Peter Huber Kältemaschinenbau Unimotive GL 10w-XT
Peter Huber Kältemaschinenbau Unimotive GL 10w-XT — Temperature control units.
The Unimotive GL 10w-XT is a water-cooled, magnetically coupled stainless-steel pump temperature control unit for water-ethylene glycol mixtures down to -45 °C and up to 150 °C, rated 24 kW heating on a 460 V 3~ 60 Hz supply. It is the extended-temperature (XT) variant of the 10w platform, distinguished from the 20w siblings by its smaller capacity class. It serves automotive test benches and process loops requiring sub-zero glycol temperature control with Pilot ONE regulation.
Peter Huber Kältemaschinenbau Unimotive Series — Model Comparison & Selection
| Model | Display resolution | Heating capacity, W | Power supply | Pump connection | Refrigerant | Safety class |
|---|---|---|---|---|---|---|
| Unimotive GL 10w-XT Current | 0, 01 | 24000 | 460V 3~ 60Hz | M38x1, 5 AG male | R-744 (A1, H280) | I / NFL |
| Unimotive 10w | 0.01 | 12000 | 400V 3~ 50Hz | M38x1,5 AG male | R-452A (A1, H280) | III / FL |
| Unimotive 10w-XT | 0, 01 | 12000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | I / NFL |
| Unimotive 20w | 0, 01 | 12000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | III / FL |
| Unimotive 20w-XT | 0, 01 | 12000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | I / NFL |
| Unimotive 26w | 0, 01 K | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | III / FL |
| Unimotive 26w-XT | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | I / NFL |
| Unimotive 27w | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | III / FL |
| Unimotive GL 07w | 0, 01 | 12000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 07w-XT | 0, 01 | 12000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | I / NFL |
| Unimotive GL 10w | 0, 01 | 24000 | 460V 3~ 60Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 20w | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 20w-XT | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | I / NFL |
| Unimotive GL 30w | 0, 01 | 24000 | 460V 3~ 60Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 50w | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 50w-XT | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | I / NFL |
| Unimotive GL 60w | 0, 01 | 48000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 60w-XT | 0, 01 | 48000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | I / NFL |
Full specifications
| Control output | POKO ONE |
| Heating capacity | 24 kW |
| Pump connection | M38x1, 5 AG male |
| Pump speed adjustable | no |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -45...150 °C |
| Weight | 600 kg |
| Current consumption | 47 A (max) |
| Power supply | 460V 3~ 60Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Refrigerant | R-744 (A1, H280) (1st stage) |
| Maximum viscosity | 200 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G3/4 male |
| Cooling water consumption | 2100 l/h (at 15°C, flow 0°C) |
| Cooling water differential pressure | 1 bar (min) |
| Cooling Water Pressure | 6 bar (max) |
| Protection rating | IP20 |
| Safety class | I / NFL |
Description
Construction and series positioning
- Immersion circulator/thermostat architecture with a magnetically coupled, stainless-steel pump (non-adjustable speed) and M38x1.5 male pump connections.
- Water-cooled refrigerant circuit using R-744 (A1, H280), 1st stage; cooling water connections G3/4 male.
- Within the Unimotive family, the GL 10w-XT is the extended-temperature version of the 10w; the 20w and 20w-XT are the larger-capacity siblings. Selection between 10w and 20w is a capacity decision, not a temperature-range decision.
- Controller: Pilot ONE with POKO ONE control output; display resolution 0.01 K; external Pt100 sensor input; interfaces Ethernet, USB (host and device) and RS232.
Working range and media behavior
- Temperature range -45...150 °C, intended for water-ethylene glycol mixtures with corrosion protection (e.g. Glysantin®) down to -45 °C; the unit is not specified for other fluid families.
- Maximum permissible kinematic viscosity of the heat transfer fluid is 200 mm²/s, which bounds fluid choice at the cold end.
- Heating capacity 24 kW. Cooling capacity is not tabulated here; per the datasheet, cooling power drops and cooling water flow may rise as cooling water temperature increases, so performance must be derated to site conditions.
- Cooling water consumption 2100 l/h at 15 °C cooling water with 0 °C flow temperature; minimum differential pressure 1 bar, maximum supply pressure 6 bar.
Installation, control and operation
- Supply 460 V 3~ 60 Hz, 47 A max current, 3x63A fusing; three-phase units below 63 A ship with cable but without plug, so a site disconnect must be provided.
