Peter Huber Kältemaschinenbau TC50-F
Peter Huber Kältemaschinenbau TC50-F — Immersion coolers with Noise level 60 dB.
The TC50-F is an air-cooled immersion cooler with a flexible 900 mm condensing probe (13 mm OD) and 1150 mm connection line, covering -50 to 50 °C with an ON-OFF controller and no pump. It is a single-stage unit rated 0.2 kW at -30 °C, and serves low-load cold-trap and bath-cooling duties.
Peter Huber Kältemaschinenbau Eintauchkuehler Kuehlfallen Series — Model Comparison & Selection
| Model | Noise level, dB | Controller | Power supply | Refrigerant | Safety class | Temperature range, °C |
|---|---|---|---|---|---|---|
| TC50-F Current | 60 | ON-OFF-Regler | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -50–50 |
| CT50 Single OLÉ | 59 | OLÉ | 220-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | III / FL | -50–50 |
| TC100 | 55 | ON-OFF-Regler | 220-240V 1~/2~ 50/60Hz | R-1150 (A3, H220) | I / NFL | -100–40 |
| TC100E-F | 55 | MPC | 110-120V 1~ 60Hz | R-1150 (A3, H220) | I / NFL | -100–40 |
| TC100-F | 58 | ON-OFF-Regler | 220-240V 1~/2~ 50/60Hz | R-1150 (A3, H220) | I / NFL | -100–40 |
| TC45 | — | ON-OFF-Regler | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -45–100 |
| TC45E | — | MPC | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -45–100 |
| TC45E-F | 54 | MPC | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -45–100 |
| TC45-F | 59 | ON-OFF-Regler | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -45–100 |
| TC50 | — | ON-OFF-Regler | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -50–50 |
| TC50E | — | MPC | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -50–50 |
| TC50E-F | 60 | MPC | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -50–50 |
Full specifications
| Noise level | 60 dB(A) |
| Pump speed adjustable | no |
| Pump type | no pump |
| Temperature range | -50...50 °C |
| Dimensions (W x D x H) mfr | 260 x 330 x 415 mm |
| Refrigeration machine mfr | air-cooled, natural refrigerant |
| Weight | 25 kg |
| Current consumption | 3 A (max.) |
| Power supply | 208-240V 1~/2~ 50/60Hz |
| Controller | ON-OFF-controller |
| Refrigerant | R-290 (A3, H220) (1st stage) |
| Refrigerant quantity 1st stage mfr | 0, 06 kg (1st stage) |
| Ambient temperature limit | 5...40 °C |
| Cooling capacity mfr | 0, 2 kW (at -30 °C) |
| Safety class | I / NFL |
Description
Construction and series positioning
The TC50-F is a single-stage, air-cooled refrigeration unit using R-290 (A3, H220) as the first-stage refrigerant (0.06 kg charge). Unlike the TC100/TC100E variants, which are larger and quieter (55 dB(A)), the TC50-F is rated at 60 dB(A) ±4 dB(A) and weighs 25 kg in a 260 × 330 × 415 mm footprint. It has no pump and no adjustable pump speed; cooling is delivered directly through a flexible 900 mm probe of 13 mm nominal diameter, connected by a 1150 mm flexible line. The controller is an ON-OFF type, so the unit cycles the refrigeration circuit rather than modulating capacity.
Thermal behavior and load limits
Cooling capacity is modest and strongly temperature-dependent: 0.3 kW at 0 °C, 0.26 kW at -20 °C and 0.2 kW at -30 °C, all rated at 20 °C room temperature. Capacity drops as ambient temperature rises, so the unit is suited to low-load immersion cooling and cold-trap duty rather than bulk heat removal. The refrigeration unit runs continuously; on/off switching is not a temperature-control strategy. The working range is -50 to 50 °C, and the unit is rated for ambient temperatures of 5 to 40 °C only.
Installation, electrical and compliance
Power supply is 208–240 V, 1~/2~, 50/60 Hz with a maximum current of 3 A and a fuse between 10 A and 16 A. Voltage tolerance is ±5 % only with a simultaneous frequency tolerance of ±2 %, and the combination of -5 % voltage with +2 % frequency is not permitted. Single/two-phase units are delivered with a power cable and a country-specific plug that must be specified at ordering. The equipment is EMC class A, group 1 (EN 55011), intended for industrial environments, and complies with US-SNAP for industrial process refrigeration. Because the refrigerant is flammable (A3), ignition sources and ventilation must be considered at the installation site, and space requirements around the air-cooled condenser must be respected.
Integration & Operating Notes
The ±5 % voltage tolerance is only valid together with a ±2 % frequency tolerance, and the specific combination of -5 % voltage with +2 % frequency is not allowed. Verify the site supply against both limits before connection, because an out-of-band combination voids the rating.
Single/two-phase devices are delivered with a power cable and a country-specific plug that must be specified when ordering. Confirm the connector type at the order stage to avoid a mismatch with the site socket.
Rated capacity is 0.3 kW at 0 °C, 0.26 kW at -20 °C and 0.2 kW at -30 °C at 20 °C room temperature, and it may drop as ambient temperature rises. Size the duty for low-load immersion cooling and cold-trap service, not bulk heat removal.
The unit uses R-290 (propane, A3, H220) with a 0.06 kg first-stage charge. Installation must account for ignition sources and ventilation in the vicinity, and the air-cooled condenser needs the space requirements stated in the operating conditions.
The unit is only rated for ambient temperatures from 5 °C to 40 °C. The installation room must stay within this band, otherwise performance and reliability are not guaranteed; plan the condenser location accordingly.
The refrigeration unit works continuously and should not be switched on/off as a temperature-control strategy. Plan energy use, noise exposure (60 dB(A) ±4 dB(A)) and compressor life around continuous operation.
Selection & Integration Q&A
Pick the TC50-F for low-load immersion cooling or cold-trap duty where 0.2 kW at -30 °C is sufficient and a 900 mm probe with 1150 mm line reaches the vessel. The TC100/TC100E are larger, quieter (55 dB(A)) alternatives when more capacity or lower noise is needed.
Provide 208–240 V, 1~/2~, 50/60 Hz with max. 3 A and a 10–16 A fuse. Voltage tolerance ±5 % only with ±2 % frequency; -5 % voltage with +2 % frequency is not allowed. The country-specific plug must be specified at ordering.
Install only where ambient stays between 5 and 40 °C, with space around the air-cooled condenser and attention to the flammable R-290 (A3) refrigerant. The refrigeration unit runs continuously; do not use on/off switching as a control strategy. Noise is 60 dB(A) ±4 dB(A).
Base delivery includes the unit with a 900 mm probe (13 mm OD) and 1150 mm flexible connection line, plus a power cable with country-specific plug only if specified at ordering. No pump is fitted; the ON-OFF controller is integral. Confirm plug type and any site-specific accessories at order.