Peter Huber Kältemaschinenbau TC100
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau TC100

Peter Huber Kältemaschinenbau TC100 — Immersion coolers with Noise level 55 dB.

The TC100 is a two-stage cascade immersion cooler with a 48 mm × 155 mm probe on a 1150 mm flexible connection, covering -100 to +40 °C. Within the TC100 series it is the base model without the E-variant electronics; the -F variants differ in noise. It serves as a direct-contact cold trap or bath cooler in laboratory and pilot vacuum/process lines.

Noise level
55 dB

Peter Huber Kältemaschinenbau Eintauchkuehler Kuehlfallen Series — Model Comparison & Selection

Model Noise level, dBControllerPower supplyRefrigerantSafety classTemperature range, °C
TC100 Current 55ON-OFF-Regler220-240V 1~/2~ 50/60HzR-1150 (A3, H220)I / NFL-100–40
CT50 Single OLÉ 59OLÉ220-240V 1~/2~ 50/60HzR-290 (A3, H220)III / FL-50–50
TC100E-F 55MPC110-120V 1~ 60HzR-1150 (A3, H220)I / NFL-100–40
TC100-F 58ON-OFF-Regler220-240V 1~/2~ 50/60HzR-1150 (A3, H220)I / NFL-100–40
TC45 —ON-OFF-Regler208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-45–100
TC45E —MPC208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-45–100
TC45E-F 54MPC208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-45–100
TC45-F 59ON-OFF-Regler208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-45–100
TC50 —ON-OFF-Regler208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-50–50
TC50E —MPC208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-50–50
TC50E-F 60MPC208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-50–50
TC50-F 60ON-OFF-Regler208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-50–50

Full specifications

Core Performance & Functional Specifications
Noise level 55 dB(A)
Pump speed adjustable no
Pump type no pump
Temperature range -100...40 °C
Dimensions (W x D x H) mfr 295 x 500 x 570 mm
Refrigeration machine mfr air-cooled, CFC- and HCFC-free
Dimensions, Weight & Mechanical Design
Weight 61 kg
Electrical, Power & Drive Specifications
Power supply 220-240V 1~/2~ 50/60Hz
Current consumption max. mfr 6, 5 A (max)
Interfaces, Connections & Control
Controller ON-OFF-controller
Materials, Media & Process Compatibility
Refrigerant R-452A (A1, H280) (1st stage)
2nd stage: R-1150 (A3, H220)
Refrigerant quantity 1st stage mfr 0, 26 kg (1st stage)
Refrigerant quantity 2nd stage mfr 0, 056 kg (2nd stage)
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling capacity mfr 0, 16 at 0 °C to 0, 01 at -100 °C kW (at °C bath temperature)
Standards, Safety & General Data
Safety class I / NFL

Description

Construction and series position

The TC100 is an air-cooled, CFC/HCFC-free cascade refrigeration unit built as a separate chassis (295 × 500 × 570 mm, 61 kg) connected by a 1150 mm flexible line to a 48 mm nominal diameter, 155 mm long immersion probe. Cooling is provided by a two-stage circuit: R-452A (A1, H280) in the first stage and R-1150 (A3, H220) in the second. Within the series, the TC100 is the base configuration; the TC100E and TC100E-F add electronics, and the TC100-F is the -F variant at 58 dB(A) versus 55 dB(A) for this model. The CT50 Single OLÉ is a smaller sibling at 59 dB(A).

Thermal behavior

Cooling capacity is quoted against bath temperature: 0.16 kW at 0 °C falling to 0.01 kW at -100 °C, so the unit is intended for low-load, low-temperature duty rather than bulk heat removal. Capacity also degrades as ambient temperature rises above the reference, so sizing must include margin for rooms warmer than 20 °C. Control is ON-OFF (no proportional or pump-assisted circulation; no pump is fitted).

Installation and operation

The unit is air-cooled and requires the specified clearance around the condenser for airflow and service. Electrical supply is 220–240 V, 1~ or 2~, 50/60 Hz, with a maximum current of 6.5 A; protection must be between 10 A and 16 A. Voltage and frequency tolerances are coupled (±5 % voltage only with ±2 % frequency, and opposite-sign deviations are not permitted). Single/two-phase units ship with a country-specific plug that must be specified at order. The equipment is EMC Class A, Group 1 (EN 55011), i.e. intended for industrial environments, and is pressure equipment under Art. 4.3 PED. The second stage uses a flammable A3 refrigerant, which governs handling and service procedures.

Integration & Operating Notes

Process Limits
Cooling capacity collapses toward -100 °C

Capacity is only 0.01 kW at -100 °C versus 0.16 kW at 0 °C, so the load at the probe must be kept very small at the low end; oversizing the thermal mass or heat leak will prevent reaching setpoint.

Datasheet, Specifications (cooling capacity
Mounting
Probe geometry must match the vessel opening

The probe is 48 mm nominal diameter and 155 mm long, and the flexible connection is 1150 mm; confirm the vessel port and the physical distance between the refrigeration chassis and the probe before ordering.

Datasheet, Specifications (probe dimensions, flexible connection
Electrical
Voltage and frequency tolerances are coupled

A -5 % voltage deviation combined with +2 % frequency is not allowed; supply quality must satisfy EN 60034-1 with matched deviation signs, otherwise the compressor is at risk.

Datasheet, Installation (mains tolerances
Electrical
Country-specific plug must be ordered

Single/two-phase units are delivered with a power cable and a country-specific plug that must be specified at ordering; three-phase units below 63 A come with cable but no plug.

Datasheet, Installation (electrical connection
Electrical
Protection must be 10–16 A

Maximum current is 6.5 A; the upstream fuse must be at least 10 A and no more than 16 A, so the supply circuit and breaker rating must be chosen within this window.

Datasheet, Specifications (fuse, current
Cooling & Media
Air-cooled condenser needs clearance and cool room air

The unit is air-cooled; space requirements around the chassis must be respected for airflow and service, and ambient temperature must stay within 5–40 °C, with capacity derating above the reference ambient.

Datasheet, Installation (space requirements, ambient temperature
Maintenance
Flammable A3 refrigerant in second stage

The second stage contains R-1150 (A3, H220); service and leak handling must follow flammable-refrigerant procedures, and only qualified personnel should open the circuit.

Datasheet, Specifications (refrigerant, safety class

Selection & Integration Q&A

When should I choose the TC100 rather than the TC100E or TC100-F?

Pick the TC100 when basic ON-OFF control and 55 dB(A) are sufficient. The TC100E/TC100E-F add electronics; the TC100-F is the -F variant at 58 dB(A). Choose a sibling only if its extra function or noise level is required.

What supply and protection does the TC100 need at the installation point?

220–240 V, 1~ or 2~, 50/60 Hz, max 6.5 A, with a 10–16 A fuse. Voltage and frequency tolerances are coupled (±5 % V only with ±2 % Hz, no opposite-sign deviation). Order the country-specific plug; three-phase units below 63 A ship without plug.

What is not included in the base delivery and what service constraints apply?

The country-specific plug must be specified at order; no pump or external circulation is included. The second stage uses flammable R-1150 (A3), so service must follow flammable-refrigerant procedures. EMC Class A means industrial-environment use only.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
TC100 Eintauchkühler
Datasheet ·79,5 КБ ·1 pp.
View PDF