Peter Huber Kältemaschinenbau Huber LCS 80
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Huber LCS 80

Peter Huber Kältemaschinenbau Huber LCS 80 — Recirculating chillers with Noise level 77 dB.

The Huber LCS 80 is a 460 V 3-phase 60 Hz recirculating chiller with a -45 to 15 °C operating range, 20 kW cooling at 15 °C falling to 8.5 kW at -40 °C, and a 53 l/min G1 pump. It is the low-temperature, high-capacity member of the Cs Chiller series, intended for process cooling where the duty point sits well below ambient.

Noise level
77 dB

Peter Huber Kältemaschinenbau Cs Chiller Series — Model Comparison & Selection

Model Noise level, dBAmbient temperature limit, °CMinimum fill volume, m³Power supplyPump connectionTemperature range, °C
Huber LCS 80 Current 77-20–500.04460V 3~ 60HzG1 AG male-45–15
CS 100 by Van der Heijden 705–320.16400V 3~ 50HzG3/4 AG male-20–15
CS 100-H by Van der Heijden 715–320.175400V 3~ 50HzG3/4 AG male-20–85
CS 200 by Van der Heijden 785–320.225400V 3~ 50HzG1 AG male-20–15
CS 200-H by Van der Heijden 785–320.225400V 3~ 50HzG1 AG male-20–85
CS 20-H by Van der Heijden 575–320.024230V 1~ 50HzG3/8 AG male-20–85
CS 25 by Van der Heijden 605–320.024230V 1~ 50HzG1/2 AG male-20–15
CS 50 by Van der Heijden 635–320.1400V 3~ 50HzG3/4 AG male-20–15
CS 75 by Van der Heijden 665–320.16400V 3~ 50HzG3/4 AG male-20–15
Huber CS 200-H 785–320.225460V 3~ 60HzG1 AG male-20–85
Huber CS 25 605–320.024208V 2~ 60HzG1/2 AG male-20–15
Huber CS 250HT OD 78-20–500.16460V 3~ 60HzG3/4 AG male10–22
Huber CS 330-H OD 78-20–500.16460V 3~ 60HzG1 AG male-20–85
Huber CS 550-H OD 79-20–500.16460V 3~ 60HzG1 AG male-20–85
Huber CS 550 OD 79-20–500.16460V 3~ 60HzG1 AG male-20–15

Full specifications

Core Performance & Functional Specifications
Noise level 77 dB(A) (sound pressure level)
Bath volume 37 l
Maximum delivery flow rate 53 l/min (max.; pressure 4, 7 bar)
Minimum fill volume 40 l (min.)
Pump connection G1 AG male
Temperature range -45...15 °C
Dimensions, Weight & Mechanical Design
Weight 880 kg (net)
Electrical, Power & Drive Specifications
Current consumption 50 A (max.)
Power supply 460V 3~ 60Hz
Interfaces, Connections & Control
Digital interface RS232
Materials, Media & Process Compatibility
Refrigerant R-449A (A1, H280) (1st stage)
Refrigerant charge 16 kg (1st stage)
Operating Conditions, Environment & Protection
Ambient temperature limit -20...50 °C (permissible)
Standards, Safety & General Data
Safety class I / NFL

Description

Construction and series positioning

Behavior under load and media

Installation, control and operation

Integration & Operating Notes

Electrical
Supply quality must stay within ±5% voltage and ±2% frequency

The unit tolerates ±5% voltage only with a simultaneous ±2% frequency tolerance; the combination of -5% voltage with +2% frequency is explicitly not allowed. Verify site power quality before commissioning, because out-of-tolerance supply can damage the compressor and void compliance.

Huber LCS 80 Technical Data, page 2
Electrical
Cable supplied, plug not included below 63 A

Three-phase devices with current consumption less than 63 A are delivered with cable but without a plug. The LCS 80 draws 50 A max and requires 3x60 A fusing, so the customer must provide and fit the site connector and size the protection device accordingly.

Huber LCS 80 Technical Data, page 2
Process Limits
Fill volume must stay between 40 l and 105 l

The bath volume ranges from 40 l minimum to 105 l maximum, with a warning at 75 l. Operating outside these limits risks insufficient heat transfer or overflow; the fill volume must be checked against the external circuit volume at commissioning.

Huber LCS 80 Technical Data, page 1
Process Limits
Use Temper -55 fluid for the -45 °C lower limit

The recommended heat transfer fluid is Temper -55, selected to match the low-temperature duty down to -45 °C. A fluid with a higher freeze point will degrade or freeze at the lower end of the operating range.

Huber LCS 80 Technical Data, page 2
Mounting
Outdoor siting allowed down to -20 °C ambient

The unit is suitable for outside mounting with a minimum environmental temperature of -20 °C and a maximum of +50 °C. Outdoor siting widens placement options but requires weather protection planning and verification of cooling capacity at the site ambient, since capacity is reduced above +35 °C.

Huber LCS 80 Technical Data, page 1
Control & I/O
Remote panel on a fixed 20 m cable

The operating panel is supplied as a remote control connected via a 20 m cable, so the operator station must be located within that cable length. An RS232 interface is available for integration with a supervisory controller.

Huber LCS 80 Technical Data, page 1
Process Limits
External circuit must match G1 male and 4.7 bar

The pump connection is G1 AG male with a maximum delivery flow of 53 l/min and a maximum delivery pressure of 4.7 bar. Hoses, fittings and the external circuit must be rated for this pressure and match the G1 thread to avoid leaks or over-pressurisation.

Huber LCS 80 Technical Data, page 1
Maintenance
Refrigerant record-keeping applies (22.35 t CO2e)

The refrigerant is R-449A (A1, H280) with a 16 kg charge and 22.35 t CO2 equivalent. F-gas/CO2-equivalent record-keeping and qualified refrigerant handling obligations apply, including leak reporting.

Huber LCS 80 Technical Data, page 1

Selection & Integration Q&A

When should I choose the LCS 80 rather than a CS 100 or CS 125 sibling?

Pick the LCS 80 when the process needs temperatures down to -45 °C and up to 20 kW cooling at 15 °C. The CS 100/CS 125 siblings are quoted only by noise level (70–75 dB(A)) and are not specified for this low-temperature window, so they suit less demanding duty points.

What site electrical infrastructure does the LCS 80 require?

A 460 V 3-phase 60 Hz supply with 50 A max current and 3x60 A fusing. The unit is delivered with cable but without a plug (below 63 A), so the customer must provide the connector. Supply must stay within ±5% voltage and ±2% frequency; -5% voltage with +2% frequency is not allowed.

Can the LCS 80 be installed outdoors, and what limits apply?

Yes, outdoor mounting is permitted down to -20 °C ambient and up to +50 °C. However, cooling capacity is reduced above +35 °C ambient, so the site ambient must be checked against the capacity tables, and weather protection should be planned.

What is included in base delivery and what must be sourced separately?

Base delivery includes the unit with cable but no plug, a remote operating panel on a 20 m cable, and an RS232 interface. The site connector, external circuit hoses/fittings matching G1 male and 4.7 bar, and the Temper -55 heat transfer fluid must be sourced separately.