Peter Huber Kältemaschinenbau KISS 220B
Peter Huber Kältemaschinenbau KISS 220B — Thermostatic baths.
The KISS 220B is a 17-litre stainless-steel bath immersion circulator with a fixed-speed pressure/suction pump, rated for 25–200 °C (20–200 °C with water cooling, −30–200 °C with refrigerator). Its 200–240 V 1~/2~ supply and 10 A draw distinguish it within the KISS series. It serves as a temperature-control source for external circuits and immersed samples in laboratory processes.
Direct drop-in replacements
Peter Huber Kältemaschinenbau Mit Edelstahlbaedern Bis 200C Series — Model Comparison & Selection
| Model | Controller | Interface | Pump connection | Temperature range, °C | Weight, kg | Max delivery flow rate, liter/minute |
|---|---|---|---|---|---|---|
| KISS 220B Current | KISS | USB (Device), RS232 Inte | M16x1 AG male (Option, # | 25–200 | 14 | 14 |
| CC-202C | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male | 45–200 | 9 | 22 |
| CC-205B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male | 45–200 | 9 | 27 |
| CC-208B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male | 25–200 | 10 | 27 |
| CC-212B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male (Option, # | 25–200 | 11 | 22 |
| CC-215B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male | 25–200 | 12 | — |
| CC-220B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male (Option, # | 25–200 | 14 | 27 |
| CC-225B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male (Option, # | 25–200 | 16 | 27 |
| KISS 202C | KISS | USB (Device), RS232 Inte | M16x1 AG male | 45–200 | 8 | 14 |
| KISS 205B Badumwälzthermostat mit Edelstahlbad | KISS | USB (Device), RS232 Inte | M16x1 AG male | 45–200 | 9 | 14 |
| KISS 208B | KISS | USB (Device), RS232 Inte | M16x1 AG male (Option, # | 25–200 | 10 | 14 |
| KISS 212B | KISS | USB (Device), RS232 Inte | M16x1 AG male | 25–200 | 11 | 14 |
| KISS 215B | KISS | USB (Device), RS232 Inte | M16x1 AG male | 25–200 | 12 | 14 |
| KISS 225B | KISS | USB (Device), RS232 Inte | M16x1 AG male | 25–200 | 16 | 14 |
Full specifications
| Alarm output | optic, acoustic |
| Bath volume | 17 l |
| Dimensions (w × D × H) | 350 x 555 x 375 mm |
| Filling volume | 10 l (min) |
| Maximum delivery flow rate | 14 l/min (pressure pump; 0, 25 bar) |
| Pump connection | M16x1 AG male (Option, #19607) (Option, #19607) |
| Pump speed adjustable | no |
| Pump type | Pressure and suction pump, speed not variable |
| Temperature range | 25...200 °C with cooling unit: -30...200 °C · with water cooling: 20...200 °C |
| Temperature stability mfr | ±0, 05 K |
| Heating capacity mfr | 1, 5 - 2, 1 kW |
| Delivery capacity suction pump max. mfr | 10, 5 l/min (suction pump; 0, 17 bar) |
| Bath opening WxD/BD mfr | 290 x 329 / 150 mm (WxD/depth) |
| Weight | 14 kg |
| Current consumption | 10 A (max) |
| Power supply | 200-240V 1~/2~ 50/60Hz |
| Controller | KISS |
| Digital interface | USB (Device), RS232 Interface |
| Temperature display | OLED Display |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
The KISS 220B is an immersion circulator with a 17-litre stainless-steel bath (minimum filling 10 l) and a bath opening of 290 × 329 mm, 150 mm deep, with 205 mm opening height. It uses the KISS controller with OLED display and provides USB (Device) and RS232 interfaces. Within the series, this model is the 220B variant; sibling models CC-202C, CC-205B, CC-208B and CC-212B are separate configurations and their specifications do not apply here. The unit weighs 14 kg and is rated IP20, safety class III/FL per DIN 12876.
Thermal behaviour and media
Temperature range depends on the cooling method: 25–200 °C without active cooling, 20–200 °C with water cooling, and −30–200 °C with a refrigerator. Temperature stability is ±0.05 K. Heating capacity is 1.5–2.1 kW. The pump is a pressure and suction type with fixed speed (not adjustable): pressure side up to 14 l/min at 0.25 bar, suction side up to 10.5 l/min at 0.17 bar. The plastic pump is not resistant to acetone or polyglycols; for freeze protection, water should be mixed with glycerine or ethylene glycol.
Installation and control
Supply is 200–240 V, 1~ or 2~, 50/60 Hz, with maximum current 10 A. Branch circuit protection must be 10–16 A and external protection per UL 489 is required. Voltage and frequency tolerances are ±10% and must not run in opposite directions. Ambient operating temperature is 5–40 °C. An autostart function restarts the unit after power failure. An optional Pt100 sensor connection (#10688, factory-fitted only) displays but does not control process temperature. An M16×1 AG male pump connection is available as option #19607.
Integration & Operating Notes
The unit draws up to 10 A and requires external branch circuit protection per UL 489 rated between 10 A and 16 A. Verify that the available supply is within 200–240 V and 50/60 Hz with ±10% tolerance, and that voltage and frequency deviations do not occur in opposite directions, to avoid operating outside the specified envelope.
Without active cooling the bath operates from 25 °C to 200 °C; water cooling extends the lower limit to 20 °C; a refrigerator extends it to −30 °C. Select the cooling configuration before specifying the process temperature, because the achievable range is not the same across configurations.
The pump's plastic wetted parts are not resistant to acetone or polyglycols (depending on manufacturer). For freeze protection, mix water with glycerine or ethylene glycol rather than using pure water. Confirm fluid compatibility before filling to prevent pump damage and contamination.
An autostart function automatically restarts the unit after a power failure. For unattended operation, assess whether this behaviour is acceptable for the process, or provide external sequencing to prevent unexpected restart of heating and circulation.
The optional Pt100 sensor connection (#10688) is factory-fitted only and is intended to display an external process temperature; it does not provide closed-loop control. Order it with the unit if external temperature monitoring is required, since it cannot be added later in the field.
Selection & Integration Q&A
Choose the KISS 220B when you need a 17-litre bath with a 25–200 °C base range and fixed-speed pressure/suction pump. Sibling CC models are separate configurations; compare their published bath volumes and temperature ranges before selecting, as their specifications do not apply to the 220B.
Provide a 200–240 V, 1~ or 2~, 50/60 Hz supply with 10–16 A branch protection per UL 489. The pressure pump delivers up to 14 l/min at 0.25 bar and suction up to 10.5 l/min at 0.17 bar; size external tubing and fittings to stay within these limits. An M16×1 AG male pump connection is available as option #19607.
The Pt100 sensor connection (#10688) and M16×1 AG male pump connection (#19607) are options ordered separately; the Pt100 is factory-fitted only. No maintenance schedule is given in the supplied datasheet; consult the operating manual for service intervals. Use glycerine or ethylene glycol for freeze protection.