Peter Huber Kältemaschinenbau Unistat 430w Dynamisches
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unistat 430w Dynamisches

Peter Huber Kältemaschinenbau Unistat 430w Dynamisches — Temperature control units with Noise level 61 dB.

The Unistat 430w Dynamisches is a water-cooled, free-standing temperature control unit with a magnetically coupled stainless-steel pump and Pilot ONE controller, covering -40 to 250 °C with 4 kW heating. It is the dynamic (high-flow) modification of the 4xx series, delivering up to 91 l/min and 1.5 bar for externally closed circuits with high pressure drop. It serves reactor and process tempering where fast, high-flow heat transfer is required.

Noise level
61 dB

Peter Huber Kältemaschinenbau Unistate Bis 45C Series — Model Comparison & Selection

Model Noise level, dBDisplay resolutionPower supplyPump connectionPump typeTemperature range, °C
Unistat 430w Dynamisches Current 610, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250
Grande Fleur Dynamisches ——————
Grande Fleur w Dynamisches —0, 01208-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable -40–200
Grande Fleur w-eo Dynamisches ——————
Petite Fleur-eo Dynamisches 580, 01 K220-240V 1~/2~ 50/60HzM16x1 AG malePressure pump, variable -40–200
Petite Fleur w Dynamisches —0, 01220-240V 1~/2~ 50/60HzM16x1 AG malePressure pump, variable -40–200
Unistat 410w Dynamisches 610, 01 K220-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable -45–250
Unistat 425w Dynamisches 610, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250
Unistat P405w Dynamisches —0, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st-45–250
Unistat P425 Dynamisches 650, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250
Unistat P425w Dynamisches 610, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250
Unistat P430 Dynamisches 650, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250
Unistat P430w Dynamisches 650, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250

Full specifications

Core Performance & Functional Specifications
Noise level 61 dB(A)
Control output POKO ONE
Filling volume 4 l
Heating capacity 4 kW
Pump connection M30x1, 5 AG male
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -40...250 °C
Dimensions, Weight & Mechanical Design
Weight 175 kg
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G1/2 male
Cooling water consumption 312 l/h (at 15°C, flow 0°C)
Cooling water differential pressure 1 bar (min.)
Cooling Water Pressure 6 bar (max.)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Thermal and hydraulic behavior

Installation, control and operation

Integration & Operating Notes

Electrical
Size the supply and RCD for >3.5 mA leakage current

The 400 V 3~ 50 Hz supply needs a 3x16 A fuse, and the unit's leakage current exceeds 3.5 mA, so the residual current device type and rating must be chosen to avoid nuisance tripping while maintaining protection. A plug must be provided on site because three-phase units below 63 A ship with cable only.

Unistat 430w Technical Data, page 2
Cooling & Media
Keep cooling water between 1 and 6 bar differential

The water-cooled condenser requires at least 1 bar differential pressure to achieve the specified 312 l/h at 15 °C intake, and must never see more than 6 bar. Intake temperature must stay within 5…20 °C; insulate piping where condensation forms and use only cooling water compatible with the copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM wetted materials.

Unistat 430w Technical Data, page 2
Process Limits
Do not install in an ATEX zone or for medical/food use

Without an Ex px cabinet the unit is not explosion-protected, so site selection must exclude ATEX zones. Use with medical devices (e.g. in vitro diagnostics) or for direct foodstuff temperature control is not permissible; FL-rated fluids require a fire point of at least 65 °C and supervised operation if ethanol is used.

Unistat Operation Manual, Section 1.3.3 / 1.3.4
Mounting
Transport upright, empty, and let it acclimate before power-on

Horizontal transport damages the compressor, and only an emptied unit may be moved. Casters are not for transport and carry only 25% of total mass symmetrically; use an industrial truck. After moving between cold and warm environments allow about 2 hours acclimation before switching on to prevent condensation-induced electrical faults.

Unistat Operation Manual, Commissioning, page 22 / 29
Control & I/O
Decide restart behavior and Process Safety compressor mode

Pilot ONE can be set to Off/Standby (default, manual start after power-up) or ON/Temperature control active (automatic restart with a brief INFO window to abort). Process Safety mode releases maximum cooling at the OT trigger and requires the compressor set to Permanently ON; choose this only where unattended restart and continuous cooling are acceptable for the process.

Unistat Operation Manual, Section 1.7.2 / 1.8.1
Maintenance
Shut off cooling water overnight and inspect hoses periodically

Close the cooling water supply even for short downtimes such as overnight to avoid water damage and unnecessary load on the circuit. Periodically check hoses and hose connections for leaks and quality and replace as required; work on the refrigerant circuit must be performed by approved refrigeration/air-conditioning contractors, not general maintenance staff.

Unistat Operation Manual, page 29 / Section 1.4.2

Selection & Integration Q&A

When should I choose the Unistat 430w Dynamisches instead of the Grande Fleur variants in the same series?

Pick the 430w Dynamisches for high-flow, high-pressure-drop externally closed circuits: its pump delivers up to 91 l/min and 1.5 bar. The Grande Fleur models are separate configurations; compare their pump curves and cooling capacities against your circuit's pressure drop before deciding.

What must I provide on the cooling water and process side to connect this unit?

Provide cooling water at 5…20 °C with 1 bar minimum differential pressure and no more than 6 bar, using G1/2 male connections. Process hoses need M30x1.5 male mating ball-socket hoses and sleeve nuts per DIN 3863; the 400 V supply needs a 3x16 A fuse and an on-site plug.

What maintenance and accessories are not included in the base delivery?

The three-phase cable ships without a plug, so a plug must be provided on site. A condensation/fluid drip tray must be installed by the responsible body. Periodic hose inspection and replacement, cooling water shutoff, and refrigerant-circuit work by approved contractors are owner responsibilities.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unistat 430w Technical Data
Datasheet ·77,1 КБ ·2 pp.
View PDF
PDF
Unistat® Operation Manual — Unistat® 5, Unistat® Pilot ONE®
Manual ·3,6 МБ ·100 pp.
View PDF
PDF
Unistat Operation Manual
Manual ·3,7 МБ ·102 pp.
View PDF