Peter Huber Kältemaschinenbau CC-212B
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau CC-212B

Peter Huber Kältemaschinenbau CC-212B — Thermostatic baths.

The CC-212B is a 10.5-litre stainless-steel bath immersion circulator with a variable-speed pressure/suction pump and Pilot ONE controller, rated for 25–200 °C without cooling, 20–200 °C with water cooling, and -30–200 °C with a refrigerator. Its 1.5–2.1 kW heater and ±0.02 K stability distinguish it within the CC-2xxB series for precise external temperature control.

Direct drop-in replacements

Eppendorf
MixMate ®
58%
match
⚖️ Compare

Peter Huber Kältemaschinenbau Mit Edelstahlbaedern Bis 200C Series — Model Comparison & Selection

Model ControllerInterfacePump connectionTemperature range, °CWeight, kgMax delivery flow rate, liter/minute
CC-212B Current Pilot ONEEthernet, USB (Host u. DM16x1 AG male (Option, #25–2001122
CC-202C Pilot ONEEthernet, USB (Host u. DM16x1 AG male45–200922
CC-205B Pilot ONEEthernet, USB (Host u. DM16x1 AG male45–200927
CC-208B Pilot ONEEthernet, USB (Host u. DM16x1 AG male25–2001027
CC-215B Pilot ONEEthernet, USB (Host u. DM16x1 AG male25–20012—
CC-220B Pilot ONEEthernet, USB (Host u. DM16x1 AG male (Option, #25–2001427
CC-225B Pilot ONEEthernet, USB (Host u. DM16x1 AG male (Option, #25–2001627
KISS 202C KISSUSB (Device), RS232 InteM16x1 AG male45–200814
KISS 205B Badumwälzthermostat mit Edelstahlbad KISSUSB (Device), RS232 InteM16x1 AG male45–200914
KISS 208B KISSUSB (Device), RS232 InteM16x1 AG male (Option, #25–2001014
KISS 212B KISSUSB (Device), RS232 InteM16x1 AG male25–2001114
KISS 215B KISSUSB (Device), RS232 InteM16x1 AG male25–2001214
KISS 220B KISSUSB (Device), RS232 InteM16x1 AG male (Option, #25–2001414
KISS 225B KISSUSB (Device), RS232 InteM16x1 AG male25–2001614

Full specifications

Core Performance & Functional Specifications
Dimensions (w × D × H) 350 x 375 x 375 mm
Filling volume 7 l (min.)
Maximum delivery flow rate 27 l/min (pressure pump max.; 0, 7 bar)
suction pump max.; 0, 4 bar: 22 l/min
Pump connection M16x1 AG male (Option, #19607) (Option, #19607)
Pump speed adjustable Yes
Pump type Pressure and suction pump, variable speed
External sensor connection Pt100
Temperature range 25...200 °C
with cooling unit: -30...200 °C · with water cooling: 20...200 °C
Temperature stability mfr ±0, 02 K
Heating capacity mfr 1, 5 - 2, 1 kW
Bath opening WxD/BD mfr 290 x 152 / 150
Bath volume mfr 10, 5 l
Dimensions, Weight & Mechanical Design
Weight 11 kg
Electrical, Power & Drive Specifications
Current consumption 10 A (max.)
Power supply 200-240V 1~/2~ 50/60Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

The CC-212B is a bath-mounted immersion circulator with a 10.5-litre stainless-steel bath, a minimum filling capacity of 7 litres, and a bath opening of 290 × 152 mm (width × depth) with a 150 mm bath depth. It uses the Pilot ONE controller and a variable-speed pressure and suction pump. Within the CC-2xxB series, the CC-212B sits between the smaller CC-208B and the larger CC-215B; the CC-205B and CC-202C are lower-capacity siblings. The CC-212B is the choice when a 10.5-litre bath and the higher pump delivery (27 l/min pressure, 22 l/min suction) are needed for external circuit work.

Thermal behaviour and media

Temperature range is configuration-dependent: 25–200 °C without cooling, 20–200 °C with water cooling, and -30–200 °C with a refrigerator. Temperature stability is ±0.02 K. Heating power scales with supply voltage from 1.5 kW at 200 V to 2.1 kW at 240 V. The pump delivers up to 27 l/min at 0.7 bar (pressure side) and 22 l/min at 0.4 bar (suction side), with adjustable speed. The bath bridge includes a hole for a cooling probe such as an immersion cooler TC45–TC100E, allowing sub-ambient operation without extra modification. When using the Huber pump adapter, polyglycol must not be used as heat transfer fluid.

