Agilent Technologies Flexible HPLC/UHPLC, Prime Analytical-Scale Purification System
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies Flexible HPLC/UHPLC, Prime Analytical-Scale Purification System

Agilent Technologies Flexible HPLC/UHPLC, Prime Analytical-Scale Purification System — Liquid chromatography.

The Agilent InfinityLab analytical-scale purification system is a flexible HPLC/UHPLC platform configured for 1–3 inch id Load & Lock columns in DAC or SAC mode. It delivers 40,606 N/m plate count at 1 mL/min with 15.18 min retention and 1.27 asymmetry, distinguishing it from fixed-configuration siblings. It serves as the purification bridge between analytical method development and preparative-scale isolation.

Agilent Technologies Preparative HPLC Systems Series — Model Comparison & Selection

Model Column inner diameter, mmTypical flow rate, mL/minPlate count, Nth/meterRetention time, minCarryover, %Sample capacity
Flexible HPLC/UHPLC, Prime Analytical-Scale Purification System Current 1...314060615.18——
1220 Infinity II Analytical-Scale LC Purification System 4.61————
1260 Analytical-Scale Purification System | Preparative HPLC 1...314060615.18——
1260 Infinity Bio-Inert Analytical-Scale LC Purification System 1...314060615.18——
1260 Infinity II Manual Preparative LC System 1...314060615.18——
1260 Infinity II Preparative LC System | UV based Prep HPLC 1...314060615.185132
1260 Prime Bio Analytical-Scale Purification System 4.61——9×10⁻⁴6144
1290 Analytical-Scale LC Purification 1...314060615.18——
1290 Infinity Autoscale Preparative system 1...314060615.18——
1290 Infinity III 4.61——9×10⁻⁴6144
1290 Infinity II Preparative LC System 1...314060615.18——
Automated LC/MSD purification| High-precision| Complex mixtures 1...314060615.18——
High-throughput Purification | Drug discovery | Biological 1...314060615.18——
Preparative LC/MSD System | Enhanced Purification & Efficiency 1...314060615.18——

Full specifications

Core Performance & Functional Specifications
Plate Count 40606 Nth/meter
Retention Time 15.18 min
Typical Flow Rate 1 mL/min
Productivity range mfr Micrograms Micrograms
Asymmetry mfr 1.27
Dimensions, Weight & Mechanical Design
Column Inner Diameter 1...3 in

Description

Construction and series positioning

Behavior under load

Installation, connection and control

Integration & Operating Notes

Process Limits
Column id must stay within 1–3 inch envelope

The system is specified for Load & Lock columns from 1 to 3 inches id (27 to 75 mm). Fitting a column outside this range invalidates the plate-count and asymmetry performance and may not mate with the packing station or column organizer.

Agilent Load & Lock Columns brochure 5994-3907EN
Process Limits
DAC and SAC modes require different packing procedures

Dynamic axial compression and static axial compression are distinct operating modes for the same column hardware. The packing station and procedure must be selected for the intended mode before the column is installed; switching modes after packing is not described in the verified documentation.

Agilent Load & Lock Columns brochure 5994-3907EN
Cooling & Media
Water-jacketed columns need a temperature-control supply

Standard stainless steel water jackets are part of the column construction. A temperature-control fluid circuit must be available at the installation site; the verified facts do not state jacket pressure, flow or temperature ratings, so these must be confirmed against the column datasheet.

Agilent Load & Lock Columns brochure 5994-3907EN
Maintenance
Packing station must match the column id

Mobile packing stations are offered separately for 1, 2 and 3 inch id columns. Using a station sized for a different id will not correctly pack the column; confirm the station model against the column id before ordering or scheduling packing.

Agilent Load & Lock Columns brochure 5994-3907EN
Control & I/O
Module set determines available detection and fractionation

The portfolio includes column organizer, column compartment, manual injector, open-bed sampler/collector, detectors, fraction collectors and LC/MSD XT. The exact modules delivered with this configuration are not itemized in the verified facts; confirm the module list against the order configuration before designing the workflow.

InfinityLab LC Purification Solutions brochure 5991-9154EN

Selection & Integration Q&A

When should I choose this analytical-scale purification system instead of the 1260 Infinity II Manual Preparative LC System or the 1220 Infinity II Analytical-Scale system?

Choose this system when the workflow needs flexible HPLC/UHPLC operation with 1–3 inch id Load & Lock columns in DAC or SAC mode at analytical-scale flow (1 mL/min typical). The 1220 Infinity II is the lower-tier analytical-scale option; the 1260 Infinity II Manual Preparative system targets manual preparative operation rather than this flexible prime configuration.

What adjacent equipment must be specified alongside the system for column packing and detection?

A mobile packing station sized to the column id (1, 2 or 3 inch) is required for DAC or SAC packing, plus a temperature-control fluid supply for the stainless steel water jackets. Detection and fractionation depend on the selected module set (detectors, fraction collectors, LC/MSD XT); confirm the module list against the order configuration.

What are the mounting and duty-cycle limits for the column and packing station?

Columns must be within the 1–3 inch id range and matched to the correct packing station for their id. DAC and SAC modes use different packing procedures, so the mode must be fixed before packing. The verified facts give no jacket pressure, flow or temperature ratings; confirm these against the column datasheet before installation.

What maintenance items and accessories are not included in the base delivery?

Mobile packing stations for 1, 2 and 3 inch id columns are separate items, as are the preparative columns and supplies listed in the portfolio brochure. Detectors, fraction collectors and LC/MSD XT are modules whose inclusion depends on configuration. The verified facts contain no service-interval schedule; maintenance intervals must be taken from the instrument manual.

Dimensional Drawings & Technical Documentation