Peter Huber Kältemaschinenbau Huber CS 550-H OD
Peter Huber Kältemaschinenbau Huber CS 550-H OD — Recirculating chillers with Noise level 79 dB.
The Huber CS 550-H OD is a 37-litre refrigerated circulating bath with a -20 to 85 °C working range, 17 kW heating capacity and a 125 l/min pump through a G1 male connection. It is the outdoor-rated, high-capacity member of the Cs Chiller series, distinguished by IP54 electricals and a 20 m remote panel. It serves as a central temperature-control unit for industrial process loops requiring a 160–195 litre fluid inventory.
Peter Huber Kältemaschinenbau Cs Chiller Series — Model Comparison & Selection
| Model | Noise level, dB | Ambient temperature limit, °C | Minimum fill volume, m³ | Power supply | Pump connection | Temperature range, °C |
|---|---|---|---|---|---|---|
| Huber CS 550-H OD Current | 79 | -20–50 | 0.16 | 460V 3~ 60Hz | G1 AG male | -20–85 |
| CS 100 by Van der Heijden | 70 | 5–32 | 0.16 | 400V 3~ 50Hz | G3/4 AG male | -20–15 |
| CS 100-H by Van der Heijden | 71 | 5–32 | 0.175 | 400V 3~ 50Hz | G3/4 AG male | -20–85 |
| CS 200 by Van der Heijden | 78 | 5–32 | 0.225 | 400V 3~ 50Hz | G1 AG male | -20–15 |
| CS 200-H by Van der Heijden | 78 | 5–32 | 0.225 | 400V 3~ 50Hz | G1 AG male | -20–85 |
| CS 20-H by Van der Heijden | 57 | 5–32 | 0.024 | 230V 1~ 50Hz | G3/8 AG male | -20–85 |
| CS 25 by Van der Heijden | 60 | 5–32 | 0.024 | 230V 1~ 50Hz | G1/2 AG male | -20–15 |
| CS 50 by Van der Heijden | 63 | 5–32 | 0.1 | 400V 3~ 50Hz | G3/4 AG male | -20–15 |
| CS 75 by Van der Heijden | 66 | 5–32 | 0.16 | 400V 3~ 50Hz | G3/4 AG male | -20–15 |
| Huber CS 200-H | 78 | 5–32 | 0.225 | 460V 3~ 60Hz | G1 AG male | -20–85 |
| Huber CS 25 | 60 | 5–32 | 0.024 | 208V 2~ 60Hz | G1/2 AG male | -20–15 |
| Huber CS 250HT OD | 78 | -20–50 | 0.16 | 460V 3~ 60Hz | G3/4 AG male | 10–22 |
| Huber CS 330-H OD | 78 | -20–50 | 0.16 | 460V 3~ 60Hz | G1 AG male | -20–85 |
| Huber CS 550 OD | 79 | -20–50 | 0.16 | 460V 3~ 60Hz | G1 AG male | -20–15 |
| Huber LCS 80 | 77 | -20–50 | 0.04 | 460V 3~ 60Hz | G1 AG male | -45–15 |
Full specifications
| Noise level | 79 dB(A) |
| Bath volume | 37 l |
| Heating capacity | 17 kW |
| Minimum fill volume | 160 l |
| Pump connection | G1 AG male |
| Temperature range | -20...85 °C |
| Temperature stability | 2 K |
| Weight | 715 kg (net) |
| Current consumption | 73 A (max) |
| Power supply | 460V 3~ 60Hz |
| Digital interface | RS232 |
| Ambient temperature limit | -20...50 °C |
| Safety class | I / NFL |
Description
Construction and series positioning
The CS 550-H OD is a refrigerated circulating bath built for outdoor installation down to -20 °C ambient, with electrical components rated IP54. Within the Cs Chiller series it is the largest and only outdoor-rated variant; the sibling CS 100/125 models are indoor units with lower noise levels (70–75 dB(A)) and smaller footprints. The 37-litre bath volume and 160–195 litre filling range define a substantial fluid inventory, and the 715 kg net weight reflects the 23 kg R-449A refrigerant charge and 17 kW heater.
Thermal and hydraulic behaviour
Cooling capacity is strongly ambient-dependent: it is reduced above +35 °C and must be sized against the actual site ambient, with a maximum of 50 °C. The -20 to 85 °C range requires a 50:50 water–monoethylene glycol mixture for freeze protection and viscosity. The pump delivers up to 125 l/min at 4.7 bar through a G1 male connection; flow is balanced by a manual bypass valve on the rear of the chiller, so rear access is mandatory for commissioning.
Electrical and control integration
The unit runs on 460 V 3~ 60 Hz with a maximum current consumption of 73 A. Because this exceeds 63 A, it is delivered without cable and without plug, so the customer must provide the full three-phase power connection. Voltage and frequency tolerances are coupled: a -5 % voltage combined with +2 % frequency is not permitted. Control is via RS232 and a remote operating panel connected by a 20 m cable, which constrains panel location during site planning. The equipment is EMC Class A Group 1 (EN55011), intended for industrial environments only.
Integration & Operating Notes
With a maximum current consumption of 73 A, the unit exceeds the 63 A threshold above which three-phase devices are delivered without cable and without plug. The installer must size and provide the full 460 V 3~ 60 Hz feed, including protection and isolation, before commissioning.
A -5 % voltage combined with +2 % frequency is explicitly not allowed. The supply must be checked as a combined condition; a supply that is individually within voltage and frequency limits but simultaneously at both extremes will violate the permitted operating envelope.
The flow rate is set by a manual bypass valve on the backside of the chiller. Site layout must leave clear access to the rear of the unit for commissioning and any later flow adjustment; a wall-backed or enclosed installation prevents correct hydraulic balancing.
Cooling power is reduced above +35 °C ambient and the unit is rated to a maximum of 50 °C. Selection must use the real installation ambient rather than a nominal capacity figure; outdoor siting also requires frost protection planning down to -20 °C.
The recommended heat transfer fluid is a 50:50 water–monoethylene glycol mixture. This ratio is required for freeze protection at the lower end of the -20 to 85 °C range and for correct viscosity through the pump and external circuit.
The unit contains 23 kg of R-449A (A1, H280), equal to 32.13 t CO2 equivalent. This charge level brings refrigerant reporting obligations and scheduled leak-check intervals; service personnel must follow H280 pressure and ventilation precautions.
The operating panel is connected via a 20 m cable. Panel location must be planned within that cable run from the chiller; the RS232 interface provides the serial control link but does not extend the panel cable.
Selection & Integration Q&A
Choose the CS 550-H OD when you need outdoor installation (to -20 °C ambient), a 160–195 litre fluid inventory, 17 kW heating and 125 l/min pump flow. The CS 100/125 siblings are indoor units with lower noise (70–75 dB(A)) and smaller baths, suited to lower-capacity indoor loops.
The 460 V 3~ 60 Hz unit draws up to 73 A, above the 63 A threshold, so it ships without cable and without plug. The customer must provide the complete three-phase feed, protection and isolation, and must verify combined voltage/frequency tolerance (-5 % V with +2 % Hz is not allowed).
It is rated for outdoor mounting down to -20 °C ambient and up to 50 °C, with cooling power reduced above +35 °C. Electrical components are IP54. Rear access is required for the manual bypass valve, and the 20 m remote panel cable constrains panel placement.
The base unit excludes the power cable and plug (73 A > 63 A). The 23 kg R-449A charge (32.13 t CO2e) requires F-gas leak-check scheduling and H280 precautions. The 50:50 water–monoethylene glycol fluid must be procured to fill the 160–195 litre bath.