Peter Huber Kältemaschinenbau KISS 108A
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau KISS 108A

Peter Huber Kältemaschinenbau KISS 108A — Thermostatic baths.

The KISS 108A is a 6-litre immersion circulator with a pressure/suction pump and 1.5–2.1 kW heater, operating from 25 to 100 °C (down to 15 °C only with an external cooling unit). It is the 6-litre member of the KISS series, distinguished by its fixed-speed pump and 200–240 V single- or two-phase supply. It serves as a temperature-control source for external circuits or open baths in laboratory processes.

Direct drop-in replacements

Eppendorf
MixMate ®
58%
match
⚖️ Compare

Peter Huber Kältemaschinenbau Mit Polycarbonatbaedern Bis 100C Series — Model Comparison & Selection

Model ControllerInterfaceMax delivery flow rate, liter/minutePump connectionWeight, kgAmbient temperature limit, °C
KISS 108A Current KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #65–40
CC-104A Pilot ONEEthernet, USB (Host u. D22M16x1 AG male65–40
CC-106A Pilot ONEEthernet, USB (Host u. D22M16x1 AG male55–40
CC-108A Pilot ONEEthernet, USB (Host u. D27M16x1 AG male (Option, #65–40
CC-110A Pilot ONEEthernet, USB (Host u. D22M16x1 AG male65–40
CC-112A Pilot ONEEthernet, USB (Host u. D27M16x1 AG male (Option, #85–40
CC-118A Pilot ONEEthernet, USB (Host u. D22M16x1 AG male (Option, #85–40
KISS 104A KISSUSB (Device), RS232 Inte14M16x1 AG male55–40
KISS 106A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #55–40
KISS 110A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #65–40
KISS 112A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #85–40
KISS 118A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #85–40

Full specifications

Core Performance & Functional Specifications
Alarm output optic, acoustic
Bath volume 6 l
Maximum delivery flow rate 14 l/min (pressure pump max.; 0, 25 bar)
Pump connection M16x1 AG male (Option, #19606) (Option, #19606)
Pump speed adjustable no
Pump type Pressure and suction pump, speed not variable
Temperature range 25...100 °C
with cooling unit: 15...100 °C · with water cooling: 20...100 °C
Temperature stability mfr ±0, 05 K
Dimensions (W x D x H) mfr 147 x 407 x 330 mm
Heating capacity mfr 1, 5 - 2, 1 kW
Delivery capacity suction pump max. mfr 10, 5 l/min (suction pump max.; 0, 17 bar)
Bath opening WxD/BD mfr 130 x 210 / 150 mm
Filling volume min. mfr 3, 5 l (min.)
Dimensions, Weight & Mechanical Design
Weight 6 kg
Electrical, Power & Drive Specifications
Current consumption 10 A (max.)
Power supply 200-240V 1~/2~ 50/60Hz V, Hz
Interfaces, Connections & Control
Controller KISS
Digital interface USB (Device), RS232 Interface
Temperature display OLED Display
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Thermal behaviour and media

Installation and operation

Integration & Operating Notes

Electrical
Branch circuit must include a UL 489 listed breaker

The datasheet requires external branch circuit protection according to UL 489. The installer must provide this breaker; without it the installation is non-compliant. Size the circuit for the maximum current of 10 A at 200–240 V.

Datasheet, Technical Data (installation
Electrical
Voltage and frequency tolerances must not oppose each other

The ±10% voltage tolerance applies only as long as the frequency tolerance does not run in the opposite direction. Verify the site supply envelope before commissioning; combined deviations in opposite directions exceed the allowed operating range.

Datasheet, Technical Data (installation
Electrical
Country-specific plug must be stated at order

Single- and two-phase devices (100 V to 240 V) are delivered with a power cable and a country-specific plug. The plug type must be specified when ordering, otherwise the unit may not connect to the site outlet.

Datasheet, Technical Data (installation
Cooling & Media
Low-temperature operation requires an ordered cooling option

The base range is 25–100 °C. Reaching 20 °C requires water cooling; reaching 15 °C requires a refrigerator. The cooling method must be selected to match the required process temperature before ordering.

Datasheet, Technical Data (selection
Cooling & Media
Plastic pump is incompatible with acetone and polyglycols

The pump is not resistant to acetone or polyglycols (depending on the manufacturer). Avoid these media or order a more resistant plastic to prevent premature pump failure.

Datasheet, Technical Data (operation
Cooling & Media
Use glysantine or ethylene glycol for freeze protection

For freeze protection, water should be mixed with glysantine or ethylene glycol rather than used pure, to avoid freezing and damage at low temperatures.

Datasheet, Technical Data (operation
Process Limits
Fill between 3.5 l minimum and 6 l bath volume

The bath has a 6 l volume with a minimum filling capacity of 3.5 l. Underfilling risks pump cavitation and poor temperature control; overfilling risks overflow.

Datasheet, Technical Data (operation
Environment
Operate only within 5–40 °C ambient temperature

The unit must only be operated within an ambient temperature of 5 °C to 40 °C. Operation outside this range affects reliability and warranty compliance.

Datasheet, Technical Data (operation
Control & I/O
Autostart restarts the unit automatically after power failure

The autostart function restarts the unit automatically after a power failure. Engineers must assess whether unattended restart is acceptable for their process, particularly where a hot bath restart could affect samples or safety.

Datasheet, Technical Data (safety
Process Limits
Pump performance limits external circuit flow and head

The pressure pump delivers up to 14 l/min at 0.25 bar; the suction side delivers up to 10.5 l/min at 0.17 bar. Check that the external circuit flow and head requirements fall within these limits.

Datasheet, Technical Data (performance
Electrical
Heating power depends on the actual supply voltage

Heating power ranges from 1.5 kW at 200 V to 2.1 kW at 240 V. Heat-up time must be matched to the actual site voltage.

Datasheet, Technical Data (performance
Environment
Class A equipment is intended for industrial EMC environments

The unit is classified as EN 55011 Class A, Group 1 for electromagnetic disturbance. Class A equipment is intended for industrial environments and may not be suitable for residential EMC environments.

Datasheet, Technical Data (compliance

Selection & Integration Q&A

When should I choose the KISS 108A rather than a sibling model such as the CC-104A or CC-110A?

Choose the KISS 108A when a 6 l bath volume with a fixed-speed pressure/suction pump and 25–100 °C base range fits the process. Select a sibling CC model when a different bath capacity is required; the 108A is the 6-litre member of the series.

What external circuit components and supply conditions must be provided for the KISS 108A?

Provide a 200–240 V, 1~ or 2~, 50/60 Hz supply with a UL 489 listed branch breaker, sized for 10 A max. The pressure pump supplies up to 14 l/min at 0.25 bar and suction up to 10.5 l/min at 0.17 bar; match external circuit head and flow to these limits.

What maintenance and consumables are not included in the base delivery of the KISS 108A?

The base delivery does not include the country-specific plug (must be specified at order), external cooling water or refrigerator for sub-25 °C operation, or freeze-protection additives (glysantine or ethylene glycol). The M16x1 AG male pump connection is an option (#19606).

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
KISS 108A Immersion Circulator
Datasheet ·104,3 КБ ·2 pp.