Peter Huber Kältemaschinenbau Unimotive GL 50w-XT
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unimotive GL 50w-XT

Peter Huber Kältemaschinenbau Unimotive GL 50w-XT — Temperature control units.

The Unimotive GL 50w-XT is a water-cooled temperature control unit built for water-ethylene glycol mixtures down to -45 °C, covering a -45...150 °C working range with 24 kW heating capacity. Within the Unimotive series it is the large GL modification with R-744 refrigerant and a magnetically coupled stainless-steel pump, intended for automotive test benches and process circuits requiring sub-zero glycol temperature control.

Peter Huber Kältemaschinenbau Unimotive Series — Model Comparison & Selection

Model Display resolutionHeating capacity, WPower supplyPump connectionRefrigerantSafety class
Unimotive GL 50w-XT Current 0, 0124000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)I / NFL
Unimotive 10w 0.0112000400V 3~ 50HzM38x1,5 AG maleR-452A (A1, H280)III / FL
Unimotive 10w-XT 0, 0112000400V 3~ 50HzM38x1, 5 AG maleR-452A (A1, H280)I / NFL
Unimotive 20w 0, 0112000400V 3~ 50HzM38x1, 5 AG maleR-452A (A1, H280)III / FL
Unimotive 20w-XT 0, 0112000400V 3~ 50HzM38x1, 5 AG maleR-452A (A1, H280)I / NFL
Unimotive 26w 0, 01 K24000400V 3~ 50HzM38x1, 5 AG maleR-452A (A1, H280)III / FL
Unimotive 26w-XT 0, 0124000400V 3~ 50HzM38x1, 5 AG maleR-452A (A1, H280)I / NFL
Unimotive 27w 0, 0124000400V 3~ 50HzM38x1, 5 AG maleR-452A (A1, H280)III / FL
Unimotive GL 07w 0, 0112000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)III / FL
Unimotive GL 07w-XT 0, 0112000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)I / NFL
Unimotive GL 10w 0, 0124000460V 3~ 60HzM38x1, 5 AG maleR-744 (A1, H280)III / FL
Unimotive GL 10w-XT 0, 0124000460V 3~ 60HzM38x1, 5 AG maleR-744 (A1, H280)I / NFL
Unimotive GL 20w 0, 0124000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)III / FL
Unimotive GL 20w-XT 0, 0124000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)I / NFL
Unimotive GL 30w 0, 0124000460V 3~ 60HzM38x1, 5 AG maleR-744 (A1, H280)III / FL
Unimotive GL 50w 0, 0124000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)III / FL
Unimotive GL 60w 0, 0148000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)III / FL
Unimotive GL 60w-XT 0, 0148000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)I / NFL

Full specifications

Core Performance & Functional Specifications
Control output POKO ONE
Heating capacity 24 kW
Pump connection M38x1, 5 AG male
Pump speed adjustable no
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -45...150 °C
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Refrigerant R-744 (A1, H280) (1st stage)
Maximum viscosity 200 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G3/4 male
Cooling water differential pressure 1 bar (min)
Cooling Water Pressure 6 bar (max)
Protection rating IP20
Standards, Safety & General Data
Safety class I / NFL

Description

Construction and series positioning

Working range and media

Installation, connection and control

Integration & Operating Notes

Cooling & Media
Cooling water loop must deliver 1 bar differential at G3/4

The water-cooled refrigeration machine needs a plant cooling water supply connected to G3/4 male ports with at least 1 bar differential pressure and no more than 6 bar static pressure. Verify the loop in the piping design before the unit is delivered, since without adequate differential pressure the refrigerant stage cannot reject heat.

Datasheet, Technical Specifications; Operation Manual, Section on Cooling Water
Cooling & Media
Cooling water intake above 20 °C degrades cooling power

The unit is designed for cooling water intake between 5 °C and 20 °C. Higher intake temperatures reduce available cooling power and may increase flow demand, so a site survey of the summer cooling water temperature is required. Below the condensation point, the internal cooling water pipes must be insulated to prevent condensation and corrosion.

Datasheet, Technical Specifications
Electrical
Voltage and frequency tolerances are coupled, not independent

The 400 V 3~ 50 Hz supply is permitted only within ±5% voltage and ±2% frequency, and certain combinations such as -5% voltage with +2% frequency are explicitly not allowed. Check the site power quality against this envelope; the unit is EN55011 Class A Group 1 and is intended for industrial EMC environments only.

Datasheet, Technical Specifications; Operation Manual, Section on Electrical Connection
Electrical
Three-phase unit arrives with cable but no plug

Three-phase devices with current consumption below 63 A are delivered with a cable but without a plug, so a suitable plug must be sourced and fitted on site. Replacement power cables are limited to 4 m for three-phase connections; hardwiring must be performed by a certified electrician and the unit must always be connected to a socket with protective earth.

Datasheet, Technical Specifications; Operation Manual, Section on Electrical Connection
Mounting
Drip tray and refrigerant room-size check are site obligations

The responsible body must install a condensation water / heat transfer fluid drip tray below the unit. Refrigerant charge (see rating plate) and room size must be checked during installation to avoid dangerous refrigerant concentration. The unit must not be installed or started inside an ATEX zone, and minimum wall clearances per the manual table must be respected.

Operation Manual, Sections on Installation and Commissioning
Process Limits
Glycol circuit must be vented to room temperature before shutdown

The heat transfer fluid must always be adjusted to room temperature before switching off to avoid overpressure in the system. The maximum system pressure is 12 bar, and pressure-sensitive downstream apparatus such as glass reactors must account for the unit's maximum flow pressure, possibly with an external bypass. Hose inside diameters must at least match the M38x1.5 pump connections.

Operation Manual, Sections on Normal Operation and Shutting Down
Maintenance
Shut off cooling water during even short downtimes

Cooling water flow can occur even when the unit is switched off, so the supply should be shut off overnight and during other downtimes to prevent water damage and unnecessary consumption. Only drinking water or filtered river water may be used; sea water is prohibited. A building-side backflow preventer is required to protect the drinking water supply.

Operation Manual, Section on Cooling Water

Selection & Integration Q&A

When should I choose the GL 50w-XT rather than a smaller Unimotive 10w or 20w variant?

Pick the GL 50w-XT when the process needs the full -45...150 °C glycol range with 24 kW heating and a water-cooled R-744 stage. The 10w/20w siblings suit lower heat loads; the GL 50w-XT is the high-capacity glycol modification of the series.

What plant-side services must be ready before the unit can be commissioned?

A 400 V 3~ 50 Hz supply within ±5% voltage / ±2% frequency (plug not included), a G3/4 cooling water loop at 1...6 bar with 5...20 °C intake, a drip tray below the unit, and a building-side backflow preventer.

What environment and duty-cycle limits apply to installation?

Ambient 5...40 °C, IP20, no ATEX zone, EN55011 Class A Group 1 industrial EMC environment only. Refrigerant charge versus room size must be checked, and cooling water must be shut off during downtimes since flow can occur with the unit switched off.

What is not included in base delivery and what maintenance/consumables should be planned?

The three-phase cable is supplied but the plug is not; the drip tray and backflow preventer are site-provided. Plan glycol mixture with corrosion protection, cooling water treatment compatible with copper/1.4401/EPDM, and periodic checks of the cooling water circuit per the manual.