Installation, control and operation

The unit accepts 200–240 V single- or two-phase supply at 50/60 Hz and draws up to 10 A maximum. External branch circuit protection according to UL 489 is required. Voltage and frequency tolerances are ±10% only if they do not combine in opposite directions. Single- and two-phase devices are delivered with a power cable and country-specific plug that must be specified at order. Interfaces include Ethernet, USB (Host and Device), and RS232. An external Pt100 sensor can be connected. The unit has adjustable overtemperature protection per DIN 12876 and safety classification III/FL. Ambient temperature during operation must stay between 5 °C and 40 °C. The equipment is Class A, Group 1 per EN55011. A drain valve with cap (#6839) is available as an accessory for emptying the bath.

Integration & Operating Notes

Electrical
Branch protection must be UL 489 listed and sized for 10 A

The CC-212B draws up to 10 A at 200–240 V; external branch circuit protection according to UL 489 is mandatory. Size the breaker and wiring for the maximum current and confirm that the supply voltage and frequency tolerances do not combine in opposite directions (e.g. -10% voltage with +3% frequency is not permitted).

Datasheet, Technical Specifications, page 2
Process Limits
Cooling method determines the usable lower temperature limit

Without auxiliary cooling the bath is limited to 25–200 °C; water cooling extends the lower limit to 20 °C, and a refrigerator is required to reach -30 °C. Plan the cooling source before specifying the unit for sub-ambient work, and note that the bath bridge has a hole for an immersion cooler such as TC45–TC100E.

Datasheet, Technical Specifications, page 1
Process Media
Polyglycol is incompatible with the Huber pump adapter

When the Huber pump adapter is fitted, polyglycol must not be used as the heat transfer fluid. Select an alternative fluid for the intended temperature range and confirm compatibility with the pump adapter and bath materials before commissioning.

Datasheet, page 2
Mounting
Order the country-specific plug with the unit

Single- and two-phase devices are delivered with a power cable and a country-specific plug that must be specified at the time of ordering. Confirm the site's outlet type and supply configuration (1~ or 2~) before placing the order to avoid a mismatch on delivery.

Datasheet, page 2
Environment
Ambient temperature must stay between 5 °C and 40 °C

Operating the unit outside the 5–40 °C ambient range may affect temperature stability and component life. Ensure the installation location is conditioned accordingly, and note that the equipment is Class A, Group 1 per EN55011, which may require additional EMC measures in residential or sensitive environments.

Datasheet, Technical Specifications, page 1
Maintenance
Use the drain valve accessory for safe bath emptying

A drain valve with cap (#6839) is available as an accessory to simplify fluid changes and maintenance. Plan for its installation if frequent fluid replacement is expected, and follow the fluid manufacturer's disposal and handling instructions.

Datasheet, page 2

Selection & Integration Q&A

When should I choose the CC-212B instead of the CC-208B or CC-215B?

Choose the CC-212B when a 10.5-litre bath and the higher pump delivery (27 l/min pressure, 22 l/min suction) are needed. The CC-208B has a smaller bath, and the CC-215B is larger; select the CC-212B for a balance of bath volume and external circuit flow.

What are the requirements for the electrical supply and branch protection?

The unit requires 200–240 V single- or two-phase 50/60 Hz supply with a maximum current of 10 A. External branch circuit protection according to UL 489 is mandatory. Voltage and frequency tolerances are ±10% only if they do not combine in opposite directions.

What ambient conditions and duty-cycle limits apply to the CC-212B?

The ambient temperature during operation must remain between 5 °C and 40 °C. The unit is rated for continuous operation within its temperature range, but the heating power depends on supply voltage (1.5 kW at 200 V to 2.1 kW at 240 V), which affects heat-up time.

What maintenance accessories and consumables are not included in the base delivery?

The base delivery does not include a drain valve with cap (#6839) or a cooling probe/immersion cooler (e.g. TC45–TC100E). The country-specific plug must be specified at order. Heat transfer fluid is also not included and must be selected according to the application.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
CC-212B Immersion Circulator
Datasheet ·70,6 КБ ·2 pp.
View PDF