Peter Huber Kältemaschinenbau TC50E
Peter Huber Kältemaschinenbau TC50E — Immersion coolers.
The TC50E is an air-cooled immersion cooler with a 48 mm × 155 mm probe on a 1150 mm flexible line, covering -50 to 50 °C with 0.3 kW at 0 °C. It is the smallest single-stage unit in the series, distinguished by its R-290 charge of only 0.06 kg. It serves low-temperature cooling of baths and vessels where no pump is required.
Peter Huber Kältemaschinenbau Eintauchkuehler Kuehlfallen Series — Model Comparison & Selection
| Model | Noise level, dB | Controller | Power supply | Refrigerant | Safety class | Temperature range, °C |
|---|---|---|---|---|---|---|
| TC50E Current | — | MPC | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -50–50 |
| CT50 Single OLÉ | 59 | OLÉ | 220-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | III / FL | -50–50 |
| TC100 | 55 | ON-OFF-Regler | 220-240V 1~/2~ 50/60Hz | R-1150 (A3, H220) | I / NFL | -100–40 |
| TC100E-F | 55 | MPC | 110-120V 1~ 60Hz | R-1150 (A3, H220) | I / NFL | -100–40 |
| TC100-F | 58 | ON-OFF-Regler | 220-240V 1~/2~ 50/60Hz | R-1150 (A3, H220) | I / NFL | -100–40 |
| TC45 | — | ON-OFF-Regler | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -45–100 |
| TC45E | — | MPC | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -45–100 |
| TC45E-F | 54 | MPC | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -45–100 |
| TC45-F | 59 | ON-OFF-Regler | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -45–100 |
| TC50 | — | ON-OFF-Regler | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -50–50 |
| TC50E-F | 60 | MPC | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -50–50 |
| TC50-F | 60 | ON-OFF-Regler | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -50–50 |
Full specifications
| Pump speed adjustable | no |
| Pump type | no pump |
| External sensor connection | Pt100 |
| Temperature range | -50...50 °C |
| Temperature stability mfr | ±0, 5 K |
| Alarm message mfr | optic, acoustic |
| Dimensions (W x D x H) mfr | 260 x 330 x 415 mm |
| Refrigeration machine mfr | air-cooled, natural refrigerant |
| Weight | 25 kg |
| Current consumption | 3 A (max.) |
| Power supply | 208-240V 1~/2~ 50/60Hz |
| Controller | MPC |
| Temperature display | LED Display |
| Refrigerant | R-290 (A3, H220) (1st stage) |
| Refrigerant quantity 1st stage mfr | 0, 06 kg (1st stage) |
| Ambient temperature limit | 5...40 °C |
| Cooling capacity mfr | 0: 0, 3; -20: 0, 26; -30: 0, 2 kW (at °C) |
| Safety class | I / NFL |
Description
Construction and series positioning
The TC50E is a direct-immersion refrigeration unit: an air-cooled condensing unit connected by a 1150 mm flexible line to a 48 mm diameter, 155 mm long probe that is placed directly into the process bath or vessel. There is no pump and no pump-speed control, so the unit cools a bath that is mixed by other means. Within the series it is the smallest single-stage model; the TC100/TC100E/TC100E-F siblings are larger, and the CT50 Single OLÉ is a different construction. The refrigerant is R-290 (A3, H220) with a first-stage charge of only 0.06 kg.
Behavior under load
Cooling capacity is temperature-dependent and must be sized at the actual process temperature: 0.3 kW at 0 °C, 0.26 kW at -20 °C and 0.2 kW at -30 °C. The working range is -50 to 50 °C, so the unit is for cooling only and cannot heat above 50 °C. Temperature stability is 0.5 K at -10 °C. The refrigeration unit runs continuously and control is performed by a solenoid valve in the cooling circuit, so the compressor is not cycled for temperature control.
Control, installation and operation
Control is by an MPC controller with LED display and a Pt100 external sensor connection; alarms are optical and acoustic. The unit requires a 208-240 V, 1~ or 2~, 50/60 Hz supply with a minimum 10 A fuse and a maximum current of 3 A. Single- and two-phase units are delivered with a power cable and a country-specific plug that must be specified at ordering. Ambient temperature must remain between 5 and 40 °C, and cooling capacity may drop as ambient rises.
Integration & Operating Notes
Single- and two-phase devices in the 100–240 V range are delivered with a power cable and a country-specific plug, which must be stated when ordering. Confirm the plug type and the 10 A minimum fuse rating of the site circuit before delivery to avoid rework at installation.
Capacity falls from 0.3 kW at 0 °C to 0.26 kW at -20 °C and 0.2 kW at -30 °C. A load specified at a warmer temperature will be undersized if the process actually runs colder, so the heat load must be evaluated at the lowest intended bath temperature.
The operating range ends at 50 °C, so the unit cannot serve processes that require heating above this value. It is a cooling device for low-temperature tasks; any process step above 50 °C must be handled by separate equipment.
R-290 is classified A3 (H220), so installation requires attention to ventilation and ignition-source control in the vicinity of the refrigeration circuit. The charge is only 0.06 kg, which keeps F-gas reporting minimal, but the flammability classification still governs site zoning and service procedures.
The unit is air-cooled, so its heat rejection depends on room conditions. Below 5 °C or above 40 °C ambient the unit is outside its permitted environment, and as ambient rises the available cooling capacity may drop, which must be accounted for in the heat balance.
The MPC controller accepts an external Pt100 sensor and displays temperature on an LED display, with optical and acoustic alarms. This allows control at the process measurement point rather than at the unit, but the sensor and its cabling are not part of the base delivery and must be provided separately.
The unit is classified Class A, Group 1 to EN55011, so it is intended for industrial environments and may not be suitable for residential or light-commercial locations without additional EMC measures.
Selection & Integration Q&A
Choose the TC50E when the heat load fits its capacity curve (0.3 kW at 0 °C down to 0.2 kW at -30 °C) and the process stays within -50 to 50 °C. The TC100/TC100E siblings are larger units for higher loads; the TC50E is the smallest single-stage immersion cooler in the series.
A 208-240 V, 1~ or 2~, 50/60 Hz supply with a minimum 10 A fuse and maximum current of 3 A. Voltage tolerance is ±5% only with ±2% frequency, and -5% voltage with +2% frequency is not allowed. Specify the country-specific plug at ordering.
Ambient must stay between 5 and 40 °C; above that, cooling capacity may drop. The probe is 48 mm diameter and 155 mm long with a 1150 mm flexible line, so the vessel port and routing must accommodate these. R-290 (A3) requires ventilation and ignition-source control.
The external Pt100 sensor and its cabling for the sensor connection are not part of the base unit and must be supplied separately. The power cable and country-specific plug are included only if specified at ordering. No pump is fitted, so bath mixing must be provided by other means.