BINDER BD 115
BINDER ·Made in Germany · In Active Production

BINDER BD 115

BINDER BD 115 — Incubators with Temperature 5…100 °C.

The BD 115 is a 112 L natural-convection incubator with a stainless-steel interior and a +5 to 100 °C working range, positioned between the smaller BD 56 and larger BD 260 in the BD series. Its 0.1 K display resolution and 0.3 K uniformity suit standard 37 °C incubation, while the 90 min heat-up to 100 °C supports occasional hot-air disinfection at 100 °C.

Temperature
5…100 °C

Direct drop-in replacements

⚖️ Compare
⚖️ Compare
⚖️ Compare
⚖️ Compare

BINDER BD Series — Model Comparison & Selection

Model Temperature, °CDepth, cmHeating up time to 100 c, minHeight, cmInterior volume, LLoad per rack, kg
BD 115 Current 5–10060.59070.511230
BD 23 5–100525949.52012
BD 260 5–10076759425340
BD 400 5–10076.5120102.540035
BD 56 5–10056.57562.55730
BD 720 5–100879015973745

Full specifications

Core Performance & Functional Specifications
Temperature 5...100 °C
Heating-up time to 100 °C 90 min
Interior volume 112 L (BD115-230V / BD115UL-120V)
Load per rack 30 kg (BD115-230V / BD115UL-120V)
Nominal power 0.35 kW
Permitted load 150 kg (BD115-230V / BD115UL-120V)
Temperature fluctuation 0.1 K
Temperature uniformity 0.3 K
Back wall clearance 160 mm (BD115-230V / BD115UL-120V)
Side wall clearance 100 mm (BD115-230V / BD115UL-120V)
Article Number mfr 9010
Recovery time after door was opened for 30 s at 37°C mfr 15 min
Dimensions, Weight & Mechanical Design
Depth 605 mm (BD115-230V / BD115UL-120V)
Height 705 mm (BD115-230V / BD115UL-120V)
Width 710 mm (BD115-230V / BD115UL-120V)
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 230 V
Unit fuse 6.3 A
Phase (Nominal voltage) mfr 1

Description

Construction and series positioning

Behavior under load and in operation

Installation, connection and control

Integration & Operating Notes

Mounting
Keep 160 mm rear and 100 mm side clearance

The chamber requires 160 mm clearance at the rear and 100 mm at the sides for ventilation; installing it in a recess or against a wall without these gaps can cause overheating and invalidate the approved operating conditions.

Data Sheet Model BD 115 en.pdf; Quick Start Guide, Requirements for the location of installation
Electrical
Provide a 6.3 A time-lag upstream fuse and stable 230 V supply

The BD 115 draws 0.35 kW at 230 V 1~ 50 Hz and relies on an external 6.3 A time-lag (T) 5×20 mm / 250 V chamber fuse; supply voltage must stay within ±10 % of nominal, so verify the circuit and supply quality before connection.

Quick Start Guide, Electrical connection data and installation
Process Limits
Load no more than half the usable volume with 10 % wall clearance

Samples must remain inside the usable volume with at least 10 % of the interior dimensions as clearance from walls, and the usable volume must not be loaded by more than half; exceeding this prevents sufficient airflow and homogeneous temperature distribution.

Quick Start Guide, Placing the charging material in the usable volume
Control & I/O
Set safety controller 2–5 °C above the working set-point

The safety controller set-point is an absolute maximum limit and must be set approximately 2 °C to 5 °C above the temperature set-point; its permissible range is 10–100 °C, and it must be re-checked after every set-point change to avoid nuisance trips or loss of protection.

Quick Start Guide, Set-point entry for temperature and Setting the safety controller set-point
Maintenance
Schedule annual maintenance and electrical safety testing

Maintenance must be performed at least once a year by licensed electricians or BINDER-authorized experts, and annual repeat electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 is recommended; replace the door gasket only when the chamber is cold to avoid damage.

Operating Manual BD-ED-FD (E2), Section 10.1 and 10.2
Cooling & Media
Hot-air disinfection at 100 °C is permitted for BD

The BD 115 can be hot-air disinfected at 100 °C / 212 °F; all inflammable goods must be removed from the interior beforehand, and racks can be removed and autoclaved separately.

Quick Start Guide, Cleaning and decontamination

Selection & Integration Q&A

When should I choose the BD 115 over the BD 56 or BD 260?

Pick the BD 115 when 112 L interior volume fits your workload and you need the same +5 to 100 °C range as all BD siblings; choose BD 56 for smaller loads or BD 260 for larger batches, since all share natural convection and 0.1 °C display resolution.

What supply and upstream protection does the BD 115 require?

The BD115-230V needs 230 V 1~ 50 Hz with ±10 % tolerance and an external 6.3 A time-lag (T) 5×20 mm / 250 V fuse; the UL variant uses 120 V 60 Hz with a NEMA 5-15 plug. Verify the circuit can supply 0.35 kW continuously.

What ambient conditions and clearances must the installation site meet?

Install indoors on a non-flammable surface at +18 to +40 °C and ≤70 % r.h. non-condensing, below 2000 m altitude, with 160 mm rear and 100 mm side clearance; do not place in unventilated recesses or explosive areas.

What maintenance and accessories are not included in the base delivery?

Annual maintenance by BINDER-authorized personnel and electrical safety testing per EN 50678/EN 50699 are required but not included; optional accessories include RS 422/APT-COM 4, additional Pt100 sensor, analog temperature output, and audible over-temperature alarm.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Data Sheet Model BD 115 en
Datasheet ·642,1 КБ ·7 pp.
View PDF
PDF
QuickStartGuide BD BF ED FD FED E31 03 2023 dt en fr
Manual ·1,7 МБ ·35 pp.
View PDF
PDF
BD ED FD (E2) 09 2026 en
Manual ·12,9 МБ ·71 pp.
View PDF