Peter Huber Kältemaschinenbau Unimotive 26w-XT
Peter Huber Kältemaschinenbau Unimotive 26w-XT — Temperature control units.
Water-cooled thermostatic bath/circulator with magnetically coupled stainless-steel pump and 24 kW heating, covering -45 to 150 °C on a 400 V 3~ 50 Hz supply. The XT modification of the Unimotive 26w adds the extended low-temperature range down to -45 °C, distinguishing it from the 10w and 20w siblings. Intended for temperature control of external circuits and bath applications with water-glycol heat transfer fluid.
Peter Huber Kältemaschinenbau Unimotive Series — Model Comparison & Selection
| Model | Display resolution | Heating capacity, W | Power supply | Pump connection | Refrigerant | Safety class |
|---|---|---|---|---|---|---|
| Unimotive 26w-XT Current | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | I / NFL |
| Unimotive 10w | 0.01 | 12000 | 400V 3~ 50Hz | M38x1,5 AG male | R-452A (A1, H280) | III / FL |
| Unimotive 10w-XT | 0, 01 | 12000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | I / NFL |
| Unimotive 20w | 0, 01 | 12000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | III / FL |
| Unimotive 20w-XT | 0, 01 | 12000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | I / NFL |
| Unimotive 26w | 0, 01 K | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | III / FL |
| Unimotive 27w | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | III / FL |
| Unimotive GL 07w | 0, 01 | 12000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 07w-XT | 0, 01 | 12000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | I / NFL |
| Unimotive GL 10w | 0, 01 | 24000 | 460V 3~ 60Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 10w-XT | 0, 01 | 24000 | 460V 3~ 60Hz | M38x1, 5 AG male | R-744 (A1, H280) | I / NFL |
| Unimotive GL 20w | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 20w-XT | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | I / NFL |
| Unimotive GL 30w | 0, 01 | 24000 | 460V 3~ 60Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 50w | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 50w-XT | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | I / NFL |
| Unimotive GL 60w | 0, 01 | 48000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 60w-XT | 0, 01 | 48000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | I / NFL |
Full specifications
| Control output | POKO ONE (potential-free contact) |
| Heating capacity | 24 kW |
| Pump connection | M38x1, 5 AG male |
| Pump speed adjustable | no |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -45...150 °C |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Refrigerant | R-452A (A1, H280) (1st stage) |
| Maximum viscosity | 200 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G3/4 male |
| Cooling water consumption | 1260 l/h (at 15°C, flow 0°C) |
| Cooling water differential pressure | 1 bar (min.) |
| Cooling Water Pressure | 6 bar (max.) |
| Protection rating | IP20 |
| Safety class | I / NFL |
Description
Construction and series positioning
- Water-cooled refrigeration circuit using R-452A (A1, H280) in the first stage; cooling water connection G3/4 male.
- Magnetically coupled stainless-steel pump, non-adjustable speed, pump connection M38x1.5 AG male.
- 24 kW heating capacity; temperature range -45...150 °C; display resolution 0.01 K.
- Within the Unimotive series, the 26w-XT is the larger, extended-range variant; the 10w/10w-XT and 20w/20w-XT siblings differ in size and range and must not be conflated with this model.
Behavior under load and media
- Cooling power falls with decreasing temperature: 28 kW at 20 °C down to 2.6 kW at -40 °C, so low-temperature duty must be sized against the reduced capacity.
- Pump delivery decreases with rising circuit pressure: 201 l/min maximum, 42 l/min at 5.0 bar.
- Designed for water-ethylene glycol mixtures with corrosion protection (e.g. Glysantin®) down to -45 °C; fluid selection must follow this requirement.
- Cooling water consumption 1260 l/h at 15 °C with 0 °C flow; minimum differential pressure 1 bar, maximum cooling water pressure 6 bar.
Installation, control and operation
- Controller Pilot ONE with POKO ONE potential-free contact output; interfaces Ethernet, USB (Host and Device) and RS232; external Pt100 sensor connection.
- 400 V 3~ 50 Hz supply, IP20, leakage current >3.5 mA; three-phase devices below 63 A are delivered with cable but without plug.
- Voltage tolerance ±5% only with simultaneous frequency tolerance ±2%; the combination -5% voltage with +2% frequency is not permitted.
- Cooling water intake temperature 5...20 °C; ambient temperature 5...40 °C; EN55011 Class A Group 1 (industrial EMC environment).
- Safety class I / NFL.
Integration & Operating Notes
Voltage may deviate ±5% only when frequency deviation is simultaneously within ±2%; the combination of -5% voltage with +2% frequency is explicitly not allowed. Verify the site supply before connection, because an out-of-tolerance combination can damage the compressor and electronics.
Three-phase devices below 63 A are delivered with a cable but without a plug, so the installer must provide a suitable plug. The unit must always be connected to a socket with protective earth; hardwiring requires a certified electrician. Leakage current exceeds 3.5 mA, which affects RCD selection.
Cooling water intake must be between 5 °C and 20 °C, with differential pressure between 1 bar and 4 bar and maximum pressure 6 bar. Only drinking water or filtered river/sea water is permitted; sea water must not be used. Circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so the water must be compatible with these.
Cooling water can flow even when the unit is switched off, so the supply should be shut off for shorter downtimes such as overnight. A building-side backflow preventer is required to protect the drinking water supply. For outdoor siting, cooling water lines must be frost-protected and the water temperature must not fall below 5 °C.
Minimum wall clearances depend on air- or water-cooling and on whether the unit runs with or without a bath; insufficient clearance prevents energy dissipation and risks overheating. The refrigerant charge and room size must be checked during installation to avoid dangerous refrigerant concentration. A drip tray for condensate/heat transfer fluid must be installed below the unit by the responsible body.
The maximum flow pressure of the unit must be considered for pressure-sensitive applications such as glass reactors; an external bypass may be required. Hose inside diameters must at least match the pump connections, with larger diameters for longer lines. Before switching off, the heat transfer fluid must be brought to room temperature to avoid overpressure.
The unit must not be installed or started inside an ATEX zone. It is a Group 1 Class A device to EN55011, suited to industrial EMC environments; installation elsewhere may cause interference. Ambient temperature must remain between 5 °C and 40 °C, and IP20 requires a dry indoor environment.
Selection & Integration Q&A
Pick the 26w-XT when the process needs the -45...150 °C range with 24 kW heating and the higher cooling capacity (28 kW at 20 °C, 2.6 kW at -40 °C). The 10w/20w variants are smaller; confirm their own ranges and capacities before substituting.
Cooling water at 5...20 °C with 1...4 bar differential pressure, max 6 bar, G3/4 male connection, and a building-side backflow preventer. Electrically: 400 V 3~ 50 Hz with a PE socket and a site-supplied plug, since three-phase units below 63 A ship with cable but no plug.
Dry indoor location, ambient 5...40 °C, no ATEX zone, industrial EMC environment (EN55011 Class A Group 1). Respect wall clearances per the manual table and check refrigerant charge against room size. Shut off cooling water during downtimes; bring fluid to room temperature before switching off.
The mains plug is not included for three-phase units below 63 A; a drip tray and backflow preventer are site responsibilities. Heat transfer fluid must be a water-ethylene glycol mixture with corrosion protection (e.g. Glysantin®) down to -45 °C. Replacement power cables are limited to 4 m for three-phase connections.