Peter Huber Kältemaschinenbau CC-208B
Peter Huber Kältemaschinenbau CC-208B — Thermostatic baths.
The CC-208B is a 7.5-litre stainless-steel bath with an immersion circulator and a variable-speed pressure/suction pump, rated for -30…200 °C only when a refrigerator is fitted, 20…200 °C with water cooling, and 25…200 °C without external cooling. Its Pilot ONE controller, Pt100 external sensor input and ±0.02 K stability distinguish it from smaller CC-202C/CC-205B siblings, while the 27 l/min pressure pump serves closed external circuits in laboratory and pilot processes.
Direct drop-in replacements
Peter Huber Kältemaschinenbau Mit Edelstahlbaedern Bis 200C Series — Model Comparison & Selection
| Model | Controller | Interface | Pump connection | Temperature range, °C | Weight, kg | Max delivery flow rate, liter/minute |
|---|---|---|---|---|---|---|
| CC-208B Current | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male | 25–200 | 10 | 27 |
| CC-202C | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male | 45–200 | 9 | 22 |
| CC-205B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male | 45–200 | 9 | 27 |
| CC-212B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male (Option, # | 25–200 | 11 | 22 |
| CC-215B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male | 25–200 | 12 | — |
| CC-220B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male (Option, # | 25–200 | 14 | 27 |
| CC-225B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male (Option, # | 25–200 | 16 | 27 |
| KISS 202C | KISS | USB (Device), RS232 Inte | M16x1 AG male | 45–200 | 8 | 14 |
| KISS 205B Badumwälzthermostat mit Edelstahlbad | KISS | USB (Device), RS232 Inte | M16x1 AG male | 45–200 | 9 | 14 |
| KISS 208B | KISS | USB (Device), RS232 Inte | M16x1 AG male (Option, # | 25–200 | 10 | 14 |
| KISS 212B | KISS | USB (Device), RS232 Inte | M16x1 AG male | 25–200 | 11 | 14 |
| KISS 215B | KISS | USB (Device), RS232 Inte | M16x1 AG male | 25–200 | 12 | 14 |
| KISS 220B | KISS | USB (Device), RS232 Inte | M16x1 AG male (Option, # | 25–200 | 14 | 14 |
| KISS 225B | KISS | USB (Device), RS232 Inte | M16x1 AG male | 25–200 | 16 | 14 |
Full specifications
| Filling volume | 5 l (minimum) |
| Maximum delivery flow rate | 27 l/min (pressure pump; also 0, 7 bar) suction pump; also 0, 4 bar: 22 l/min |
| Pump connection | M16x1 AG male (Option, #19607) |
| Pump speed adjustable | Yes |
| Pump type | Pressure and suction pump, variable speed |
| External sensor connection | Pt100 |
| Temperature range | 25...200 °C with cooling unit: -30...200 °C · with water cooling: 20...200 °C |
| Temperature stability mfr | ±0, 02 K |
| Dimensions (W x D x H) mfr | 290 x 350 x 375 mm |
| Heating capacity mfr | 1, 5 - 2, 1 kW |
| Bath opening WxD/BD mfr | 230 x 127 / 150 mm |
| Bath volume mfr | 7, 5 l |
| Weight | 10 kg |
| Current consumption | 10 A (max.) |
| Power supply | 200-240V 1~/2~ 50/60Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- The CC-208B is a 7.5 l bath (minimum filling 5 l) with a 230 × 127 mm opening, 150 mm bath depth, and a stainless-steel bath bridge drilled for a cooling probe.
- Overall dimensions are 290 × 350 × 375 mm at 10 kg; the unit is IP20 and safety class III / FL with adjustable overtemperature protection to DIN 12876.
- Within the CC-20xB series, the 208B sits between the smaller CC-205B and the larger CC-212B/CC-215B; the CC-202C is a different, lower-capacity variant. Choose the 208B when the 7.5 l volume and 1.5–2.1 kW heating capacity match the load, and step to 212B/215B only for larger bath volumes.
