Peter Huber Kältemaschinenbau Unimotive 10w-XT
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unimotive 10w-XT

Peter Huber Kältemaschinenbau Unimotive 10w-XT — Temperature control units.

The Unimotive 10w-XT is a water-cooled temperature control unit with a magnetically coupled stainless-steel pump and Pilot ONE controller, covering -45 to 150 °C for water-ethylene glycol mixtures. Its 12 kW heating and 14 kW cooling at 20 °C distinguish it within the Unimotive series, while cooling capacity falls to 0.8 kW at -40 °C. It serves automotive test benches and process circuits requiring sub-zero temperature control.

Peter Huber Kältemaschinenbau Unimotive Series — Model Comparison & Selection

Model Display resolutionHeating capacity, WPower supplyPump connectionRefrigerantSafety class
Unimotive 10w-XT Current 0, 0112000400V 3~ 50HzM38x1, 5 AG maleR-452A (A1, H280)I / NFL
Unimotive 10w 0.0112000400V 3~ 50HzM38x1,5 AG maleR-452A (A1, H280)III / FL
Unimotive 20w 0, 0112000400V 3~ 50HzM38x1, 5 AG maleR-452A (A1, H280)III / FL
Unimotive 20w-XT 0, 0112000400V 3~ 50HzM38x1, 5 AG maleR-452A (A1, H280)I / NFL
Unimotive 26w 0, 01 K24000400V 3~ 50HzM38x1, 5 AG maleR-452A (A1, H280)III / FL
Unimotive 26w-XT 0, 0124000400V 3~ 50HzM38x1, 5 AG maleR-452A (A1, H280)I / NFL
Unimotive 27w 0, 0124000400V 3~ 50HzM38x1, 5 AG maleR-452A (A1, H280)III / FL
Unimotive GL 07w 0, 0112000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)III / FL
Unimotive GL 07w-XT 0, 0112000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)I / NFL
Unimotive GL 10w 0, 0124000460V 3~ 60HzM38x1, 5 AG maleR-744 (A1, H280)III / FL
Unimotive GL 10w-XT 0, 0124000460V 3~ 60HzM38x1, 5 AG maleR-744 (A1, H280)I / NFL
Unimotive GL 20w 0, 0124000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)III / FL
Unimotive GL 20w-XT 0, 0124000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)I / NFL
Unimotive GL 30w 0, 0124000460V 3~ 60HzM38x1, 5 AG maleR-744 (A1, H280)III / FL
Unimotive GL 50w 0, 0124000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)III / FL
Unimotive GL 50w-XT 0, 0124000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)I / NFL
Unimotive GL 60w 0, 0148000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)III / FL
Unimotive GL 60w-XT 0, 0148000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)I / NFL

Full specifications

Core Performance & Functional Specifications
Control output POKO ONE
Heating capacity 12 kW
Pump connection M38x1, 5 AG male
Pump speed adjustable no
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -45...150 °C
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Refrigerant R-452A (A1, H280) (1st stage)
Maximum viscosity 200 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G3/4 male
Cooling water consumption 620 l/h (at 15°C, flow 0°C)
Cooling water differential pressure 1 bar (min.)
Cooling Water Pressure 6 bar (max.)
Protection rating IP20
Standards, Safety & General Data
Safety class I / NFL

Description

Construction and series positioning

Behavior under load and media

Installation, connection and control

Integration & Operating Notes

Process Limits
Maximum heat transfer fluid viscosity 200 mm²/s

The maximum permissible kinematic viscosity of the heat transfer fluid is 200 mm²/s. Exceeding this can cause pump overload or poor circulation, especially at low temperatures.

Datasheet, Technical Specifications, page 2
Process Limits
Cooling capacity drops to 0.8 kW at -40 °C

Cooling capacity falls significantly at low temperatures, from 14 kW at 20 °C to 0.8 kW at -40 °C. Verify that the process heat load at the lowest operating temperature does not exceed this reduced capacity.

Datasheet, Technical Specifications, page 1
Safety
Do not install or start up inside an ATEX zone

The unit must not be installed or started inside an ATEX zone, so hazardous-area installations are excluded. Additionally, observe the refrigerant charge (2.9 kg R-452A, 6.2 t CO2 equivalent) and room size to avoid dangerous refrigerant concentration.

Operation Manual, Installation, page 13 and 15
Electrical
Delivered without plug for three-phase devices below 63 A

The unit requires a 400 V three-phase 50 Hz supply and is delivered with cable but without a plug for three-phase devices below 63 A. The electrician must provide a suitable plug/connection on site. Replacement power cables are length-limited to 4 m for 3-phase connections.

Datasheet, Technical Specifications, page 2; Operation Manual, Electrical Connection, page 31

Selection & Integration Q&A

When should I choose the Unimotive 10w-XT over the Unimotive 10w or 20w-XT?

Choose the 10w-XT when you need -45 °C capability with water-ethylene glycol and moderate heating/cooling capacity (12 kW heating, 14 kW cooling at 20 °C). The 10w lacks the extended low-temperature range; the 20w-XT offers higher capacity but may be oversized for smaller loads.

What are the cooling water supply requirements and adjacent plumbing for this unit?

Provide cooling water at 5–20 °C with a differential pressure of 1–4 bar and maximum supply pressure of 6 bar. Use G3/4 male connections, install a backflow preventer on the building side, and ensure only drinking water or filtered river/sea water (no sea water) is used.

What are the ambient and duty-cycle limits for installation?

Operate within an ambient temperature of 5–40 °C. Do not install in an ATEX zone. Ensure minimum wall clearances per the manual, install a drip tray, and shut off cooling water during downtimes. The unit is EN 55011 Class A Group 1, suitable for industrial EMC environments.

What maintenance and accessories are required that are not included in the base delivery?

The base delivery excludes a plug for three-phase devices below 63 A; the electrician must provide a suitable plug/connection. A drip tray and backflow preventer are site responsibilities. Regular maintenance includes checking cooling water quality and refrigerant charge; no specific service intervals are listed in the provided digests.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unimotive 10w-XT Immersion Circulator
Datasheet ·84,5 КБ ·2 pp.
View PDF
PDF
Unimotive Operation Manual
Manual ·2,8 МБ ·82 pp.
View PDF