Peter Huber Kältemaschinenbau TC45-F
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau TC45-F

Peter Huber Kältemaschinenbau TC45-F — Immersion coolers with Noise level 59 dB.

The TC45-F is an air-cooled immersion cooler with a 13 mm × 600 mm probe and 1050 mm flexible connection, rated for -45 to 100 °C using R-290 refrigerant. Its 0.24 kW at 0 °C falls to 0.05 kW at -40 °C, so it suits low-load setpoint holding rather than high-heat-load pull-down. It is the compact low-capacity member of the series, intended for direct immersion into vessels or baths.

Noise level
59 dB

Peter Huber Kältemaschinenbau Eintauchkuehler Kuehlfallen Series — Model Comparison & Selection

Model Noise level, dBControllerPower supplyRefrigerantSafety classTemperature range, °C
TC45-F Current 59ON-OFF-Regler208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-45–100
CT50 Single OLÉ 59OLÉ220-240V 1~/2~ 50/60HzR-290 (A3, H220)III / FL-50–50
TC100 55ON-OFF-Regler220-240V 1~/2~ 50/60HzR-1150 (A3, H220)I / NFL-100–40
TC100E-F 55MPC110-120V 1~ 60HzR-1150 (A3, H220)I / NFL-100–40
TC100-F 58ON-OFF-Regler220-240V 1~/2~ 50/60HzR-1150 (A3, H220)I / NFL-100–40
TC45 —ON-OFF-Regler208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-45–100
TC45E —MPC208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-45–100
TC45E-F 54MPC208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-45–100
TC50 —ON-OFF-Regler208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-50–50
TC50E —MPC208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-50–50
TC50E-F 60MPC208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-50–50
TC50-F 60ON-OFF-Regler208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-50–50

Full specifications

Core Performance & Functional Specifications
Noise level 59 dB(A)
Pump speed adjustable no
Pump type no pump
Temperature range -45...100 °C
Dimensions (W x D x H) mfr 190 x 295 x 360 mm
Refrigeration machine mfr air-cooled, natural refrigerant
-20 mfr 0, 18 kW (at -20 °C)
-30 mfr 0, 1 kW (at -30 °C)
-40 mfr 0, 05 kW (at -40 °C)
Dimensions, Weight & Mechanical Design
Weight 16 kg
Electrical, Power & Drive Specifications
Power supply 208-240V 1~/2~ 50/60Hz
Current consumption max. mfr 1, 4 A (max)
Interfaces, Connections & Control
Controller ON-OFF-controller
Materials, Media & Process Compatibility
Refrigerant R-290 (A3, H220) (1st stage)
Refrigerant quantity 1st stage mfr 0, 041 kg (1st stage)
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Cooling capacity mfr 0, 24 kW (at 0 °C)
Standards, Safety & General Data
Safety class I / NFL

Description

Construction and series positioning

Behavior under load

Installation and operation

Integration & Operating Notes

Electrical
Supply tolerance is a combined voltage/frequency rule

The ±5% voltage tolerance is only valid together with ±2% frequency tolerance, and the combination of -5% voltage with +2% frequency is explicitly not allowed. Verify the site supply against this combined window before commissioning, because a supply that passes each limit separately may still be outside the permitted combination.

TC45-F Datasheet, Specifications
Process Limits
Cooling capacity collapses at the cold end

Capacity falls from 0.24 kW at 0 °C to 0.05 kW at -40 °C. The required heat load at the lowest operating temperature must be checked against this curve; sizing on the 0 °C figure will under-cool at deep setpoints.

TC45-F Datasheet, Specifications
Mounting
Probe geometry fixes vessel opening and immersion depth

The probe is 13 mm in diameter and 600 mm long with a 1050 mm flexible connection. Confirm the vessel port accepts the 13 mm probe and that 600 mm insertion reaches the required liquid depth without the flexible connection being strained.

TC45-F Datasheet, Specifications
Environment
IP20 and flammable refrigerant restrict siting

IP20 requires a dry indoor location protected from water and dust. R-290 is a flammable A3 refrigerant, so the installation area must be assessed for ignition sources and ventilation before the unit is placed.

TC45-F Datasheet, Specifications
Electrical
Upstream protection must sit within 10–16 A

The datasheet specifies a minimum fuse of 10 A and a maximum of 16 A. Size the upstream protection inside this window; a breaker outside it is not covered by the manufacturer's installation requirements.

TC45-F Datasheet, Specifications
Electrical
Plug type must be ordered, not assumed

Single- and two-phase devices are delivered with a power cable and a country-specific plug that must be specified at ordering. Confirm the plug standard with the customer before release to avoid rework at commissioning.

TC45-F Datasheet, Specifications
Cooling & Media
Continuous refrigeration adds a permanent room heat load

The refrigeration unit works continuously rather than cycling, so the room receives a continuous heat rejection. Account for this in room ventilation or air-conditioning sizing, especially in small or closed laboratories.

TC45-F Datasheet, Specifications

Selection & Integration Q&A

When should I choose the TC45-F rather than a TC100/TC100E sibling?

Pick the TC45-F for low heat loads at deep setpoints down to -45 °C where its 0.05 kW at -40 °C suffices. Choose a TC100-series sibling when higher capacity or lower noise (55 dB(A) vs 59 dB(A)) is needed.

What supply and upstream protection does the TC45-F require?

208–240 V, 1~/2~, 50/60 Hz, max 1.4 A, with a 10–16 A upstream fuse. Voltage tolerance ±5% applies only with ±2% frequency tolerance; -5% voltage with +2% frequency is not allowed. Specify the country plug at order.

Where can the TC45-F be installed and how is it operated?

Dry indoor location, ambient 5–40 °C, IP20, EMC Class A industrial environment. R-290 flammable refrigerant requires ignition-source and ventilation assessment. ON-OFF control, no pump; refrigeration runs continuously.

What is not included in base delivery and what must be planned for service?

The country-specific plug must be specified at order; no pump is fitted. The 600 mm probe and 1050 mm flexible connection define the immersion interface. Continuous operation means ongoing room heat rejection must be planned.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
TC45-F Datasheet