Peter Huber Kältemaschinenbau Unistat 410w Dynamisches
Peter Huber Kältemaschinenbau Unistat 410w Dynamisches — Temperature control units with Noise level 61 dB.
The Unistat 410w Dynamisches is a water-cooled, free-standing temperature control unit with a variable-speed pressure pump and Pilot ONE controller, covering -45 to 250 °C with 3 kW heating. Its water-cooled refrigeration (R-1270, A3) distinguishes it from air-cooled siblings, requiring a 5–20 °C cooling water circuit. It serves precise temperature control of externally closed or open applications in process engineering.
Peter Huber Kältemaschinenbau Unistate Bis 45C Series — Model Comparison & Selection
| Model | Noise level, dB | Display resolution | Power supply | Pump connection | Pump type | Temperature range, °C |
|---|---|---|---|---|---|---|
| Unistat 410w Dynamisches Current | 61 | 0, 01 K | 220-240V 1~/2~ 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | -45–250 |
| Grande Fleur Dynamisches | — | — | — | — | — | — |
| Grande Fleur w Dynamisches | — | 0, 01 | 208-240V 1~/2~ 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | -40–200 |
| Grande Fleur w-eo Dynamisches | — | — | — | — | — | — |
| Petite Fleur-eo Dynamisches | 58 | 0, 01 K | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | Pressure pump, variable | -40–200 |
| Petite Fleur w Dynamisches | — | 0, 01 | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | Pressure pump, variable | -40–200 |
| Unistat 425w Dynamisches | 61 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
| Unistat 430w Dynamisches | 61 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
| Unistat P405w Dynamisches | — | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st | -45–250 |
| Unistat P425 Dynamisches | 65 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
| Unistat P425w Dynamisches | 61 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
| Unistat P430 Dynamisches | 65 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
| Unistat P430w Dynamisches | 65 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
Full specifications
| Noise level | 61 dB(A) (sound pressure level) |
| Control output | POKO ONE (potential-free contact) |
| Heating capacity | 3 kW |
| Pump connection | M24x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | Pressure pump, variable speed |
| External sensor connection | Pt100 |
| Temperature range | -45...250 °C |
| Weight | 68 kg (net) |
| Current consumption | 16 A (max) |
| Fuse rating | 16 A |
| Power supply | 220-240V 1~/2~ 50/60Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s (max permissible) |
| Ambient temperature limit | 5...40 °C (permissible) |
| Cooling water connection | G1/2 male |
| Cooling water differential pressure | 1 bar (min) |
| Cooling Water Pressure | 6 bar (max) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Free-standing Unistat 4xx platform, water-cooled ('w' suffix) and 'Dynamisches' (dynamic) modification; no plate heat exchanger is fitted on the Unistat 410(w) per the operation manual, so the usual warning against rapid, repeated high-temperature cycling does not apply in the same way.
- Sibling models in the series include Grande Fleur variants (Dynamisches, -eo, w, w-eo); this 410w is the water-cooled dynamic unit, whereas air-cooled versions need no cooling water circuit.
- Refrigerant R-1270 (A3, H220, flammable) with a factory-tested tight circuit (indicator sensitivity 10⁻⁴ Pa·m³/s or better); refrigerant work is restricted to approved refrigeration contractors.
Working range and behavior
- Temperature range -45 to 250 °C; heating capacity 3 kW; display resolution 0.01 K.
- Variable-speed pressure pump with M24x1.5 male connection; maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s.
- Cooling power depends on cooling water temperature: performance drops and required flow rises as inlet water warms; inlet must stay between 5 °C and 20 °C.
- Process Safety mode releases maximum cooling at the overtemperature trigger instead of switching off all poles; it requires the compressor set to Permanently ON.
Installation, control and operation
- Water circuit: G1/2 male connections, minimum 1 bar differential pressure, maximum 6 bar; insulate cooling water pipes when water is below the condensation point.
- Electrical: 220–240 V 1~/2~ 50/60 Hz, max 16 A, 16 A fuse; voltage tolerance ±5% and frequency ±2% must not combine unfavorably; hard-wired connection by an electrician.
- Pilot ONE controller with POKO ONE potential-free contact, Ethernet, USB host/device and RS232; external Pt100 sensor input; IP20, ambient 5–40 °C; EMC Class A Group 1.
- Leveling feet/parking brakes must be set before commissioning; allow ~2 h acclimation after cold/warm moves; expansion vessel must remain unlocked for externally closed applications and locked for open bath applications.
Integration & Operating Notes
The water-cooled refrigeration circuit is designed for inlet water between 5 °C and 20 °C; warmer water reduces cooling power and raises required flow, while pressure below 1 bar differential or above 6 bar risks inadequate flow or overpressure. Insulate cooling water pipes when water temperature falls below the condensation point to prevent moisture damage.
The unit draws up to 16 A and requires a 16 A fuse; voltage tolerance ±5% and frequency tolerance ±2% must not be combined in the unfavorable direction. At 230 V the maximum current is just under 16 A and the unit can be limited to 13 A or 10 A at first start-up, which affects upstream breaker sizing. Hard-wired connection must be performed by an electrician.
For externally closed applications the expansion vessel must remain unlocked; locking it causes overpressure damage. For externally open (bath) applications the expansion vessel must be locked to prevent fluid loss. This valve setting must be verified before filling and venting.
The variable-speed pressure pump is limited to a maximum permissible kinematic viscosity of 50 mm²/s; selecting a fluid that is too viscous at low temperatures will overload the pump and compromise circulation. For FL-rated fluids, the fire point must be ≥65 °C, and ethanol use requires supervised operation.
The unit must be transported and stored upright only, moved only when emptied, and set on leveling feet or parking brakes before commissioning. A condensation water/heat transfer fluid drip tray must be installed below the unit. Installation inside an ATEX zone is prohibited; explosion protection requires an Ex px cabinet.
In Process Safety mode the controller releases maximum cooling capacity at the overtemperature trigger instead of switching off all poles; this mode requires the compressor to be set to Permanently ON. Power Failure Auto can resume temperature control automatically after an outage if it was active, so choose the restart behavior according to whether unattended restart is acceptable.
The cooling water supply should be shut off even for short downtimes such as overnight to prevent water damage and unnecessary load on the circuit. Hoses and connections must be periodically checked for leaks and quality and replaced as needed. Refrigerant-circuit work is restricted to approved refrigeration/air-conditioning contractors.
Selection & Integration Q&A
Choose the 410w when a cooling water circuit (5–20 °C, 1–6 bar differential) is available and air-cooled heat rejection is impractical. The 'Dynamisches' modification suits rapid temperature changes; the 410(w) has no plate heat exchanger, so the usual fatigue warning for rapid cycling does not apply in the same way.
Cooling water: G1/2 male, inlet 5–20 °C, minimum 1 bar differential, maximum 6 bar. Pump circuit: M24x1.5 male. Insulate cooling water pipes if water is below the condensation point. Ensure heat transfer fluid viscosity ≤50 mm²/s.
Supply 220–240 V 1~/2~ 50/60 Hz, max 16 A, 16 A fuse; voltage ±5% and frequency ±2% must not combine unfavorably. Ambient 5–40 °C, IP20, EMC Class A Group 1. Hard-wired connection by an electrician; allow ~2 h acclimation after cold/warm moves.
Cooling water hoses/fittings (G1/2 male) and pump hoses/adapters (M24x1.5 male) are not included. A drip tray must be provided by the installer. Periodic checks of hoses and connections are required; refrigerant work only by approved contractors. Optional gas detection sensor and processing unit are stand-alone accessories.