Peter Huber Kältemaschinenbau Unimotive 26w
Peter Huber Kältemaschinenbau Unimotive 26w — Temperature control units.
The Unimotive 26w is a water-cooled thermostatic bath/circulator with a 24 kW heater and a magnetically coupled stainless-steel pump, covering -45 to 95 °C for water-glycol service. It is the mid-capacity water-cooled variant in the Unimotive series, distinguished from the 10w and 20w by its higher heating capacity and cooling-water demand. It is intended for automotive temperature simulation and material testing where deep cooling and high heat rejection are required.
Peter Huber Kältemaschinenbau Unimotive Series — Model Comparison & Selection
| Model | Display resolution | Heating capacity, W | Power supply | Pump connection | Refrigerant | Safety class |
|---|---|---|---|---|---|---|
| Unimotive 26w Current | 0, 01 K | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | III / FL |
| Unimotive 10w | 0.01 | 12000 | 400V 3~ 50Hz | M38x1,5 AG male | R-452A (A1, H280) | III / FL |
| Unimotive 10w-XT | 0, 01 | 12000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | I / NFL |
| Unimotive 20w | 0, 01 | 12000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | III / FL |
| Unimotive 20w-XT | 0, 01 | 12000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | I / NFL |
| Unimotive 26w-XT | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | I / NFL |
| Unimotive 27w | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | III / FL |
| Unimotive GL 07w | 0, 01 | 12000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 07w-XT | 0, 01 | 12000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | I / NFL |
| Unimotive GL 10w | 0, 01 | 24000 | 460V 3~ 60Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 10w-XT | 0, 01 | 24000 | 460V 3~ 60Hz | M38x1, 5 AG male | R-744 (A1, H280) | I / NFL |
| Unimotive GL 20w | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 20w-XT | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | I / NFL |
| Unimotive GL 30w | 0, 01 | 24000 | 460V 3~ 60Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 50w | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 50w-XT | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | I / NFL |
| Unimotive GL 60w | 0, 01 | 48000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 60w-XT | 0, 01 | 48000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | I / NFL |
Full specifications
| Control output | POKO ONE |
| Heating capacity | 24 kW |
| Pump connection | M38x1, 5 AG male |
| Pump speed adjustable | no |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -45...95 °C |
| Weight | 462 kg |
| Current consumption | 38 A (max) |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Refrigerant | R-452A (A1, H280) (1st stage) |
| Maximum viscosity | 200 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G3/4 male |
| Cooling water consumption | 1260 l/h (at 15°C, flow 0°C) |
| Cooling water differential pressure | 1 bar (min) |
| Cooling Water Pressure | 6 bar (max) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
The Unimotive 26w is a water-cooled temperature control unit built for operation with water-ethylene glycol mixtures containing corrosion protection down to -45 °C. Its pump is magnetically coupled and made of stainless steel, with a fixed (non-adjustable) speed and an M38x1.5 male pump connection. The cooling circuit uses R-452A in the first stage. Within the Unimotive series, the 26w sits above the 10w and 20w models in heating capacity (24 kW) and cooling-water consumption (1260 l/h at 15 °C, flow 0 °C), making it the choice when larger thermal loads must be handled; the smaller siblings are preferable where facility cooling water is limited or the load is lower. The 10w-XT and 20w-XT variants are not described here.
Behavior under load and media constraints
Cooling capacity falls steeply at low temperatures: 28 kW at 20 °C but only 2.6 kW at -40 °C, so deep-cooling duty must be sized against the low-temperature figure, not the nominal. The pump delivers 201 l/min at zero head but only 42 l/min at 5 bar, meaning external circuit pressure drop directly limits available flow. The heat transfer fluid must not exceed 200 mm²/s kinematic viscosity, which restricts very viscous fluids at low temperature. The unit is not suitable for pressure-sensitive glass reactors without accounting for maximum flow pressure and possibly an external bypass.
Installation, connection and control
The controller is Pilot ONE with POKO ONE control output, a display resolution of 0.01 K, and interfaces Ethernet, USB (host and device) and RS232; an external Pt100 sensor can be connected. Electrical supply is 400 V 3~ 50 Hz with a maximum current of 38 A and 3x40 A fusing; three-phase devices below 63 A are delivered with a cable but without a plug. Leakage current exceeds 3.5 mA, so additional protective measures may be required. Cooling water must be supplied at 5–20 °C with a differential pressure of 1–4 bar and a maximum supply pressure of 6 bar; a backflow preventer on the building side is mandatory, and sea water must not be used. The unit is IP20 and must be installed in an ambient temperature of 5–40 °C, outside any ATEX zone, with a drip tray below it and minimum wall clearances observed.
Integration & Operating Notes
The cooling water intake temperature is limited to 5–20 °C, and the differential pressure must be between 1 and 4 bar with a maximum supply pressure of 6 bar. Insulate pipes if the water temperature can fall below the dew point, and provide a backflow preventer on the building side to protect the drinking water supply.
Cooling capacity drops from 28 kW at 20 °C to 2.6 kW at -40 °C, so low-temperature loads must be sized against the reduced capacity. The heat transfer fluid must also remain below 200 mm²/s kinematic viscosity, which may limit fluid selection at the low end of the -45 to 95 °C range.
Because the current consumption is below 63 A, the unit is delivered with a cable but without a plug. The customer must provide a suitable plug and socket with protective earth, or have a certified electrician perform a fixed connection. Replacement cables are limited to 4 m for three-phase connections.
Selection & Integration Q&A
Choose the 26w when the thermal load requires up to 24 kW heating and cooling-water flow around 1260 l/h at 15 °C is available. The 10w and 20w have lower capacities and are preferable where facility cooling water is limited or the load is smaller. The 26w also provides deeper cooling capacity at low temperatures than the smaller models.
Cooling water must be 5–20 °C, with a differential pressure of 1–4 bar and maximum supply pressure of 6 bar. Connections are G3/4 male. A backflow preventer on the building side is mandatory, sea water is prohibited, and pipes must be insulated if condensation can occur. Only drinking water or filtered river water may be used.
Install indoors or with frost protection, ambient 5–40 °C, IP20, outside ATEX zones, with a drip tray below and specified wall clearances. Cooling water must not fall below 5 °C. The unit is Group 1 Class A for industrial EMC. Shut off cooling water during downtimes and bring fluid to room temperature before switching off.
The base delivery includes a power cable but no plug; the customer must supply a suitable plug or fixed connection. A building-side backflow preventer and a drip tray are required but not supplied. Regular maintenance includes checking cooling water quality and refrigerant charge; R-452A charge is 3.6 kg with GWP 2141, so leak checks and F-gas reporting apply.