Peter Huber Kältemaschinenbau Unimotive GL 50w
Peter Huber Kältemaschinenbau Unimotive GL 50w — Temperature control units.
The Unimotive GL 50w is a water-cooled temperature control unit with a magnetically coupled stainless-steel pump, 24 kW heating capacity and a -45...95 °C working range, built for water-ethylene glycol mixtures with corrosion protection. Within the Unimotive series it is the larger GL modification, distinguished by its 24 kW heater and G3/4 cooling-water connection. It serves automotive test benches and process circuits requiring sub-ambient to near-boiling temperature control.
Peter Huber Kältemaschinenbau Unimotive Series — Model Comparison & Selection
| Model | Display resolution | Heating capacity, W | Power supply | Pump connection | Refrigerant | Safety class |
|---|---|---|---|---|---|---|
| Unimotive GL 50w Current | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive 10w | 0.01 | 12000 | 400V 3~ 50Hz | M38x1,5 AG male | R-452A (A1, H280) | III / FL |
| Unimotive 10w-XT | 0, 01 | 12000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | I / NFL |
| Unimotive 20w | 0, 01 | 12000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | III / FL |
| Unimotive 20w-XT | 0, 01 | 12000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | I / NFL |
| Unimotive 26w | 0, 01 K | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | III / FL |
| Unimotive 26w-XT | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | I / NFL |
| Unimotive 27w | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | III / FL |
| Unimotive GL 07w | 0, 01 | 12000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 07w-XT | 0, 01 | 12000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | I / NFL |
| Unimotive GL 10w | 0, 01 | 24000 | 460V 3~ 60Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 10w-XT | 0, 01 | 24000 | 460V 3~ 60Hz | M38x1, 5 AG male | R-744 (A1, H280) | I / NFL |
| Unimotive GL 20w | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 20w-XT | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | I / NFL |
| Unimotive GL 30w | 0, 01 | 24000 | 460V 3~ 60Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 50w-XT | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | I / NFL |
| Unimotive GL 60w | 0, 01 | 48000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 60w-XT | 0, 01 | 48000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | I / NFL |
Full specifications
| Control output | POKO ONE (potential-free contact) |
| Heating capacity | 24 kW |
| Pump connection | M38x1, 5 AG male |
| Pump speed adjustable | no |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -45...95 °C |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Refrigerant | R-744 (A1, H280) (1st stage) |
| Maximum viscosity | 200 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G3/4 male |
| Cooling water differential pressure | 1 bar (min) |
| Cooling Water Pressure | 6 bar (max) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- The GL 50w is a water-cooled unit using R-744 (A1, H280) refrigerant in the first stage, with a magnetically coupled stainless-steel pump (M38x1.5 AG male connection, non-adjustable speed).
- Heating capacity is 24 kW; the working range is -45...95 °C, set against the smaller Unimotive 10w/20w and their -XT variants in the same series.
- The heat transfer circuit is built for water-ethylene glycol mixture with corrosion protection (e.g. Glysantin®) down to -45 °C, not for general-purpose fluids.
- Cooling water is connected via G3/4 male ports; the circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.
Behavior under load and media
- Published cooling capacities apply only at 20 °C room temperature, 15 °C cooling water inlet and 1 bar differential pressure; higher cooling water temperature reduces cooling power and may raise the required flow rate.
- Cooling water intake must be between 5 °C and 20 °C; below the condensation point the internal cooling water pipes must be insulated.
- The pump is not speed-adjustable, so flow pressure is fixed by the pump characteristic; pressure-sensitive applications such as glass reactors must account for the maximum flow pressure, possibly with an external bypass.
- Adjustable overtemperature protection to DIN 12876 is provided and must be set for the application.
Installation, control and operation
- Control is via the Pilot ONE controller with POKO ONE potential-free contact output; interfaces include Ethernet, USB (host and device) and RS232, plus a Pt100 external sensor connection.
- Electrical supply is 400 V 3~ 50 Hz with ±5% voltage tolerance and ±2% frequency tolerance; the combination -5% voltage with +2% frequency is not permitted. Protection rating is IP20.
- Three-phase units below 63 A are supplied with cable but without a plug; units above 63 A are supplied without cable and without plug.
- The unit is Class A, Group 1 to EN55011 and must not be installed in an ATEX zone. A drip tray for condensate/heat transfer fluid must be installed below the unit, and refrigerant charge versus room size must be checked.
- Cooling water must be drinking water or filtered river/sea water (sea water prohibited); a building-side backflow preventer is required, and cooling water should be shut off during downtimes.
Integration & Operating Notes
The unit is designed for cooling water intake between 5 °C and 20 °C. Outside this window the published cooling capacity is invalid and the site supply must be verified or modified before installation; below the condensation point the internal cooling water pipes must be insulated.
Supply tolerance is ±5% voltage with ±2% frequency, but the combination of -5% voltage and +2% frequency is explicitly not allowed. Plan the site supply and any generator/transformer configuration so this corner condition cannot occur.
Three-phase devices drawing less than 63 A are delivered with cable but without a plug, so a suitable plug must be provided on site. Devices above 63 A are supplied without cable and without plug, requiring full on-site cabling by a certified electrician.
The responsible body must install a condensation water / heat transfer fluid drip tray below the unit. Refrigerant charge and room size must be checked during installation to avoid dangerous refrigerant concentration from the R-744 charge.
The unit must not be installed or started inside an ATEX zone. Site classification must be confirmed before placement, regardless of the IP20 protection rating of the enclosure.
Only drinking water or filtered river/sea water may be used; sea water is prohibited. A building-side backflow preventer is required to protect the drinking water supply, and the mixed-material circuit (copper, 1.4401, MS, PA, PPE, PTFE, EPDM) constrains water treatment.
Cooling water can flow even when the unit is switched off, so the supply should be shut off even for short downtimes such as overnight. Outdoor installation additionally requires frost-protected lines and a minimum cooling water temperature of 5 °C.
The pump is not speed-adjustable, so the maximum flow pressure is fixed. For pressure-sensitive applications such as glass reactors, the maximum flow pressure must be taken into account and an external bypass may be required. Hose inside diameters must at least match the M38x1.5 pump connections.
The heat transfer fluid must always be adjusted to room temperature before switching off to avoid overpressure in the system. The adjustable overtemperature protection to DIN 12876 must be set for the application before commissioning.
Selection & Integration Q&A
Pick the GL 50w when the duty needs the 24 kW heater and the -45...95 °C range with water-ethylene glycol. The 10w/20w and their -XT variants are the smaller series members; confirm heating capacity and circuit size against your load before downsizing.
Provide G3/4 male cooling-water connections with 1 bar minimum differential and 6 bar maximum pressure, drinking or filtered river water (no sea water), and a building-side backflow preventer. Electrically, 400 V 3~ 50 Hz within ±5%/±2%; below 63 A the unit arrives with cable but no plug.
Not inside an ATEX zone. Install with the specified wall clearances, a drip tray below the unit, and a room-size check against the R-744 charge. Cooling water intake must be 5–20 °C; outdoor siting needs frost-protected lines and a minimum 5 °C cooling water temperature.
The base delivery excludes the power plug (below 63 A) or cable and plug (above 63 A), the drip tray, the building-side backflow preventer, and the water-ethylene glycol mixture with corrosion protection. Cooling water must be shut off during downtimes; the adjustable DIN 12876 overtemperature protection must be configured on site.