Peter Huber Kältemaschinenbau Unimotive GL 10w
Peter Huber Kältemaschinenbau Unimotive GL 10w — Temperature control units.
The Unimotive GL 10w is a water-cooled, R-744 (CO2) refrigeration temperature control unit built for water-ethylene glycol mixtures from -45 to 95 °C, with 24 kW heating and a magnetically coupled stainless-steel pump. It is the GL variant of the 10w series, adding a water-cooled condenser for facilities with a cooling-water loop instead of the air-cooled Unimotive 10w. It serves automotive test benches and process circuits requiring deep sub-ambient temperature control.
Peter Huber Kältemaschinenbau Unimotive Series — Model Comparison & Selection
| Model | Display resolution | Heating capacity, W | Power supply | Pump connection | Refrigerant | Safety class |
|---|---|---|---|---|---|---|
| Unimotive GL 10w Current | 0, 01 | 24000 | 460V 3~ 60Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive 10w | 0.01 | 12000 | 400V 3~ 50Hz | M38x1,5 AG male | R-452A (A1, H280) | III / FL |
| Unimotive 10w-XT | 0, 01 | 12000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | I / NFL |
| Unimotive 20w | 0, 01 | 12000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | III / FL |
| Unimotive 20w-XT | 0, 01 | 12000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | I / NFL |
| Unimotive 26w | 0, 01 K | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | III / FL |
| Unimotive 26w-XT | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | I / NFL |
| Unimotive 27w | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | III / FL |
| Unimotive GL 07w | 0, 01 | 12000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 07w-XT | 0, 01 | 12000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | I / NFL |
| Unimotive GL 10w-XT | 0, 01 | 24000 | 460V 3~ 60Hz | M38x1, 5 AG male | R-744 (A1, H280) | I / NFL |
| Unimotive GL 20w | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 20w-XT | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | I / NFL |
| Unimotive GL 30w | 0, 01 | 24000 | 460V 3~ 60Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 50w | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 50w-XT | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | I / NFL |
| Unimotive GL 60w | 0, 01 | 48000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 60w-XT | 0, 01 | 48000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | I / NFL |
Full specifications
| Control output | POKO ONE |
| Heating capacity | 24 kW |
| Pump connection | M38x1, 5 AG male |
| Pump speed adjustable | no |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -45...95 °C |
| Weight | 600 kg |
| Current consumption | 47 A (max) |
| Power supply | 460V 3~ 60Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Refrigerant | R-744 (A1, H280) (1st stage) |
| Maximum viscosity | 200 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G3/4 male |
| Cooling water consumption | 2100 l/h (at 15°C, flow 0°C) |
| Cooling water differential pressure | 1 bar (min) |
| Cooling Water Pressure | 6 bar (max) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Water-cooled condenser (cooling-water connection G3/4 male) distinguishes the GL 10w from the air-cooled Unimotive 10w and 10w-XT siblings; the 20w variants are larger-capacity units of the same series.
- Two-stage refrigeration with R-744 (A1, H280) as first-stage refrigerant, charge 3.3 kg, GWP 1.
- Magnetically coupled stainless-steel pump, non-adjustable speed, pump connection M38x1.5 AG male.
- Safety class III / FL with adjustable overtemperature protection per DIN 12876; IP20 enclosure; 600 kg, 730x804x1738 mm.
Working range and media behavior
- Temperature range -45 to 95 °C, intended for water-ethylene glycol mixtures with corrosion protection (e.g. Glysantin®) down to -45 °C.
- Maximum permissible kinematic viscosity 200 mm²/s limits fluid selection at the low end.
- Pump delivers 201 l/min at zero head and 42 l/min at 5.0 bar; system pressure drop must be checked against this curve.
- Cooling capacity derates as cooling-water temperature rises above 15 °C, and cooling-water flow may need to increase; nominal consumption 2100 l/h at 15 °C water / 0 °C flow.
Control, installation and operation
- Pilot ONE controller with POKO ONE control output; display resolution 0.01 K; interfaces Ethernet, USB host and device, RS232; external Pt100 sensor connection.
- Electrical supply 460 V 3~ 60 Hz, 47 A max, 3x63 A fuses, external branch circuit protection per UL 489 required; cable supplied without plug.
- Cooling water 5–20 °C intake, differential pressure 1–4 bar, absolute maximum 6 bar; backflow preventer on building side; shut off cooling water during downtimes.
- EMC class A group 1 to EN55011 (industrial environment); not for ATEX zones.
Integration & Operating Notes
The unit draws 47 A at 460 V 3~ 60 Hz and requires 3x63 A fuses plus external branch circuit protection according to UL 489. Three-phase devices below 63 A are delivered with cable but without plug, so the site must provide the mating connection and protective earth.
The cooling-water circuit requires a differential pressure between 1 and 4 bar and must never exceed 6 bar absolute. Intake temperature is limited to 5–20 °C; below the condensation point the internal cooling-water pipes must be insulated. Only drinking water or filtered river water is permitted; sea water is prohibited.
Nominal cooling-water consumption is 2100 l/h at 15 °C water and 0 °C flow. As cooling-water temperature increases, cooling power drops and the required flow rate may rise, so capacity must be derated for the actual site water temperature rather than assumed from the nominal point.
The unit is constructed for water-ethylene glycol mixtures with corrosion protection (e.g. Glysantin®) down to -45 °C. The maximum permissible kinematic viscosity is 200 mm²/s, which limits fluid choice at the lowest operating temperatures; other fluid families are outside the intended envelope.
Selection & Integration Q&A
Pick the GL 10w when a facility cooling-water loop (5–20 °C, 1–4 bar differential) is available and heat must be rejected to water rather than room air. Choose the air-cooled 10w where no water loop exists; move to the 20w only if the required cooling or heating capacity exceeds this unit's 24 kW heating and its pump curve.
A 460 V 3~ 60 Hz supply within the combined ±5%/±2% envelope, 3x63 A fuses and UL 489 branch protection, plus a mating plug since the cable is supplied without one. Hydraulically: G3/4 male cooling-water connection with 1–4 bar differential, a building-side backflow preventer, and hoses sized at least to the M38x1.5 pump connections.
Ambient 5–40 °C, IP20, not inside an ATEX zone, EMC class A group 1 (industrial only). Cooling water must not fall below 5 °C and lines need frost protection outdoors. Bring fluid to room temperature before switch-off, and shut off cooling water during downtimes since flow can occur with the unit off.
Base delivery excludes the power plug, the drip tray (responsible body must install), the building-side backflow preventer and the heat transfer fluid itself. Cooling-water circuit materials (copper, 1.4401, MS, PA, PPE, PTFE, EPDM) dictate water quality; only drinking or filtered river water is allowed. Replacement power cables are limited to 4 m for three-phase connections.