220) | I / NFL | -100–40 |
| TC100E-F | 55 | MPC | 110-120V 1~ 60Hz | R-1150 (A3, H220) | I / NFL | -100–40 |
| TC100-F | 58 | ON-OFF-Regler | 220-240V 1~/2~ 50/60Hz | R-1150 (A3, H220) | I / NFL | -100–40 |
| TC45 | — | ON-OFF-Regler | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -45–100 |
| TC45E | — | MPC | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -45–100 |
| TC45-F | 59 | ON-OFF-Regler | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -45–100 |
| TC50 | — | ON-OFF-Regler | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -50–50 |
| TC50E | — | MPC | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -50–50 |
| TC50E-F | 60 | MPC | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -50–50 |
| TC50-F | 60 | ON-OFF-Regler | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -50–50 |
Full specifications
| Noise level | 54 dB(A) |
| Pump speed adjustable | no |
| Pump type | no pump |
| External sensor connection | Pt100 |
| Temperature range | -45...100 °C |
| Temperature stability mfr | ±0, 5 K |
| Alarm message mfr | optic, acoustic |
| Dimensions (W x D x H) mfr | 190 x 295 x 360 mm |
| Refrigeration machine mfr | air-cooled, natural refrigerant |
| -20 mfr | 0, 18 kW (at -20 °C) |
| -30 mfr | 0, 1 kW (at -30 °C) |
| -40 mfr | 0, 05 kW (at -40 °C) |
| Weight | 16 kg |
| Power supply | 208-240V 1~/2~ 50/60Hz |
| Current consumption max. mfr | 1, 4 A (max.) |
| Controller | MPC |
| Temperature display | LED Display |
| Refrigerant | R-290 (A3, H220) (1st stage) |
| Refrigerant quantity 1st stage mfr | 0, 041 kg (1st stage) |
| Ambient temperature limit | 5...40 °C |
| Cooling capacity mfr | 0, 24 kW (at 0 °C) |
| Safety class | I / NFL |
Description
Construction and positioning
- The TC45E-F is a direct-expansion immersion cooler: a flexible probe (600 mm long, 13 mm diameter) connected by a 1050 mm flexible line to an air-cooled refrigeration unit using R-290 (A3, H220) as the first-stage refrigerant.
- Unlike sibling models such as the TC100 or TC100E-F, which are larger refrigerated/heating circulators with pumps, this unit has no pump and no pump-speed adjustment; cooling is delivered solely through the immersed probe.
- The refrigeration machine is air-cooled and runs continuously; temperature is regulated by a solenoid valve in the cooling circuit, not by cycling the compressor.
Performance and behaviour
- The working range is -45 to 100 °C, but cooling capacity falls steeply as the setpoint drops: 0.24 kW at 0 °C, 0.18 kW at -20 °C, 0.1 kW at -30 °C, and only 0.05 kW at -40 °C. Low-temperature duty therefore requires careful heat-load sizing.
- Rated capacities are valid at 20 °C room temperature; higher ambient temperatures reduce available cooling output.
- Temperature stability is ±0.5 K, with an LED display and MPC controller. An external Pt100 sensor can be connected.
Installation and operation
- The unit requires a 208–240 V, 1~/2~, 50/60 Hz supply with a fuse between 10 A and 16 A; maximum current is 1.4 A. Single- and two-phase devices are delivered with a power cable and country-specific plug, which must be specified at ordering.
- Voltage and frequency tolerances are linked: ±5 % voltage is only permitted with ±2 % frequency, and the combination of -5 % voltage with +2 % frequency is not allowed.
- Ambient temperature must remain between 5 and 40 °C. The refrigerant R-290 is flammable (A3), so ventilation and ignition-source requirements apply. The unit is EMC class A, group 1, intended for industrial environments.
Integration & Operating Notes
At -40 °C only 0.05 kW is available, versus 0.24 kW at 0 °C. The heat load of the bath or vessel must be calculated at the coldest required setpoint, not at 0 °C, or the unit will fail to reach or hold temperature.
A ±5 % voltage deviation is only allowed if frequency stays within ±2 %, and the combination of -5 % voltage with +2 % frequency is explicitly not allowed. Sites with unstable mains must be assessed against this rule to avoid compressor damage.
The air-cooled refrigeration circuit is rated at 20 °C room temperature; higher ambient reduces cooling capacity, and operation outside 5–40 °C can trip the refrigeration circuit. The installation room must be kept within this band.
R-290 is classified A3 (H220), so the installation location must meet ventilation and ignition-source requirements for flammable refrigerants. The refrigerant charge is 0.041 kg in the first stage.
The compressor runs continuously and temperature is regulated by a solenoid valve in the cooling circuit. Wiring of the valve and any external Pt100 sensor must be correct for stable control; the controller is MPC with LED display and optic/acoustic alarm.
Single- and two-phase devices are supplied with a power cable and country-specific plug only if specified at ordering. The supply must be fused between 10 A and 16 A; maximum current draw is 1.4 A.
Selection & Integration Q&A
Choose the TC45E-F for small heat loads where the immersed probe can reach the bath directly and no pump is needed. Its capacity is only 0.05 kW at -40 °C; for higher loads or pumped circulation, a larger sibling is required.
A 208–240 V, 1~/2~, 50/60 Hz supply fused between 10 A and 16 A. Maximum current is 1.4 A. The country-specific plug must be specified at ordering; voltage and frequency tolerances are interlocked.
Ambient temperature must remain between 5 and 40 °C; capacity is rated at 20 °C and falls at higher ambient. R-290 refrigerant requires ventilation and ignition-source control. EMC class A group 1 means industrial environments only.
The unit includes the 600 mm probe, 1050 mm flexible connection, and power cable with country-specific plug only if specified at ordering. No pump is fitted. External Pt100 sensor is an interface option, not a standard accessory.