Peter Huber Kältemaschinenbau Unistat T305 Dynamisches

Peter Huber Kältemaschinenbau Unistat T305 Dynamisches — Temperature control units.

The Unistat T305 Dynamisches is a non-refrigerated, dynamically controlled temperature unit with a 65–300 °C working range and 2.8 L expansion volume. It is distinguished from the T305 HT and T305w HT siblings by its dynamic pressure control and absence of active cooling, requiring application heat loss or water cooling to reach 65 °C. It serves as a precise, pressure-controlled circulator for closed or open external circuits in industrial processes.

Peter Huber Kältemaschinenbau Unistate Bis 425C Series — Model Comparison & Selection

Model Display resolutionPower supplyPump connectionPump typeTemperature range, °CExpansion volume, m³
Unistat T305 Dynamisches Current 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable 65–300—
Chili Dynamisches 0.01200-240V 1~/2~ 50/60HzM24x1,5 AG malePressure pump, variable 65–3000.002
Unistat T305 HT Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable 65–300—
Unistat T305w HT Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable 65–300—
Unistat T320 Dynamisches 0, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st65–3000.024
Unistat T320w HT Dynamisches 0, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st65–3000.024
Unistat T330 HT Dynamisches 0, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st65–3000.024
Unistat T330w HT Dynamisches 0, 01 K400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st65–3000.024
Unistat T340 Dynamisches 0, 01460V 3~ 60HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T340 HT Dynamisches 0, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T345 Dynamisches 0, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T345 HT Dynamisches 0, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T350 HT Dynamisches 0, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T402 Dynamisches 0, 01400V 3~N 50/60HzM24x1, 5 AG malePressure pump, variable 80–425—
Unistat TR401 Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 (Flow) M24x1, 5Pressure pump, variable 50–4000.005
Unistat TR402 Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 (Flow) M24x1, 5Pressure pump, variable 80–425—

Full specifications

Core Performance & Functional Specifications
Control output POKO ONE
Pump connection M24x1, 5 AG male
Pump speed adjustable Yes
Pump type Pressure pump, variable speed
External sensor connection Pt100
Temperature range 65...300 °C
Dimensions, Weight & Mechanical Design
Weight 37 kg
Electrical, Power & Drive Specifications
Power supply 200-240V 1~/2~ 50/60Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and Series Positioning

The Unistat T305 Dynamisches is a pressure-pump-based temperature control unit with a variable-speed, soft-start pump and integrated pressure control. Unlike the Unistat T305 HT and T305w HT siblings, this model lacks active refrigeration and relies on application heat loss or external water cooling to achieve its lower temperature limit of 65 °C. The expansion tank is not thermoregulated and is lockable for open bath configurations. The unit is rated IP20 and weighs 37 kg.

Behavior Under Load and Media

Installation, Control and Operation

Integration & Operating Notes

Process Limits
Minimum temperature depends on application heat loss

Without active refrigeration, reaching 65 °C requires either sufficient heat loss from the application (0.1–2.8 kW depending on unit) or suitable water cooling. Verify the application's heat loss falls within this range before commissioning.

Unistat T305 Technical Data, page 2
Mounting
Install only in upright position with leveling feet engaged

Transport and store the unit upright only; horizontal transport damages the compressor. Before commissioning, unscrew/activate leveling feet and engage parking brakes if casters are present. Casters are not for transport and carry only 25% of total mass symmetrically.

Unistat Operation Manual, page 22
Electrical
Hard-wired connection must be performed by an electrician

Supply voltage and frequency must match the rating plate. At 230 V the maximum current is just under 16 A; the unit can be limited to 13 A or 10 A at first start-up. Voltage tolerance ±10% per EN60034-1 applies only if frequency deviation does not run in the opposite direction.

Unistat Operation Manual, page 37; Unistat T305 Technical Data, page 2
Process Limits
Expansion vessel lock state depends on open or closed application

For externally closed applications the expansion vessel must remain unlocked; locking it causes overpressure damage. For externally open applications the expansion vessel must be locked to prevent fluid loss.

Unistat Operation Manual, page 33
Maintenance
Shut off cooling water during downtimes

Shut off the cooling water supply even for shorter downtimes such as overnight to prevent water damage and unnecessary load on the cooling water circuit. Periodically check hoses and connections for leaks and quality, and replace as needed.

Unistat Operation Manual, page 29
Control & I/O
Power Failure Auto resumes control automatically after outage

If temperature control is active during a power outage, the process will automatically continue after the power outage. Assess whether unattended restart is acceptable for the process; if not, disable this function in the controller settings.

Unistat Operation Manual, page 20
Mounting
Allow 2 hours acclimation after temperature changes

When moving the unit between cold and warm environments, allow about 2 hours acclimation before power-on to prevent condensation-induced electrical faults.

Unistat Operation Manual, page 29

Selection & Integration Q&A

When should I choose the Unistat T305 Dynamisches over the T305 HT or T305w HT siblings?

Choose the T305 Dynamisches for applications where heat loss or water cooling can handle the lower end (65 °C) and dynamic pressure control is needed. The HT variants offer active refrigeration for lower temperatures; select those if sub-ambient cooling is required.

What are the requirements for connecting the pump to my external circuit?

The pump connection is M24x1.5 male with bore shape Y (60°) per DIN 3863. Use hoses with ball socket per DIN 3863 and sleeve nut per DIN 3870. Ensure the circuit pressure drop does not exceed the pump's 0.9 bar maximum delivery pressure at 45 l/min.

What are the ambient and duty-cycle limits for installing this unit?

Install in ambient temperature 5–40 °C, IP20 environment, not in ATEX zones. Transport and store upright only. Allow 2 hours acclimation after temperature changes. For air-cooled operation, application heat loss must be 0.1–2.8 kW to reach 65 °C.

What maintenance is required and what accessories are not included?

Periodically check hoses and connections for leaks and quality; replace as needed. Shut off cooling water during downtimes. Power cable and country-specific plug must be specified at order; external pressure sensor is optional. No active refrigeration means no compressor maintenance.