Peter Huber Kältemaschinenbau CC-220B
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau CC-220B

Peter Huber Kältemaschinenbau CC-220B — Thermostatic baths.

The CC-220B is a 17-litre stainless-steel bath with a Pilot ONE immersion circulator, rated for 25...200 °C without cooling, 20...200 °C with water cooling and -30...200 °C with a refrigerator. Its variable-speed pressure/suction pump delivers up to 27 l/min at 0.7 bar and 22 l/min at 0.4 bar. It serves as a precision thermostating source for external circuits and open-bath work at ±0.02 K stability.

Direct drop-in replacements

Eppendorf
MixMate ®
53%
match
⚖️ Compare

Peter Huber Kältemaschinenbau Mit Edelstahlbaedern Bis 200C Series — Model Comparison & Selection

Model ControllerInterfacePump connectionTemperature range, °CWeight, kgMax delivery flow rate, liter/minute
CC-220B Current Pilot ONEEthernet, USB (Host u. DM16x1 AG male (Option, #25–2001427
CC-202C Pilot ONEEthernet, USB (Host u. DM16x1 AG male45–200922
CC-205B Pilot ONEEthernet, USB (Host u. DM16x1 AG male45–200927
CC-208B Pilot ONEEthernet, USB (Host u. DM16x1 AG male25–2001027
CC-212B Pilot ONEEthernet, USB (Host u. DM16x1 AG male (Option, #25–2001122
CC-215B Pilot ONEEthernet, USB (Host u. DM16x1 AG male25–20012—
CC-225B Pilot ONEEthernet, USB (Host u. DM16x1 AG male (Option, #25–2001627
KISS 202C KISSUSB (Device), RS232 InteM16x1 AG male45–200814
KISS 205B Badumwälzthermostat mit Edelstahlbad KISSUSB (Device), RS232 InteM16x1 AG male45–200914
KISS 208B KISSUSB (Device), RS232 InteM16x1 AG male (Option, #25–2001014
KISS 212B KISSUSB (Device), RS232 InteM16x1 AG male25–2001114
KISS 215B KISSUSB (Device), RS232 InteM16x1 AG male25–2001214
KISS 220B KISSUSB (Device), RS232 InteM16x1 AG male (Option, #25–2001414
KISS 225B KISSUSB (Device), RS232 InteM16x1 AG male25–2001614

Full specifications

Core Performance & Functional Specifications
Bath volume 17 l
Dimensions (w × D × H) 350 x 555 x 375 mm
Filling volume 12 l (minimum)
Maximum delivery flow rate 27 l/min (pressure pump; 0, 7 bar)
suction pump; 0, 4 bar: 22 l/min
Pump connection M16x1 AG male (Option, #19607) (Option, #19607)
Pump speed adjustable Yes
Pump type Pressure and suction pump, variable speed
External sensor connection Pt100
Temperature range 25...200 °C
with cooling unit: -30...200 °C · with water cooling: 20...200 °C
Temperature stability mfr ±0, 02 K
Heating capacity mfr 1, 5 - 2, 1 kW
Bath opening WxD/BD mfr 290 x 329 / 150 mm (WxD/depth)
Dimensions, Weight & Mechanical Design
Weight 14 kg
Electrical, Power & Drive Specifications
Current consumption 10 A (max.)
Power supply 200-240V 1~/2~ 50/60Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Thermal behaviour and media

Installation, control and operation

Integration & Operating Notes

Electrical
Branch protection must sit between 10 A and 16 A

The datasheet specifies a minimum fuse of 10 A and a maximum of 16 A with a maximum current draw of 10 A, and requires external branch circuit protection to UL 489. Oversizing the breaker beyond 16 A leaves the unit without the protection the manufacturer assumes; undersizing causes nuisance trips at run-up.

CC-220B Immersion Circulator Datasheet, Specifications / installation notes
Electrical
Voltage and frequency tolerances must not oppose each other

Supply may deviate ±10% in voltage only as long as the frequency tolerance does not run in the opposite direction. A site with low voltage and high frequency simultaneously falls outside the permitted envelope even though each deviation alone would be acceptable, so site power quality should be measured before commissioning.

CC-220B Immersion Circulator Datasheet, installation notes
Cooling & Media
No polyglycol when the Huber pump adapter is fitted

With the Huber pump adapter installed, polyglycol is not permissible as heat transfer fluid. Fluid selection must therefore be fixed together with the pump-connection decision (optional M16x1 male, #19607), because changing one later invalidates the other and may require draining and flushing the bath and external loop.

CC-220B Immersion Circulator Datasheet, operation notes
Mounting
IP20 enclosure rules out splash and wet locations

Degree of protection is IP20, so the unit has no water-ingress protection and must be sited dry, away from dripping or splashing liquid and from filling or draining operations of adjacent equipment. Ambient must stay within 5...40 °C; the 14 kg mass and 350 x 555 x 375 mm footprint define the bench space and lifting plan.

CC-220B Immersion Circulator Datasheet, Specifications
Maintenance
Fit a drain valve before the first fill

A drain valve with cap (#6839) is an accessory for emptying the bath; without it, fluid removal at change intervals means lifting or siphoning a 12–17 l charge. Decide on the drain valve at order time, since retrofitting after commissioning requires draining the bath anyway.

CC-220B Immersion Circulator Datasheet, accessories

Selection & Integration Q&A

When should I choose the CC-220B rather than the smaller baths in the same series?

Pick the 220B when the 17 l volume and 290 x 329 mm opening are needed for immersed samples or as buffer for an external loop; the smaller CC-202C/205B/208B/212B siblings suit smaller loads. Pump head and temperature range are not the differentiator, so size on volume and opening first.

What must I provide on the electrical side before the CC-220B can be commissioned?

A 200–240 V, 1~ or 2~, 50/60 Hz supply with a 10–16 A branch fuse and external protection to UL 489, within ±10% voltage and frequency tolerances that do not oppose. Specify the country-specific plug at order, since single/two-phase units ship with cable and plug only on request.

Can I control temperature at my external application rather than in the bath?

Yes. The Pilot ONE controller accepts an external Pt100 sensor, so the control point can be placed at the load. The variable-speed pressure/suction pump then has to be set against the external circuit's pressure drop, using 27 l/min at 0.7 bar on the pressure side and 22 l/min at 0.4 bar on the suction side as the limits.

What accessories and consumables are not in the base delivery?

The M16x1 male pump connection (#19607), the drain valve with cap (#6839) and any immersion cooler (TC45–TC100E) for sub-ambient work are options, as is the refrigerator for -30 °C. Heat transfer fluid is not supplied; if the Huber pump adapter is used, polyglycol is excluded.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
CC-220B Immersion Circulator Datasheet
Datasheet ·66,0 КБ ·2 pp.
View PDF