Peter Huber Kältemaschinenbau Unimotive 26w-XT
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unimotive 26w-XT

Peter Huber Kältemaschinenbau Unimotive 26w-XT — Temperature control units.

Water-cooled thermostatic bath/circulator with magnetically coupled stainless-steel pump and 24 kW heating, covering -45 to 150 °C on a 400 V 3~ 50 Hz supply. The XT modification of the Unimotive 26w adds the extended low-temperature range down to -45 °C, distinguishing it from the 10w and 20w siblings. Intended for temperature control of external circuits and bath applications with water-glycol heat transfer fluid.

Peter Huber Kältemaschinenbau Unimotive Series — Model Comparison & Selection

Model Display resolutionHeating capacity, WPower supplyPump connectionRefrigerantSafety class
Unimotive 26w-XT Current 0, 0124000400V 3~ 50HzM38x1, 5 AG maleR-452A (A1, H280)I / NFL
Unimotive 10w 0.0112000400V 3~ 50HzM38x1,5 AG maleR-452A (A1, H280)III / FL
Unimotive 10w-XT 0, 0112000400V 3~ 50HzM38x1, 5 AG maleR-452A (A1, H280)I / NFL
Unimotive 20w 0, 0112000400V 3~ 50HzM38x1, 5 AG maleR-452A (A1, H280)III / FL
Unimotive 20w-XT 0, 0112000400V 3~ 50HzM38x1, 5 AG maleR-452A (A1, H280)I / NFL
Unimotive 26w 0, 01 K24000400V 3~ 50HzM38x1, 5 AG maleR-452A (A1, H280)III / FL
Unimotive 27w 0, 0124000400V 3~ 50HzM38x1, 5 AG maleR-452A (A1, H280)III / FL
Unimotive GL 07w 0, 0112000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)III / FL
Unimotive GL 07w-XT 0, 0112000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)I / NFL
Unimotive GL 10w 0, 0124000460V 3~ 60HzM38x1, 5 AG maleR-744 (A1, H280)III / FL
Unimotive GL 10w-XT 0, 0124000460V 3~ 60HzM38x1, 5 AG maleR-744 (A1, H280)I / NFL
Unimotive GL 20w 0, 0124000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)III / FL
Unimotive GL 20w-XT 0, 0124000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)I / NFL
Unimotive GL 30w 0, 0124000460V 3~ 60HzM38x1, 5 AG maleR-744 (A1, H280)III / FL
Unimotive GL 50w 0, 0124000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)III / FL
Unimotive GL 50w-XT 0, 0124000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)I / NFL
Unimotive GL 60w 0, 0148000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)III / FL
Unimotive GL 60w-XT 0, 0148000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)I / NFL

Full specifications

Core Performance & Functional Specifications
Control output POKO ONE (potential-free contact)
Heating capacity 24 kW
Pump connection M38x1, 5 AG male
Pump speed adjustable no
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -45...150 °C
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Refrigerant R-452A (A1, H280) (1st stage)
Maximum viscosity 200 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G3/4 male
Cooling water consumption 1260 l/h (at 15°C, flow 0°C)
Cooling water differential pressure 1 bar (min.)
Cooling Water Pressure 6 bar (max.)
Protection rating IP20
Standards, Safety & General Data
Safety class I / NFL

Description

Construction and series positioning

Behavior under load and media

Installation, control and operation

Integration & Operating Notes

Electrical
Supply tolerance pairing is restricted

Voltage may deviate ±5% only when frequency deviation is simultaneously within ±2%; the combination of -5% voltage with +2% frequency is explicitly not allowed. Verify the site supply before connection, because an out-of-tolerance combination can damage the compressor and electronics.

Unimotive 26w-XT Thermostatic Bath/Circulator Datasheet, Technical Specifications
Electrical
Provide a plug and PE socket on site

Three-phase devices below 63 A are delivered with a cable but without a plug, so the installer must provide a suitable plug. The unit must always be connected to a socket with protective earth; hardwiring requires a certified electrician. Leakage current exceeds 3.5 mA, which affects RCD selection.

