Peter Huber Kältemaschinenbau Grande Fleur Dynamisches
Peter Huber Kältemaschinenbau Grande Fleur Dynamisches — Temperature control units with Noise level 61 dB.
The Grande Fleur Dynamisches is an air-cooled, variable-speed pressure-pump circulator with a -40 to 200 °C working range and 0.6 kW cooling capacity down to 0 °C, falling to 0.04 kW at -40 °C. It is the base Dynamisches variant of the Grande Fleur series, distinguished from the -eo and w siblings by its configuration rather than by a different temperature envelope. It serves as a temperature-control source for external circuits requiring 47 l/min maximum delivery through an M24x1.5 male pump connection.
Peter Huber Kältemaschinenbau Unistate Bis 45C Series — Model Comparison & Selection
| Model | Noise level, dB | Display resolution | Power supply | Pump connection | Pump type | Temperature range, °C |
|---|---|---|---|---|---|---|
| Grande Fleur Dynamisches Current | — | — | — | — | — | — |
| Grande Fleur w Dynamisches | — | 0, 01 | 208-240V 1~/2~ 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | -40–200 |
| Grande Fleur w-eo Dynamisches | — | — | — | — | — | — |
| Petite Fleur-eo Dynamisches | 58 | 0, 01 K | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | Pressure pump, variable | -40–200 |
| Petite Fleur w Dynamisches | — | 0, 01 | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | Pressure pump, variable | -40–200 |
| Unistat 410w Dynamisches | 61 | 0, 01 K | 220-240V 1~/2~ 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | -45–250 |
| Unistat 425w Dynamisches | 61 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
| Unistat 430w Dynamisches | 61 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
| Unistat P405w Dynamisches | — | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st | -45–250 |
| Unistat P425 Dynamisches | 65 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
| Unistat P425w Dynamisches | 61 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
| Unistat P430 Dynamisches | 65 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
| Unistat P430w Dynamisches | 65 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
Full specifications
Description
Construction and series positioning
- Air-cooled refrigeration machine using R-290 (A3, H220) natural refrigerant, GWP 0.02, first stage; pressure equipment category 4.3 PED.
- Variable-speed pressure pump, maximum delivery 47 l/min, pump pressure 0.9 bar, M24x1.5 male connection.
- Within the Grande Fleur Dynamisches family, the base model shares the 47 l/min flow figure with the Grande Fleur-eo Dynamisches; the w and w-eo variants are separate configurations. The smaller Petite Fleur Dynamisches is the lower-noise (58 dB(A)) alternative when acoustic budget dominates.
- Controller Pilot ONE with 5.7" touchscreen, 0.01 K display resolution and stated 0.01 K temperature stability at -10 °C; control outputs POKO ONE, control input ECS ONE, external Pt100 sensor connection.
Behavior across the working range
- Cooling capacity is flat at 0.6 kW from 0 °C up to 200 °C, holds 0.35 kW at -20 °C, then collapses to 0.2 kW at -30 °C and 0.04 kW at -40 °C. Low-temperature duty near the lower limit therefore supports only small heat loads.
- Heating capacity is stated as 1.5–2 kW; the unit covers the full -40 to 200 °C span without a separate stage.
- Heat-transfer fluid must stay at or below 50 mm²/s kinematic viscosity, which constrains fluid choice at the cold end.
Installation, connection and control
- Supply 208–240 V, 1~ or 2~, 50/60 Hz; maximum current 11.5 A, 16 A fuse. Voltage tolerance ±5% only with ±2% frequency tolerance, and the combination -5% voltage with +2% frequency is explicitly not allowed.
- Delivered with power cable and country-specific plug, which must be specified at order.
- IP20 enclosure and air-cooled condenser require a dry, protected location with adequate ventilation; ambient range 5–40 °C.
- Interfaces: Ethernet, USB host and device, RS232; alarm output optic, acoustic and relay.
- Fluid charge: minimum filling 1.5 l, 3 l expansion tank.
Integration & Operating Notes
Near the lower temperature limit the available cooling capacity falls from 0.6 kW at 0 °C to 0.2 kW at -30 °C and 0.04 kW at -40 °C. Low-temperature processes must be sized against these figures, not the nominal 0.6 kW, or the unit will be unable to pull down or hold setpoint under load.
Voltage tolerance is ±5% only when frequency tolerance is simultaneously ±2%, and the specific combination of -5% voltage and +2% frequency is not permitted. Supply quality must be checked at the installation point before commissioning, since an out-of-envelope supply invalidates the rating.
Single- and two-phase units ship with a power cable and a country-specific plug that must be stated when ordering. The 208–240 V, 1~/2~, 50/60 Hz supply, 11.5 A maximum current and 16 A fuse define the site circuit; a missing plug specification delays commissioning.
The refrigeration machine is air-cooled, so the installation site must provide airflow and clearance for heat rejection. The IP20 enclosure is not water-protected and is unsuitable for wet or washdown areas; ambient must remain within 5–40 °C, and cooling capacity may drop as ambient temperature rises.
The maximum permissible kinematic viscosity is 50 mm²/s. At low temperatures many fluids thicken beyond this, overloading the pump and reducing flow; fluid selection must be checked against the coldest operating point, not only at room temperature.
Minimum filling capacity is 1.5 l and the expansion tank holds 3 l. The charge must cover the external circuit plus the unit's own volume while leaving headroom for thermal expansion across the -40 to 200 °C range, otherwise underfilling or overflow occurs.
The pump connection is M24x1.5 male, so flexible tempering hoses or adapters on the external circuit must carry the matching female thread. Mismatched fittings cannot be adapted at the port without a correct adapter, and the 47 l/min maximum delivery sets the hose and circuit pressure-drop budget.
The first stage uses R-290 (A3, H220), a flammable natural refrigerant with GWP 0.02 and UN number 3358. Installation and safety planning must account for the A3 classification; the unit is also EN55011 Class A, Group 1, so EMC measures may be needed in sensitive or residential environments.
Selection & Integration Q&A
Pick the base Dynamisches when the standard 47 l/min variable-speed pressure pump and the -40 to 200 °C range fit the duty. The -eo shares the 47 l/min flow; the w and w-eo are separate configurations to be evaluated on their own data. Choose the Petite Fleur Dynamisches only if 58 dB(A) acoustics outweigh the smaller platform.
The pump port is M24x1.5 male, so hoses or adapters need the matching female thread. The circuit plus unit must take at least 1.5 l charge with 3 l expansion volume, and the fluid must stay at or below 50 mm²/s viscosity at the coldest setpoint. Maximum delivery is 47 l/min at 0.9 bar pump pressure.
Dry, protected location: IP20 and air-cooled condenser need ventilation, ambient 5–40 °C. Supply 208–240 V, 1~/2~, 50/60 Hz, 11.5 A max, 16 A fuse; ±5% voltage only with ±2% frequency, and -5% voltage with +2% frequency is not allowed. Specify the country-specific plug at order.
The power cable and country-specific plug are supplied only if the plug type is specified at order. Heat-transfer fluid is not part of the unit charge and must be selected within the 50 mm²/s viscosity limit. External Pt100 sensors and circuit hoses/adapters for the M24x1.5 port are separate items to be provided.