Peter Huber Kältemaschinenbau KISS 225B
Peter Huber Kältemaschinenbau KISS 225B — Thermostatic baths.
The KISS 225B is a 23.5 l stainless-steel bath with an integrated KISS immersion circulator, rated 25–200 °C without cooling, 20–200 °C with water cooling and -30–200 °C with a refrigerator. Its non-variable pressure/suction pump delivers 14 l/min at 0.25 bar and 10.5 l/min at 0.17 bar, distinguishing it from smaller CC-series baths. It serves as a temperature-controlled bath for external circulation or direct immersion in laboratory processes.
Direct drop-in replacements
Peter Huber Kältemaschinenbau Mit Edelstahlbaedern Bis 200C Series — Model Comparison & Selection
| Model | Controller | Interface | Pump connection | Temperature range, °C | Weight, kg | Max delivery flow rate, liter/minute |
|---|---|---|---|---|---|---|
| KISS 225B Current | KISS | USB (Device), RS232 Inte | M16x1 AG male | 25–200 | 16 | 14 |
| CC-202C | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male | 45–200 | 9 | 22 |
| CC-205B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male | 45–200 | 9 | 27 |
| CC-208B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male | 25–200 | 10 | 27 |
| CC-212B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male (Option, # | 25–200 | 11 | 22 |
| CC-215B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male | 25–200 | 12 | — |
| CC-220B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male (Option, # | 25–200 | 14 | 27 |
| CC-225B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male (Option, # | 25–200 | 16 | 27 |
| KISS 202C | KISS | USB (Device), RS232 Inte | M16x1 AG male | 45–200 | 8 | 14 |
| KISS 205B Badumwälzthermostat mit Edelstahlbad | KISS | USB (Device), RS232 Inte | M16x1 AG male | 45–200 | 9 | 14 |
| KISS 208B | KISS | USB (Device), RS232 Inte | M16x1 AG male (Option, # | 25–200 | 10 | 14 |
| KISS 212B | KISS | USB (Device), RS232 Inte | M16x1 AG male | 25–200 | 11 | 14 |
| KISS 215B | KISS | USB (Device), RS232 Inte | M16x1 AG male | 25–200 | 12 | 14 |
| KISS 220B | KISS | USB (Device), RS232 Inte | M16x1 AG male (Option, # | 25–200 | 14 | 14 |
Full specifications
| Filling volume | 15 l |
| Maximum delivery flow rate | 14 l/min (pressure pump; 0, 25 bar) |
| Pump connection | M16x1 AG male (Option, #19607) |
| Pump speed adjustable | no |
| Pump type | Pressure and suction pump, speed not variable |
| Temperature range | 25...200 °C with cooling unit: -30...200 °C · with water cooling: 20...200 °C |
| Temperature stability mfr | ±0, 05 K |
| Alarm message mfr | optic, acoustic |
| Dimensions (W x D x H) mfr | 350 x 555 x 425 mm |
| Heating capacity mfr | 1, 5 - 2, 1 kW |
| Delivery capacity suction pump max. mfr | 10, 5 l/min (suction pump; 0, 17 bar) |
| Bath opening WxD/BD mfr | 290 x 329 / 200 mm |
| Bath volume mfr | 23, 5 l |
| Weight | 16 kg |
| Current consumption | 10 A |
| Power supply | 200-240V 1~/2~ 50/60Hz |
| Controller | KISS |
| Digital interface | USB (Device), RS232 Interface |
| Temperature display | OLED Display |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- The KISS 225B combines a 23.5 l stainless-steel bath (minimum filling 15 l) with a KISS controller and OLED display; overall dimensions are 350 × 555 × 425 mm and weight 16 kg.
- The bath opening measures 290 × 329 mm with a 200 mm depth. The bath bridge includes a hole for a cooling probe (e.g. immersion cooler TC45–TC100E), with a holder available as an accessory.
- Within the series, the KISS 225B is the larger-bath variant; sibling CC-202C/205B/208B/212B models are not described here and should be selected only when their own verified data applies.
Thermal behavior and media
- The usable range is configuration-dependent: 25–200 °C without cooling, 20–200 °C with water cooling, and -30–200 °C with a refrigerator. Temperature stability is ±0.05 K; heating capacity is 1.5–2.1 kW.
- The plastic pump is not resistant to acetone or polyglycols (depending on manufacturer); for freeze protection, water should be mixed with glysantine or ethylene glycol rather than used pure.
- Safety class III / FL with adjustable overtemperature protection per DIN 12876; alarm messages are optic and acoustic.
Installation, control and operation
- Supply is 200–240 V 1~/2~ 50/60 Hz with max. current 10 A; branch protection must be between 10 A and 16 A, and an external branch circuit protection device according to UL 489 is required.
- Voltage and frequency tolerances are ±10% only if they do not combine in the opposite direction. Single- and two-phase devices are delivered with a power cable and country-specific plug, which must be specified at order.
- Interfaces are USB (Device) and RS232. The autostart function restarts the unit automatically after a power failure. Enclosure is IP20; ambient temperature must remain 5–40 °C.
Integration & Operating Notes
The unit draws up to 10 A at 200–240 V; the datasheet requires an external branch circuit protection device according to UL 489 and permits a fuse between 10 A and 16 A. Undersizing will trip on heat-up; oversizing leaves the supply cable unprotected.
The standard plastic pump is not resistant to acetone or polyglycols (depending on manufacturer). Using these media will degrade the pump; select an alternative fluid or consult the factory for a more resistant pump before filling the 23.5 l bath.
When operating near or below ambient, water should be mixed with glysantine or ethylene glycol for freeze protection. Pure water can freeze and damage the bath or pump; the bath bridge also provides a hole for an immersion cooler TC45–TC100E if active cooling is needed.
The autostart function automatically restarts the unit after a power failure. For unattended processes this may be desired, but where a restart could be hazardous or invalidate a test, the function must be disabled or the process interlocked externally.
Single- and two-phase devices are delivered with a power cable and country-specific plug; the plug type must be stated when ordering. Missing this detail delays commissioning because the cable is not field-configurable to all plug standards.
Selection & Integration Q&A
Choose the KISS 225B when the process needs a 23.5 l bath (15 l minimum fill) and up to 200 °C with 1.5–2.1 kW heating. Smaller CC models should be selected only when their own verified volume and temperature data match the application; the KISS 225B is the larger-bath option in this series.
Without cooling the range is 25–200 °C; with water cooling 20–200 °C; with a refrigerator -30–200 °C. The bath bridge has a hole for an immersion cooler TC45–TC100E (holder accessory). For freeze protection, mix water with glysantine or ethylene glycol, not pure water.
Supply is 200–240 V 1~/2~ 50/60 Hz, max. 10 A. Provide an external branch circuit protection device per UL 489 with a fuse between 10 A and 16 A. Voltage and frequency may deviate ±10% only if not in opposite directions. Specify the country-specific plug at order.
The base unit includes the bath and immersion circulator but not the country-specific plug unless specified, nor the immersion cooler TC45–TC100E or its holder. The plastic pump is not resistant to acetone or polyglycols; alternative pump material must be requested. No maintenance schedule is given in the supplied datasheet.