Peter Huber Kältemaschinenbau Unistat T340 HT Dynamisches

Peter Huber Kältemaschinenbau Unistat T340 HT Dynamisches — Temperature control units.

The Unistat T340 HT Dynamisches is a high-temperature dynamic temperature control unit with a 65–300 °C working range, 48 kW heating and 300 kW cooling at 30 °C. It is the largest high-temperature modification in the Unistat T305/T340 series, distinguished by its 56 l expansion volume and 200 l/min pump. It serves closed-loop process temperature control in industrial plants requiring three-phase 400 V supply.

Peter Huber Kältemaschinenbau Unistate Bis 425C Series — Model Comparison & Selection

Model Display resolutionPower supplyPump connectionPump typeTemperature range, °CExpansion volume, m³
Unistat T340 HT Dynamisches Current 0, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Chili Dynamisches 0.01200-240V 1~/2~ 50/60HzM24x1,5 AG malePressure pump, variable 65–3000.002
Unistat T305 Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable 65–300—
Unistat T305 HT Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable 65–300—
Unistat T305w HT Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable 65–300—
Unistat T320 Dynamisches 0, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st65–3000.024
Unistat T320w HT Dynamisches 0, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st65–3000.024
Unistat T330 HT Dynamisches 0, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st65–3000.024
Unistat T330w HT Dynamisches 0, 01 K400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st65–3000.024
Unistat T340 Dynamisches 0, 01460V 3~ 60HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T345 Dynamisches 0, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T345 HT Dynamisches 0, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T350 HT Dynamisches 0, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T402 Dynamisches 0, 01400V 3~N 50/60HzM24x1, 5 AG malePressure pump, variable 80–425—
Unistat TR401 Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 (Flow) M24x1, 5Pressure pump, variable 50–4000.005
Unistat TR402 Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 (Flow) M24x1, 5Pressure pump, variable 80–425—

Full specifications

Core Performance & Functional Specifications
Alarm output optic, acoustic, relay
Control output POKO ONE
Expansion volume 56 l
Heating capacity 48 kW
Pump connection M38x1, 5 AG male
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range 65...300 °C
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling capacity 300 kW (at 30 °C)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Control and process behavior

Electrical and installation

Integration & Operating Notes

Electrical
Supply tolerance envelope is linked, not independent

Voltage and frequency tolerances must be evaluated together; combinations such as -5% voltage with +2% frequency are explicitly not allowed. Verify the actual site supply against the combined envelope before commissioning, because a supply that looks acceptable on each parameter alone can still be outside the permitted operating window.

Unistat T340 HT Technical Data, page 2
Electrical
Final three-phase connection is site-supplied

The unit is delivered with cable but without a plug for three-phase supplies below 63 A. The customer must provide the final connection, and the hard-wired power connection must be carried out by an electrician with supply voltage and frequency matched to the rating plate.

Unistat T340 HT Technical Data, page 2; Unistat Operation Manual, page 37
Electrical
Leakage current above 3.5 mA affects RCD choice

Protective earthing and residual current device selection must account for a leakage current exceeding 3.5 mA. Standard 30 mA RCDs may not be appropriate; the earthing design and RCD rating must be engineered for this leakage level.

Unistat T340 HT Technical Data, page 2
Mounting
Upright transport only; empty before moving

The unit must be transported and stored upright only, because horizontal transport damages the compressor. It may only be moved when emptied of thermal fluid. Casters, if present, are not suitable for transport and carry only 25% of total mass symmetrically; use an industrial truck. Leveling feet must be activated and parking brakes engaged before commissioning.

Unistat Operation Manual, page 22
Mounting
Expansion vessel lock state depends on circuit type

For externally closed applications the expansion vessel must remain unlocked; locking it causes overpressure damage. For externally open applications it must be locked to prevent fluid loss. Confirm the circuit type before setting the vessel lock, as the wrong state damages the application or loses fluid.

Unistat Operation Manual, page 33
Process Limits
Fluid viscosity must stay below 50 mm²/s

The maximum permissible kinematic viscosity is 50 mm²/s. Thermofluid selection must be checked against this limit at the lowest operating temperature, because a fluid too viscous for the pump at 65 °C will not circulate correctly.

Unistat T340 HT Technical Data, page 1
Process Limits
External loop pressure drop must fit 5.5 bar pump head

The pump delivers up to 200 l/min at 5.5 bar. The pressure drop of the customer's external circuit must be checked against this head at the required flow; a loop with excessive pressure drop will not achieve the design flow rate.

Unistat T340 HT Technical Data, page 1
Control & I/O
Process Safety mode requires compressor permanently on

When Process Safety mode is used for exothermic process protection, the compressor must be set to Permanently ON so that cooling remains active at the overtemperature trigger. Without this setting the safety function will not have cooling available when it trips.

Unistat Operation Manual, page 19
Control & I/O
Power Failure Auto restarts control automatically

If temperature control is active during a power outage, Power Failure Auto resumes the process automatically after power returns. Assess whether unattended automatic restart is acceptable for the process before enabling this behavior.

Unistat Operation Manual, page 20
Maintenance
Shut off cooling water even for overnight stops

The cooling water supply should be shut off even for short downtimes such as overnight. This prevents water damage and unnecessary load on the cooling water circuit during idle periods.

Unistat Operation Manual, page 29
Maintenance
Hoses and connections need periodic leak and quality checks

Hoses and hose connections must be periodically checked for leaks and quality and replaced as required. This is the preventive maintenance scope for the fluid circuit; degraded hoses are a leakage risk at up to 300 °C.

Unistat Operation Manual, page 29
Environment
No ATEX zone installation; IP20 requires dry location

The unit must not be installed or started inside an ATEX zone; explosion protection would require an Ex px cabinet. With IP20 protection it must be placed in a dry, protected location away from water ingress, and ambient temperature must stay between 5 and 40 °C.

Unistat Operation Manual, page 15; Unistat T340 HT Technical Data, page 1
Environment
Allow 2 hours acclimation before first power-on

When moving the unit between cold and warm environments, allow about 2 hours acclimation before turning it on. Starting immediately risks condensation-induced electrical faults.

Unistat Operation Manual, page 29

Selection & Integration Q&A

When should I choose the T340 HT over the T305 HT or T305w HT siblings?

Choose the T340 HT when the process needs the larger 56 l expansion volume and the higher 48 kW heating / 300 kW cooling capacity at 30 °C. The T305 and T305w HT are smaller high-temperature variants; select them when the T340's capacity and expansion volume exceed the duty. All share the 65–300 °C range.

What must the external circuit and its connections be rated for?

The pump connections are M38x1.5 AG male, and the external loop pressure drop must be checked against the pump's 5.5 bar head at up to 200 l/min. The thermofluid must stay below 50 mm²/s kinematic viscosity at the lowest operating temperature.

What site conditions and duty limits apply to installation?

Install in a dry, protected location (IP20, no ATEX zone) with ambient 5–40 °C. Transport and store upright only, move only when emptied, and allow 2 hours acclimation before first power-on. The 400 V 3~ 50 Hz supply must meet the linked voltage/frequency tolerance envelope.

What maintenance and delivery-scope items should be planned for?

Periodically check hoses and connections for leaks and quality and replace as needed; shut off cooling water even overnight. The unit ships with cable but without a plug for three-phase connection below 63 A, so the final connection hardware and an electrician are site-supplied.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unistat T340 HT Technical Data