Thermal behaviour and media
- The achievable temperature range is set by the cooling configuration, not by the circulator alone: 25…200 °C without external cooling, 20…200 °C with water cooling, and -30…200 °C with a refrigerator. The cooling option must therefore be fixed at order.
- Temperature stability is ±0.02 K; an external Pt100 sensor can be connected for control at the point of use.
- The variable-speed pressure/suction pump delivers up to 27 l/min at 0.7 bar on the pressure side and 22 l/min at 0.4 bar on the suction side, so external loop pressure drop must be checked against these figures.
- Polyglycol must not be used as heat-transfer fluid when a Huber pump adapter is fitted.
Control, installation and operation
- The Pilot ONE controller provides Ethernet, USB host/device and RS232 interfaces; pump speed is adjustable.
- Supply is 200–240 V, 1~ or 2~, 50/60 Hz, max. 10 A, with a branch circuit rated 10–16 A and external UL 489 protection required. Voltage and frequency tolerances are each ±10% but must not run in opposite directions.
- Single- and two-phase units ship with a power cable and country-specific plug, which must be specified on the order.
- Ambient operating range is 5–40 °C; EMC classification is Class A, Group 1 to EN55011, so the unit is intended for non-domestic environments. The equipment falls under Article 4.3 PED.
Integration & Operating Notes
External branch circuit protection to UL 489 is mandatory and is not included in the unit. Size the branch for 10–16 A at 200–240 V, 1~ or 2~, 50/60 Hz, and confirm the site feed can carry the 10 A maximum current without nuisance tripping.
Each of voltage and frequency may deviate ±10%, but the two tolerances must not run in opposite directions at the same time. Verify the site supply envelope before connecting, especially on weak grids or generator-fed installations.
Without external cooling the bath is limited to 25…200 °C; water cooling extends it to 20…200 °C, and only a refrigerator reaches -30…200 °C. Specify the cooling method at order, because the range cannot be extended later by controller settings alone.
The bath holds 7.5 l and requires at least 5 l. Underfilling risks pump cavitation and poor temperature control; overfilling risks overflow as the fluid expands at high temperature. Check the level after the loop is filled and at operating temperature.
When a Huber pump adapter is fitted, polyglycol is not permissible as heat-transfer fluid. Select an alternative fluid compatible with the wetted materials and the 200 °C upper limit, and confirm compatibility before filling.
The unit is IP20 and must run between 5 °C and 40 °C ambient, so wet or dusty locations are excluded. EMC classification Class A, Group 1 to EN55011 means it is intended for non-domestic environments; plan the site accordingly.
A drain valve with cap (#6839) is an accessory, not part of base delivery. Without it, emptying the 7.5 l bath for fluid changes or service requires manual draining; include it in the initial order to avoid later rework.
Selection & Integration Q&A
Pick the 208B when the 7.5 l bath and 1.5–2.1 kW heating match the load. The 205B is smaller; the 212B/215B offer larger volumes. The 202C is a different, lower-capacity variant. Step up only if the external loop or sample size exceeds the 208B's volume.
A 200–240 V, 1~ or 2~, 50/60 Hz feed rated 10–16 A with external UL 489 branch protection, which is not included. Specify the country-specific plug at order. Voltage and frequency may each deviate ±10% but not in opposite directions.
Yes. The variable-speed pressure/suction pump delivers up to 27 l/min at 0.7 bar pressure and 22 l/min at 0.4 bar suction. Check the external loop's pressure drop against these figures. A Huber pump adapter is available, but polyglycol must not be used with it.
Operate between 5 °C and 40 °C ambient; IP20 excludes wet or dusty sites, and Class A EMC suits non-domestic environments. The drain valve (#6839) and the immersion cooler for sub-ambient work are accessories, not included; the bath bridge is pre-drilled for the cooling probe.