Peter Huber Kältemaschinenbau Unistat 430w Dynamisches
Peter Huber Kältemaschinenbau Unistat 430w Dynamisches — Temperature control units with Noise level 61 dB.
The Unistat 430w Dynamisches is a water-cooled, free-standing temperature control unit with a magnetically coupled stainless-steel pump and Pilot ONE controller, covering -40 to 250 °C with 4 kW heating. It is the dynamic (high-flow) modification of the 4xx series, delivering up to 91 l/min and 1.5 bar for externally closed circuits with high pressure drop. It serves reactor and process tempering where fast, high-flow heat transfer is required.
Peter Huber Kältemaschinenbau Unistate Bis 45C Series — Model Comparison & Selection
| Model | Noise level, dB | Display resolution | Power supply | Pump connection | Pump type | Temperature range, °C |
|---|---|---|---|---|---|---|
| Unistat 430w Dynamisches Current | 61 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
| Grande Fleur Dynamisches | — | — | — | — | — | — |
| Grande Fleur w Dynamisches | — | 0, 01 | 208-240V 1~/2~ 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | -40–200 |
| Grande Fleur w-eo Dynamisches | — | — | — | — | — | — |
| Petite Fleur-eo Dynamisches | 58 | 0, 01 K | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | Pressure pump, variable | -40–200 |
| Petite Fleur w Dynamisches | — | 0, 01 | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | Pressure pump, variable | -40–200 |
| Unistat 410w Dynamisches | 61 | 0, 01 K | 220-240V 1~/2~ 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | -45–250 |
| Unistat 425w Dynamisches | 61 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
| Unistat P405w Dynamisches | — | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st | -45–250 |
| Unistat P425 Dynamisches | 65 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
| Unistat P425w Dynamisches | 61 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
| Unistat P430 Dynamisches | 65 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
| Unistat P430w Dynamisches | 65 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
Full specifications
| Noise level | 61 dB(A) |
| Control output | POKO ONE |
| Filling volume | 4 l |
| Heating capacity | 4 kW |
| Pump connection | M30x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -40...250 °C |
| Weight | 175 kg |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1/2 male |
| Cooling water consumption | 312 l/h (at 15°C, flow 0°C) |
| Cooling water differential pressure | 1 bar (min.) |
| Cooling Water Pressure | 6 bar (max.) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Free-standing 4xx-class unit, water-cooled condenser (suffix 'w'), dynamic pump variant; distinguishes from air-cooled and CO2 (GL) siblings and from the Grande Fleur models in the same series listing.
- Magnetically coupled stainless-steel circulation pump, speed-adjustable, with M30x1.5 male pump connection (DIN 3863 Y bore).
- Pilot ONE controller with POKO ONE control output; 0.01 K display resolution; Pt100 external sensor input.
- 4 l internal filling volume; 175 kg; IP20; safety class III / FL.
Thermal and hydraulic behavior
- Working range -40…250 °C; heating capacity 4 kW.
- Cooling capacity is strongly temperature-dependent: 3.5 kW at 0 °C falling to 0.3 kW at -40 °C with ethanol — low-temperature duty must be sized from the curve, not the nominal figure.
- Pump delivers max. 91 l/min at max. 1.5 bar; flow falls as external circuit pressure rises, so match the pump curve to the application pressure drop.
- Cooling water circuit: G1/2 male connections, 312 l/h consumption at 15 °C intake / 0 °C flow, min. 1 bar differential pressure, max. 6 bar.
Installation, control and operation
- 400 V 3~ 50 Hz supply with 3x16 A fuse; delivered with cable but without plug (three-phase <63 A); leakage current >3.5 mA affects RCD selection.
- Cooling water intake 5…20 °C; insulate pipes where condensation occurs; shut off cooling water during downtimes.
- Interfaces: Ethernet, USB host and device, RS232; firmware update via interface box.
- Ambient 5…40 °C, indoor dry location only; not for ATEX zones, medical devices or direct food contact.
Integration & Operating Notes
The 400 V 3~ 50 Hz supply needs a 3x16 A fuse, and the unit's leakage current exceeds 3.5 mA, so the residual current device type and rating must be chosen to avoid nuisance tripping while maintaining protection. A plug must be provided on site because three-phase units below 63 A ship with cable only.
The water-cooled condenser requires at least 1 bar differential pressure to achieve the specified 312 l/h at 15 °C intake, and must never see more than 6 bar. Intake temperature must stay within 5…20 °C; insulate piping where condensation forms and use only cooling water compatible with the copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM wetted materials.
Without an Ex px cabinet the unit is not explosion-protected, so site selection must exclude ATEX zones. Use with medical devices (e.g. in vitro diagnostics) or for direct foodstuff temperature control is not permissible; FL-rated fluids require a fire point of at least 65 °C and supervised operation if ethanol is used.
Horizontal transport damages the compressor, and only an emptied unit may be moved. Casters are not for transport and carry only 25% of total mass symmetrically; use an industrial truck. After moving between cold and warm environments allow about 2 hours acclimation before switching on to prevent condensation-induced electrical faults.
Pilot ONE can be set to Off/Standby (default, manual start after power-up) or ON/Temperature control active (automatic restart with a brief INFO window to abort). Process Safety mode releases maximum cooling at the OT trigger and requires the compressor set to Permanently ON; choose this only where unattended restart and continuous cooling are acceptable for the process.
Close the cooling water supply even for short downtimes such as overnight to avoid water damage and unnecessary load on the circuit. Periodically check hoses and hose connections for leaks and quality and replace as required; work on the refrigerant circuit must be performed by approved refrigeration/air-conditioning contractors, not general maintenance staff.
Selection & Integration Q&A
Pick the 430w Dynamisches for high-flow, high-pressure-drop externally closed circuits: its pump delivers up to 91 l/min and 1.5 bar. The Grande Fleur models are separate configurations; compare their pump curves and cooling capacities against your circuit's pressure drop before deciding.
Provide cooling water at 5…20 °C with 1 bar minimum differential pressure and no more than 6 bar, using G1/2 male connections. Process hoses need M30x1.5 male mating ball-socket hoses and sleeve nuts per DIN 3863; the 400 V supply needs a 3x16 A fuse and an on-site plug.
The three-phase cable ships without a plug, so a plug must be provided on site. A condensation/fluid drip tray must be installed by the responsible body. Periodic hose inspection and replacement, cooling water shutoff, and refrigerant-circuit work by approved contractors are owner responsibilities.