- External branch circuit protection to UL 489 is mandatory; voltage tolerance ±5% and frequency ±2% must not combine unfavorably.
- Cooling water intake 5...20 °C; circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water may be used; a building-side backflow preventer is required.
- Ambient 5...40 °C, IP20, safety class I/NFL; Group 1 Class A (EN 55011) — industrial EMC environment only.
- Minimum filling capacity 8.9 l with a 19.9 l expansion tank; weight 600 kg.
Integration & Operating Notes
The 460 V 3~ 60 Hz unit draws up to 47 A and requires 3x63A fusing with external branch circuit protection according to UL 489. Because three-phase devices under 63 A are delivered with cable but without a plug, the installer must supply a matching receptacle or hardwired disconnect; hardwiring must be performed by a certified electrician.
Voltage tolerance is ±5% and frequency tolerance ±2%, but the datasheet explicitly forbids the unfavorable combination (e.g. -5% voltage together with +2% frequency). Verify the actual site supply envelope before ordering, since a supply at the edge of both bands simultaneously is outside the permitted operating window.
The cooling water intake must stay between 5 °C and 20 °C, with a minimum differential pressure of 1 bar and a maximum supply pressure of 6 bar. Consumption is 2100 l/h at 15 °C cooling water with 0 °C flow temperature; higher cooling water temperatures reduce cooling power and may increase required flow, so the site loop must be sized with margin.
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Only drinking water or filtered river/sea water is permitted, and sea water is explicitly prohibited. A building-side backflow preventer is required to protect the drinking water supply, and cooling water must be shut off during downtimes because flow can occur even with the unit switched off.
The unit is constructed for water-ethylene glycol mixtures with corrosion protection down to -45 °C. The maximum permissible kinematic viscosity is 200 mm²/s, which limits how cold and how concentrated the mixture may be at the pump. Other fluid families are not covered by the datasheet and should not be assumed compatible with the stainless-steel magnetic pump and EPDM seals.
The responsible body must install a condensation water/heat transfer fluid drip tray below the unit. Minimum wall clearances depend on whether the unit is air- or water-cooled and on operation with or without a bath; insufficient clearance prevents adequate energy dissipation. The 600 kg weight also requires a load-rated floor at the installation point.
The unit must not be installed or started inside an ATEX zone. Because the refrigerant is R-744 (A1, H280) with a 3.3 kg charge, the room size must be checked against the charge during installation to avoid dangerous refrigerant concentration. The unit is Group 1 Class A (EN 55011) and is intended for industrial EMC environments only.
The heat transfer fluid must always be adjusted to room temperature before switching off to avoid overpressure in the system. For pressure-sensitive applications such as glass reactors, the maximum flow pressure of the unit must be taken into account and an external bypass may be needed. Hose inside diameters must at least match the M38x1.5 pump connections, with larger diameters for longer lines.
Commissioning requires a minimum filling capacity of 8.9 l with a 19.9 l expansion tank, so the fluid charge and expansion allowance must be planned before start-up. Cooling water lines must be frost-protected if the unit is sited outdoors, and the cooling water temperature must never fall below 5 °C.
Selection & Integration Q&A
Pick the 10w-XT when the duty fits the smaller 10w capacity class but the process needs the extended -45...150 °C range with water-glycol. Move to a 20w sibling when heating/cooling demand exceeds the 24 kW heating class; the non-XT 10w if the extended cold end is not required.
A 460 V 3~ 60 Hz supply at 47 A max with 3x63A fusing and a UL 489 breaker plus a site disconnect (no plug shipped). Cooling water at 5...20 °C, 1...4 bar differential, 6 bar max, 2100 l/h at 15 °C, with a backflow preventer and a drip tray.
Indoor, non-ATEX, 5...40 °C ambient, IP20, Group 1 Class A EMC environment. Respect wall clearances for energy dissipation, check R-744 charge against room volume, and shut off cooling water during downtimes since flow can occur with the unit off.
No plug for the 460 V three-phase cable, no UL 489 breaker, no drip tray, no backflow preventer, and no M38x1.5 pump fittings or G3/4 cooling water hoses. The heat transfer fluid (water-glycol with corrosion protection) and the minimum 8.9 l charge are also site-supplied.