Peter Huber Kältemaschinenbau TC50
Peter Huber Kältemaschinenbau TC50 — Immersion coolers.
The TC50 is an air-cooled immersion cooler with a flexible 1150 mm connection between the refrigerating unit and a 48 mm × 155 mm probe, rated for -50 to 50 °C. It is the smallest model in the series, distinguished by its 0.3 kW cooling capacity at 0 °C and R-290 refrigerant. It is used for direct cooling of process fluids in vessels or baths.
Peter Huber Kältemaschinenbau Eintauchkuehler Kuehlfallen Series — Model Comparison & Selection
| Model | Noise level, dB | Controller | Power supply | Refrigerant | Safety class | Temperature range, °C |
|---|---|---|---|---|---|---|
| TC50 Current | — | ON-OFF-Regler | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -50–50 |
| CT50 Single OLÉ | 59 | OLÉ | 220-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | III / FL | -50–50 |
| TC100 | 55 | ON-OFF-Regler | 220-240V 1~/2~ 50/60Hz | R-1150 (A3, H220) | I / NFL | -100–40 |
| TC100E-F | 55 | MPC | 110-120V 1~ 60Hz | R-1150 (A3, H220) | I / NFL | -100–40 |
| TC100-F | 58 | ON-OFF-Regler | 220-240V 1~/2~ 50/60Hz | R-1150 (A3, H220) | I / NFL | -100–40 |
| TC45 | — | ON-OFF-Regler | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -45–100 |
| TC45E | — | MPC | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -45–100 |
| TC45E-F | 54 | MPC | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -45–100 |
| TC45-F | 59 | ON-OFF-Regler | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -45–100 |
| TC50E | — | MPC | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -50–50 |
| TC50E-F | 60 | MPC | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -50–50 |
| TC50-F | 60 | ON-OFF-Regler | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -50–50 |
Full specifications
| Pump speed adjustable | no |
| Pump type | no pump |
| Temperature range | -50...50 °C |
| Dimensions (W x D x H) mfr | 260 x 330 x 415 mm |
| Refrigeration machine mfr | air-cooled, natural refrigerant |
| -20 mfr | 0, 26 kW (at -20 °C) |
| -30 mfr | 0, 2 kW (at -30 °C) |
| Weight | 26 kg |
| Current consumption | 3 A (max.) |
| Power supply | 208-240V 1~/2~ 50/60Hz |
| Controller | ON-OFF-controller |
| Refrigerant | R-290 (A3, H220) (1st stage) |
| Refrigerant quantity 1st stage mfr | 0, 06 kg (1st stage) |
| Ambient temperature limit | 5...40 °C |
| Cooling capacity mfr | 0, 3 kW (at 0 °C) |
| Safety class | I / NFL |
Description
Construction and positioning
- The TC50 is a split immersion cooler: an air-cooled refrigerating unit connected by a 1150 mm flexible line to a probe with 48 mm nominal diameter and 155 mm length.
- It is the base model of the series; siblings TC100, TC100E and TC100E-F offer higher capacity (noise levels 55–59 dB(A) are listed for those variants, not for the TC50).
- Refrigerant is R-290 (A3, H220) with a charge of 0.06 kg (CO2 equivalent 0.0 t).
- The unit has no pump and no adjustable pump speed; cooling is delivered directly through the probe.
Performance and media behavior
- Cooling capacity is 0.3 kW at 0 °C, 0.26 kW at -20 °C and 0.2 kW at -30 °C, valid at 20 °C room temperature; capacity may drop as ambient temperature rises.
- Operating temperature range is -50 to 50 °C; the ON-OFF controller regulates within this envelope.
- The probe is intended for direct immersion in the process fluid; no pump means circulation must be provided externally if required.
Installation, control and operation
- Power supply: 208–240 V 1~/2~ 50/60 Hz, max current 3 A, fuse 10–16 A. Voltage tolerance ±5% with simultaneous frequency tolerance ±2%; the combination -5% voltage and +2% frequency is not allowed.
- Single/two-phase devices are delivered with a power cable and country-specific plug that must be specified at ordering.
- Ambient temperature must be 5–40 °C. EMC classification is Class A, Group 1 to EN55011 (industrial environment).
- US-SNAP end-use is industrial process refrigeration.
Integration & Operating Notes
The vessel or bath opening must accept a 48 mm nominal diameter probe, and the immersion depth must be at least 155 mm to use the full probe length. Check these dimensions before mounting to avoid mechanical interference or insufficient cooling.
The refrigerating unit must be positioned within 1150 mm of the probe connection point. Routing must avoid stretching or kinking the flexible line, which could damage the refrigerant circuit.
Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%; the combination of -5% voltage and +2% frequency is not allowed. Verify the site supply before commissioning to avoid invalid operating conditions.
Single/two-phase devices are delivered with a power cable and country-specific plug. The plug type must be stated when ordering; otherwise the unit may arrive without a suitable connector for the installation site.
Operation outside this range may degrade cooling performance or damage the unit. The installation location must be controlled within this range, and cooling capacity may drop as ambient temperature rises above 20 °C.
R-290 is a flammable refrigerant (A3, H220). Installation must account for A3 safety requirements, including room ventilation and avoidance of ignition sources, even though the charge is only 0.06 kg.
The equipment is EMC Class A, Group 1 to EN55011, intended for industrial environments. It may cause interference in residential areas, so additional EMC measures may be needed if installed near sensitive or residential loads.
Selection & Integration Q&A
Choose the TC50 when the required cooling capacity at your process temperature is at or below 0.3 kW at 0 °C (0.26 kW at -20 °C, 0.2 kW at -30 °C) and the 48 mm × 155 mm probe fits your vessel. The TC100 series offers higher capacity but is physically larger; select it only if the TC50's capacity is insufficient.
The ambient temperature must remain between 5 °C and 40 °C. Cooling capacity is rated at 20 °C room temperature and may drop as ambient temperature rises. The unit is air-cooled, so space requirements for ventilation must be respected; insufficient clearances can impair condenser performance.
The base delivery includes the refrigerating unit, probe, and power cable with country-specific plug (if specified). No pump is included. Maintenance requirements are not detailed in the available datasheet; consult the operating manual for service intervals. A suitable fuse (10–16 A) and electrical protection must be provided by the installer.