Peter Huber Kältemaschinenbau CC-112A
Peter Huber Kältemaschinenbau CC-112A — Thermostatic baths.
The CC-112A is a 12-litre polycarbonate-bath immersion circulator with a variable-speed pressure/suction pump and Pilot ONE controller. Its 25–100 °C base range extends to 20–100 °C with water cooling and 15–100 °C with a refrigerator, distinguishing it from smaller CC-104A–CC-110A siblings. It serves as a temperature-control source for external circuits and open bath work up to 100 °C.
Direct drop-in replacements
Peter Huber Kältemaschinenbau Mit Polycarbonatbaedern Bis 100C Series — Model Comparison & Selection
| Model | Controller | Interface | Max delivery flow rate, liter/minute | Pump connection | Weight, kg | Ambient temperature limit, °C |
|---|---|---|---|---|---|---|
| CC-112A Current | Pilot ONE | Ethernet, USB (Host u. D | 27 | M16x1 AG male (Option, # | 8 | 5–40 |
| CC-104A | Pilot ONE | Ethernet, USB (Host u. D | 22 | M16x1 AG male | 6 | 5–40 |
| CC-106A | Pilot ONE | Ethernet, USB (Host u. D | 22 | M16x1 AG male | 5 | 5–40 |
| CC-108A | Pilot ONE | Ethernet, USB (Host u. D | 27 | M16x1 AG male (Option, # | 6 | 5–40 |
| CC-110A | Pilot ONE | Ethernet, USB (Host u. D | 22 | M16x1 AG male | 6 | 5–40 |
| CC-118A | Pilot ONE | Ethernet, USB (Host u. D | 22 | M16x1 AG male (Option, # | 8 | 5–40 |
| KISS 104A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male | 5 | 5–40 |
| KISS 106A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 5 | 5–40 |
| KISS 108A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 6 | 5–40 |
| KISS 110A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 6 | 5–40 |
| KISS 112A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 8 | 5–40 |
| KISS 118A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 8 | 5–40 |
Full specifications
| Bath volume | 12 l |
| Filling volume | 8 l (min.) |
| Maximum delivery flow rate | 27 l/min (pressure pump max.; also 0, 7 bar) suction pump max.; also 0, 4 bar: 22 l/min |
| Pump connection | M16x1 AG male (Option, #19606) (Option, #19606) |
| Pump speed adjustable | Yes |
| Pump type | Pressure and suction pump, variable speed |
| External sensor connection | Pt100 |
| Temperature range | 25...100 °C with cooling unit: 15...100 °C · with water cooling: 20...100 °C |
| Temperature stability mfr | ±0, 02 K |
| Dimensions (W x D x H) mfr | 333 x 360 x 335 mm |
| Heating capacity mfr | 1, 5 - 2, 1 kW |
| Bath opening WxD/BD mfr | 275 x 161 / 150 mm |
| Weight | 8 kg |
| Current consumption | 10 A (max.) |
| Power supply | 200-240V 1~/2~ 50/60Hz V, Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
The CC-112A is the largest bath in the CC-10xA polycarbonate series, carrying a 12 l bath volume with a minimum filling capacity of 8 l. The bath opening measures 275 x 161 mm with a 150 mm depth, and the unit occupies 333 x 360 x 335 mm at 8 kg. Unlike the smaller CC-104A through CC-110A siblings, this modification provides the full 12 l working volume and the highest heating capacity of the group at 1.5–2.1 kW. The bath bridge includes a bore hole for a cooling probe, allowing an immersion cooler such as the TC45–TC100E to be added without changing the circulator.
Thermal behaviour and media
- Base operating range is 25–100 °C; adding water cooling lowers the minimum to 20 °C, and a refrigerator configuration reaches 15 °C. The achievable minimum must be matched to the chosen cooling method before selection.
- Temperature stability is ±0.02 K, maintained by the Pilot ONE controller with an external Pt100 sensor connection.
- The variable-speed pressure and suction pump delivers up to 27 l/min at 0.7 bar on the pressure side and 22 l/min at 0.4 bar on the suction side, supporting both open-bath circulation and closed external circuits.
- Safety classification III / FL per DIN 12876 with adjustable overtemperature protection; the unit is rated IP20 and must be kept dry.
Installation and control
Power is supplied from 200–240 V, single- or two-phase, 50/60 Hz, with a maximum current of 10 A. The datasheet requires a branch fuse of at least 10 A and no more than 16 A, plus external branch circuit protection to UL 489. Voltage tolerance is ±10% provided the frequency tolerance does not run in the opposite direction. Single- and two-phase units ship with a power cable and country-specific plug that must be specified at order. Communication and control interfaces include Ethernet, USB host and device, and RS232. Ambient temperature must stay within 5–40 °C.
Integration & Operating Notes
The datasheet specifies a minimum fuse of 10 A and a maximum fuse of 16 A for the 200–240 V supply. Undersizing will trip on the 10 A maximum current draw; oversizing removes the intended protection. External branch circuit protection to UL 489 is also required.
A ±10% voltage tolerance is permitted only when the frequency tolerance does not run in the opposite direction. Verify the site supply against both limits together, not independently, before commissioning.
The unit reaches 25 °C without auxiliary cooling, 20 °C with water cooling, and 15 °C only with a refrigerator. Select the cooling configuration before specifying the process setpoint; the bath bridge bore hole accepts an immersion cooler such as TC45–TC100E for intermediate needs.
When a Huber pump adapter is fitted, polyglycol must not be used as the heat transfer fluid. Confirm fluid compatibility with the pump adapter before filling; the M16x1 male pump connection is an option (#19606).
The IP20 rating provides no protection against water ingress. Install the unit away from splashing water and maintain ambient temperature between 5 °C and 40 °C. Respect the manufacturer's space requirements for the operating conditions before final placement.
Single- and two-phase devices in the 100–240 V class are delivered with a power cable and a country-specific plug that must be specified at the time of ordering. Omitting this detail delays commissioning because the plug type is not universal.
The equipment is classified Class A, Group 1 under EN55011. In residential or EMC-sensitive environments, additional filtering or separation from sensitive instrumentation may be needed to avoid disturbance.
Selection & Integration Q&A
Pick the CC-112A when the application needs the full 12 l bath volume, an 8 l minimum fill, and the 1.5–2.1 kW heating capacity for faster recovery or larger immersed loads. Smaller siblings suit lower-volume work where the 275 x 161 mm opening and 12 l volume are unnecessary.
A 200–240 V single- or two-phase, 50/60 Hz supply with a maximum current of 10 A, protected by a branch fuse of at least 10 A and no more than 16 A, plus external UL 489 branch circuit protection. Voltage tolerance is ±10% only if frequency tolerance does not oppose it.
No. The enclosure is IP20 and not protected against water ingress. Install it in a dry location with ambient temperature between 5 °C and 40 °C, away from splashing water, and respect the manufacturer's space requirements for the operating conditions.
The country-specific plug must be specified at ordering; the M16x1 male pump connection is an option (#19606); and an immersion cooler such as TC45–TC100E is a separate accessory for the bath bridge bore hole. External branch circuit protection to UL 489 is also a site responsibility.