Peter Huber Kältemaschinenbau TC45E
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau TC45E

Peter Huber Kältemaschinenbau TC45E — Immersion coolers.

The TC45E is an air-cooled immersion cooler with a natural R-290 refrigerant circuit, rated for -45 to 100 °C and controlled by an MPC unit with 0.1 K display resolution. It delivers 0.24 kW at 0 °C but only 0.05 kW at -40 °C, so it suits low-load bath cooling and cold-trap duty rather than high-heat processes. No pump is fitted; circulation must come from the bath or an external loop.

Peter Huber Kältemaschinenbau Eintauchkuehler Kuehlfallen Series — Model Comparison & Selection

Model Noise level, dBControllerPower supplyRefrigerantSafety classTemperature range, °C
TC45E Current —MPC208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-45–100
CT50 Single OLÉ 59OLÉ220-240V 1~/2~ 50/60HzR-290 (A3, H220)III / FL-50–50
TC100 55ON-OFF-Regler220-240V 1~/2~ 50/60HzR-1150 (A3, H220)I / NFL-100–40
TC100E-F 55MPC110-120V 1~ 60HzR-1150 (A3, H220)I / NFL-100–40
TC100-F 58ON-OFF-Regler220-240V 1~/2~ 50/60HzR-1150 (A3, H220)I / NFL-100–40
TC45 —ON-OFF-Regler208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-45–100
TC45E-F 54MPC208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-45–100
TC45-F 59ON-OFF-Regler208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-45–100
TC50 —ON-OFF-Regler208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-50–50
TC50E —MPC208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-50–50
TC50E-F 60MPC208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-50–50
TC50-F 60ON-OFF-Regler208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-50–50

Full specifications

Core Performance & Functional Specifications
Pump speed adjustable no
Pump type no pump
External sensor connection Pt100
Temperature range -45...100 °C
Temperature stability 0.5 K
Alarm message mfr optic, acoustic
Dimensions (W x D x H) mfr 190 x 295 x 360 mm
Refrigeration machine mfr air-cooled, natural refrigerant
Dimensions, Weight & Mechanical Design
Weight 16 kg
Electrical, Power & Drive Specifications
Current consumption 1.4 A (max.)
Power supply 208-240V 1~/2~ 50/60Hz
Interfaces, Connections & Control
Controller MPC
Display resolution 0.1 K
Temperature display LED Display
Materials, Media & Process Compatibility
Refrigerant R-290 (A3, H220) (1st stage)
Refrigerant quantity 0.041 kg (1st stage)
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling capacity 0.24 kW (0 °C)
-20 °C: 0.18 kW · -30 °C: 0.1 kW · -40 °C: 0.05 kW
Standards, Safety & General Data
Safety class I / NFL

Description

Construction and series positioning

Behavior under load

Installation and operation

Integration & Operating Notes

Electrical
Verify site supply and 10 A protection before connection

The unit draws up to 1.4 A on a 208-240 V, 1~ or 2~, 50/60 Hz supply and the datasheet calls for a minimum 10 A fuse. Voltage and frequency tolerances are coupled: ±5 % voltage with ±2 % frequency is acceptable, but opposite-sign deviations are explicitly not allowed, so supply quality must be checked before commissioning.

TC45E Technical Data Sheet, Technical Specifications
Mounting
Keep ambient between 5 and 40 °C for rated capacity

The air-cooled refrigeration circuit rejects heat to the room, and capacity is rated at 20 °C room temperature and drops as ambient rises. Installing the unit in a cold room below 5 °C or a hot enclosure above 40 °C is outside the permitted ambient band, and space requirements around the unit must be respected for airflow and service access.

TC45E Technical Data Sheet, Technical Specifications
Process Limits
Size the load against capacity at the actual setpoint

Cooling capacity falls from 0.24 kW at 0 °C to 0.05 kW at -40 °C, so a load acceptable at near-zero temperatures may be unreachable near -45 °C. The immersion probe must be matched to the bath volume and heat ingress, and the 0.5 K stability figure applies only when the load stays within the available capacity at that operating point.

TC45E Technical Data Sheet, Technical Specifications
Control & I/O
External Pt100 sensor is included for point-of-use control

The MPC controller reads an external Pt100 sensor connected by cable and plug, and a Pt100 sensor is supplied with the unit, so external temperature control is possible without extra accessories. The refrigeration unit runs continuously and the controller actuates a solenoid valve in the cooling circuit, which defines the control behavior and the wear pattern of the valve.

TC45E Technical Data Sheet, Technical Specifications
Electrical
Specify the country-specific plug at ordering

Single- and two-phase devices are delivered with a power cable and a country-specific plug that must be stated when ordering; three-phase devices below 63 A come with cable but without plug, so the customer must provide and fit one. Omitting the plug specification delays commissioning because the unit cannot be connected on arrival.

TC45E Technical Data Sheet, page 2
Process Limits
R-290 charge imposes flammable-refrigerant site rules

The first stage uses 0.041 kg of R-290 (A3, H220), a flammable natural refrigerant. Installation must account for room ventilation and ignition-source requirements applicable to A3 refrigerants; the unit is not certified for use in explosive atmospheres and must not be placed where refrigerant can accumulate in a confined space.

TC45E Technical Data Sheet, Technical Specifications

Selection & Integration Q&A

When should I choose the TC45E rather than a TC100/TC100E or CT50 Single OLÉ from the same series?

Pick the TC45E for compact, low-load bath or cold-trap duty down to -45 °C where no pump is needed. The TC100/TC100E/TC100E-F and CT50 Single OLÉ are larger siblings; choose them when higher cooling capacity or a different noise/configuration profile is required, since the TC45E is the 16 kg, 190 x 295 x 360 mm unit with 0.24 kW at 0 °C.

What supply, protection and plug does the TC45E need at the installation site?

208-240 V, 1~ or 2~, 50/60 Hz, max 1.4 A, with a minimum 10 A fuse. Voltage and frequency tolerances are coupled (±5 % / ±2 %; opposite-sign deviations not allowed). Single- and two-phase units ship with cable and a country-specific plug that must be specified at ordering.

Where can the TC45E be installed, and what ambient and EMC limits apply?

Ambient must stay between 5 and 40 °C; capacity is rated at 20 °C room temperature and falls as ambient rises, so avoid cold rooms and hot enclosures. The unit is EMC class A, group 1 (EN 55011), intended for industrial environments, and space requirements around the air-cooled unit must be respected for heat rejection and service access.

What is included in base delivery and what must be ordered or provided separately?

A Pt100 external sensor with cable and plug is included, enabling external temperature control out of the box. The country-specific power plug must be specified at ordering for single- and two-phase units; three-phase units below 63 A come with cable but no plug, which the customer must provide and fit. No pump is fitted, so bath circulation must come from elsewhere.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
TC45E Immersion Circulator — Technical Data Sheet
Datasheet ·91,3 КБ ·2 pp.
View PDF