Unimotive 26w-XT Thermostatic Bath/Circulator Datasheet, Technical Specifications; Unimotive Operation Manual, Electrical Connection
Cooling & Media
Cooling water quality and temperature window

Cooling water intake must be between 5 °C and 20 °C, with differential pressure between 1 bar and 4 bar and maximum pressure 6 bar. Only drinking water or filtered river/sea water is permitted; sea water must not be used. Circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so the water must be compatible with these.

Unimotive 26w-XT Thermostatic Bath/Circulator Datasheet, Technical Specifications; Unimotive Operation Manual, Installation
Cooling & Media
Shut off cooling water during downtimes

Cooling water can flow even when the unit is switched off, so the supply should be shut off for shorter downtimes such as overnight. A building-side backflow preventer is required to protect the drinking water supply. For outdoor siting, cooling water lines must be frost-protected and the water temperature must not fall below 5 °C.

Unimotive Operation Manual, Installation
Mounting
Respect wall clearances and refrigerant room size

Minimum wall clearances depend on air- or water-cooling and on whether the unit runs with or without a bath; insufficient clearance prevents energy dissipation and risks overheating. The refrigerant charge and room size must be checked during installation to avoid dangerous refrigerant concentration. A drip tray for condensate/heat transfer fluid must be installed below the unit by the responsible body.

Unimotive Operation Manual, Installation
Process Limits
Pressure-sensitive circuits may need a bypass

The maximum flow pressure of the unit must be considered for pressure-sensitive applications such as glass reactors; an external bypass may be required. Hose inside diameters must at least match the pump connections, with larger diameters for longer lines. Before switching off, the heat transfer fluid must be brought to room temperature to avoid overpressure.

Unimotive Operation Manual, Normal Operation
Environment
Not for ATEX zones; industrial EMC only

The unit must not be installed or started inside an ATEX zone. It is a Group 1 Class A device to EN55011, suited to industrial EMC environments; installation elsewhere may cause interference. Ambient temperature must remain between 5 °C and 40 °C, and IP20 requires a dry indoor environment.

Unimotive Operation Manual, Installation; Unimotive 26w-XT Thermostatic Bath/Circulator Datasheet, Technical Specifications

Selection & Integration Q&A

When should I choose the Unimotive 26w-XT rather than a 10w or 20w sibling?

Pick the 26w-XT when the process needs the -45...150 °C range with 24 kW heating and the higher cooling capacity (28 kW at 20 °C, 2.6 kW at -40 °C). The 10w/20w variants are smaller; confirm their own ranges and capacities before substituting.

What must be provided on the cooling water and electrical side before commissioning?

Cooling water at 5...20 °C with 1...4 bar differential pressure, max 6 bar, G3/4 male connection, and a building-side backflow preventer. Electrically: 400 V 3~ 50 Hz with a PE socket and a site-supplied plug, since three-phase units below 63 A ship with cable but no plug.

Where can the unit be installed and what duty-cycle limits apply?

Dry indoor location, ambient 5...40 °C, no ATEX zone, industrial EMC environment (EN55011 Class A Group 1). Respect wall clearances per the manual table and check refrigerant charge against room size. Shut off cooling water during downtimes; bring fluid to room temperature before switching off.

What is not included in base delivery and what maintenance/consumables should be planned?

The mains plug is not included for three-phase units below 63 A; a drip tray and backflow preventer are site responsibilities. Heat transfer fluid must be a water-ethylene glycol mixture with corrosion protection (e.g. Glysantin®) down to -45 °C. Replacement power cables are limited to 4 m for three-phase connections.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unimotive 26w-XT Thermostatic Bath/Circulator Datasheet
Datasheet ·84,5 КБ ·2 pp.
View PDF
PDF
Unimotive Operation Manual
Manual ·2,8 МБ ·82 pp.
View PDF