BINDER BD 115
BINDER BD 115 — Incubators with Temperature 5…100 °C.
The BD 115 is a 112 L natural-convection incubator with a stainless-steel interior and a +5 to 100 °C working range, positioned between the smaller BD 56 and larger BD 260 in the BD series. Its 0.1 K display resolution and 0.3 K uniformity suit standard 37 °C incubation, while the 90 min heat-up to 100 °C supports occasional hot-air disinfection at 100 °C.
Direct drop-in replacements
BINDER BD Series — Model Comparison & Selection
Full specifications
| Temperature | 5...100 °C |
| Heating-up time to 100 °C | 90 min |
| Interior volume | 112 L (BD115-230V / BD115UL-120V) |
| Load per rack | 30 kg (BD115-230V / BD115UL-120V) |
| Nominal power | 0.35 kW |
| Permitted load | 150 kg (BD115-230V / BD115UL-120V) |
| Temperature fluctuation | 0.1 K |
| Temperature uniformity | 0.3 K |
| Back wall clearance | 160 mm (BD115-230V / BD115UL-120V) |
| Side wall clearance | 100 mm (BD115-230V / BD115UL-120V) |
| Article Number mfr | 9010 |
| Recovery time after door was opened for 30 s at 37°C mfr | 15 min |
| Depth | 605 mm (BD115-230V / BD115UL-120V) |
| Height | 705 mm (BD115-230V / BD115UL-120V) |
| Width | 710 mm (BD115-230V / BD115UL-120V) |
| Power frequency | 50 Hz |
| Rated voltage | 230 V |
| Unit fuse | 6.3 A |
| Phase (Nominal voltage) mfr | 1 |
Description
Construction and series positioning
- The BD 115 uses natural convection (no fan) and a stainless-steel 1.4301 (AISI 304) inner chamber, pre-heating chamber and door interior.
- With 112 L interior volume and 710 × 705 × 605 mm net dimensions, it sits between the BD 56 and BD 260; all BD siblings share the +5 to 100 °C range, so the choice is driven by usable volume and load.
- The chamber is equipped with a temperature safety device class 3.1 (DIN 12880) and an overtemperature protective device class 1 that trips at 120 °C.
- The controller displays temperature with 0.1 °C resolution, unlike the 1 °C resolution of ED/FD ovens.
Behavior under load and in operation
- Temperature fluctuation is 0.1 K and temperature uniformity 0.3 K; recovery after a 30 s door opening at 37 °C is 15 min.
- Heating from ambient to 100 °C takes 90 min at 0.35 kW nominal power.
- The usable volume must not be loaded by more than half, and samples must keep at least 10 % of the interior dimensions as wall clearance to maintain airflow and homogeneity.
- The safety controller set-point is an absolute maximum and should be set 2–5 °C above the working set-point; its permissible range is 10–100 °C.
Installation, connection and control
- The BD 115 is supplied as a 230 V 1~ 50 Hz unit (BD115-230V, Art. 9010-0325) with a CEE 7/7 plug; a 120 V 60 Hz UL variant (BD115UL-120V, Art. 9010-0326) uses a NEMA 5-15 plug.
- The external chamber fuse is 6.3 A time-lag (T) 5×20 mm / 250 V; supply fluctuations must stay within ±10 % of nominal voltage.
- Installation requires 160 mm rear clearance and 100 mm side clearance, indoor use on a non-flammable surface, ambient temperature +18 to +40 °C, humidity ≤70 % r.h. non-condensing, and altitude ≤2000 m.
- The chamber is IP 20 and must not be installed in unventilated recesses or explosive areas.
- Options include an RS 422 interface for APT-COM 4 software, an additional Pt100 sensor, an analog temperature output, and a disconnectable audible over-temperature alarm.
Integration & Operating Notes
The chamber requires 160 mm clearance at the rear and 100 mm at the sides for ventilation; installing it in a recess or against a wall without these gaps can cause overheating and invalidate the approved operating conditions.
The BD 115 draws 0.35 kW at 230 V 1~ 50 Hz and relies on an external 6.3 A time-lag (T) 5×20 mm / 250 V chamber fuse; supply voltage must stay within ±10 % of nominal, so verify the circuit and supply quality before connection.
Samples must remain inside the usable volume with at least 10 % of the interior dimensions as clearance from walls, and the usable volume must not be loaded by more than half; exceeding this prevents sufficient airflow and homogeneous temperature distribution.
The safety controller set-point is an absolute maximum limit and must be set approximately 2 °C to 5 °C above the temperature set-point; its permissible range is 10–100 °C, and it must be re-checked after every set-point change to avoid nuisance trips or loss of protection.
Maintenance must be performed at least once a year by licensed electricians or BINDER-authorized experts, and annual repeat electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 is recommended; replace the door gasket only when the chamber is cold to avoid damage.
The BD 115 can be hot-air disinfected at 100 °C / 212 °F; all inflammable goods must be removed from the interior beforehand, and racks can be removed and autoclaved separately.
Selection & Integration Q&A
Pick the BD 115 when 112 L interior volume fits your workload and you need the same +5 to 100 °C range as all BD siblings; choose BD 56 for smaller loads or BD 260 for larger batches, since all share natural convection and 0.1 °C display resolution.
The BD115-230V needs 230 V 1~ 50 Hz with ±10 % tolerance and an external 6.3 A time-lag (T) 5×20 mm / 250 V fuse; the UL variant uses 120 V 60 Hz with a NEMA 5-15 plug. Verify the circuit can supply 0.35 kW continuously.
Install indoors on a non-flammable surface at +18 to +40 °C and ≤70 % r.h. non-condensing, below 2000 m altitude, with 160 mm rear and 100 mm side clearance; do not place in unventilated recesses or explosive areas.
Annual maintenance by BINDER-authorized personnel and electrical safety testing per EN 50678/EN 50699 are required but not included; optional accessories include RS 422/APT-COM 4, additional Pt100 sensor, analog temperature output, and audible over-temperature alarm.