Peter Huber Kältemaschinenbau TC100-F
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau TC100-F

Peter Huber Kältemaschinenbau TC100-F — Immersion coolers with Noise level 58 dB.

The TC100-F is an air-cooled two-stage immersion cooler with a 13 mm × 900 mm probe and 1150 mm flexible connection, rated for -100 to 40 °C using flammable A3 refrigerants (R-1270 first stage, R-1150 second stage). It is the -F variant of the TC100 series, distinguished by its fixed probe assembly for direct immersion into vessels and baths. It serves as a point-of-use cold source for small heat loads in laboratory and industrial processes requiring temperatures down to -100 °C.

Noise level
58 dB

Peter Huber Kältemaschinenbau Eintauchkuehler Kuehlfallen Series — Model Comparison & Selection

Model Noise level, dBControllerPower supplyRefrigerantSafety classTemperature range, °C
TC100-F Current 58ON-OFF-Regler220-240V 1~/2~ 50/60HzR-1150 (A3, H220)I / NFL-100–40
CT50 Single OLÉ 59OLÉ220-240V 1~/2~ 50/60HzR-290 (A3, H220)III / FL-50–50
TC100 55ON-OFF-Regler220-240V 1~/2~ 50/60HzR-1150 (A3, H220)I / NFL-100–40
TC100E-F 55MPC110-120V 1~ 60HzR-1150 (A3, H220)I / NFL-100–40
TC45 —ON-OFF-Regler208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-45–100
TC45E —MPC208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-45–100
TC45E-F 54MPC208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-45–100
TC45-F 59ON-OFF-Regler208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-45–100
TC50 —ON-OFF-Regler208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-50–50
TC50E —MPC208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-50–50
TC50E-F 60MPC208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-50–50
TC50-F 60ON-OFF-Regler208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-50–50

Full specifications

Core Performance & Functional Specifications
Noise level 58 dB(A)
Pump speed adjustable no
Pump type no pump
Temperature range -100...40 °C
Dimensions (W x D x H) mfr 295 x 500 x 570 mm
Refrigeration machine mfr air-cooled, natural refrigerant
Dimensions, Weight & Mechanical Design
Weight 57 kg
Electrical, Power & Drive Specifications
Current consumption 7 A (max.)
Power supply 220-240V 1~/2~ 50/60Hz
Interfaces, Connections & Control
Controller ON-OFF-controller
Materials, Media & Process Compatibility
Refrigerant R-1270 (A3, H220) (1st stage)
2nd stage: R-1150 (A3, H220)
Refrigerant quantity 1st stage mfr 0, 114 kg (1st stage)
Refrigerant quantity 2nd stage mfr 0, 056 kg (2nd stage)
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling capacity mfr 0: 0, 16; -20: 0, 15; -30: 0, 14; -50: 0, 13; -6 (at °C)
Standards, Safety & General Data
Safety class I / NFL

Description

Construction and series positioning

Behavior under load

Installation, connection and operation

Integration & Operating Notes

Process Limits
Cooling capacity collapses below -80 °C

At -90 °C only 0.07 kW and at -100 °C only 0.01 kW remain available, so the lowest temperatures are usable only with very small heat loads. Size the process heat load against the capacity at the actual setpoint, not at 0 °C.

TC100-F Temperature Controller Datasheet, Technical Specifications
Electrical
Voltage and frequency tolerances are coupled

A ±5% voltage deviation is permitted only with ±2% frequency deviation; combinations such as -5% voltage with +2% frequency are explicitly not allowed. Verify the site supply envelope before connection to avoid operating outside the permitted range.

TC100-F Temperature Controller Datasheet, Technical Specifications
Electrical
Country-specific plug must be specified at ordering

Single- and two-phase devices (100 V to 240 V) are delivered with a power cable and a country-specific plug that must be stated when ordering. Plan the order configuration and the site receptacle accordingly.

TC100-F Temperature Controller Datasheet, Technical Specifications
Mounting
Ambient must stay between 5 °C and 40 °C

The air-cooled unit may only be operated with ambient temperature between 5 °C and 40 °C, and cooling capacity is published at 20 °C room temperature and may drop as ambient rises. Keep the installation location within this range and respect space requirements for heat rejection.

TC100-F Temperature Controller Datasheet, Technical Specifications
Cooling & Media
Flammable A3 refrigerants require site precautions

Both circuits use flammable A3 refrigerants: R-1270 (0.114 kg) in the first stage and R-1150 (0.056 kg) in the second stage. Ventilation, ignition-source control and handling procedures must account for A3 classification at the installation site.

TC100-F Temperature Controller Datasheet, Technical Specifications
Control & I/O
ON-OFF control only; no pump or pump speed adjustment

The controller is an ON-OFF type and the unit has no pump and no adjustable pump speed. Process control must tolerate the resulting temperature cycling, and circulation, if required, must be provided externally.

TC100-F Temperature Controller Datasheet, Technical Specifications
Maintenance
Refrigeration unit runs continuously

The refrigeration unit works continuously, so heat rejection, noise (58 dB(A)) and wear are continuous during operation. Plan energy and heat-load budgets for uninterrupted running rather than intermittent duty.

TC100-F Temperature Controller Datasheet, Technical Specifications

Selection & Integration Q&A

When should I choose the TC100-F rather than another TC100 series variant?

Choose the TC100-F when the application needs a fixed 13 mm × 900 mm probe with 1150 mm flexible connection for direct immersion. Other TC100 series variants differ in configuration; select the variant whose probe or interface matches your vessel and routing.

What supply and protection must be provided for the TC100-F?

Provide 220–240 V single- or two-phase 50/60 Hz with a 10–16 A fuse; maximum current is 7 A. Voltage and frequency tolerances are coupled (±5% V only with ±2% Hz), and the country-specific plug must be specified at ordering.

Where can the TC100-F be installed and how must it be operated?

Install only where ambient stays between 5 °C and 40 °C, with space for air-cooled heat rejection. It is EMC Class A for industrial environments, uses flammable A3 refrigerants, and the refrigeration unit runs continuously.

What is included in delivery and what must be ordered or provided separately?

Single- and two-phase units ship with a power cable and country-specific plug, which must be specified at ordering. No pump is fitted, so any circulation must be provided externally; confirm probe fit (13 mm × 900 mm, 1150 mm flexible connection) before installation.