Peter Huber Kältemaschinenbau Unistat T345 Dynamisches
Peter Huber Kältemaschinenbau Unistat T345 Dynamisches — Temperature control units.
The Unistat T345 Dynamisches is a dynamically controlled, magnetically coupled stainless-steel pump temperature control unit with 72 kW heating and a 65–300 °C working range, positioned above the T305 siblings in heating capacity. It serves high-temperature process loops where the 56 l unregulated expansion tank and 200 l/min pump flow support closed or open external circuits.
Peter Huber Kältemaschinenbau Unistate Bis 425C Series — Model Comparison & Selection
| Model | Display resolution | Power supply | Pump connection | Pump type | Temperature range, °C | Expansion volume, m³ |
|---|---|---|---|---|---|---|
| Unistat T345 Dynamisches Current | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st | 65–300 | 0.056 |
| Chili Dynamisches | 0.01 | 200-240V 1~/2~ 50/60Hz | M24x1,5 AG male | Pressure pump, variable | 65–300 | 0.002 |
| Unistat T305 Dynamisches | 0, 01 | 200-240V 1~/2~ 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | 65–300 | — |
| Unistat T305 HT Dynamisches | 0, 01 | 200-240V 1~/2~ 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | 65–300 | — |
| Unistat T305w HT Dynamisches | 0, 01 | 200-240V 1~/2~ 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | 65–300 | — |
| Unistat T320 Dynamisches | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st | 65–300 | 0.024 |
| Unistat T320w HT Dynamisches | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | 65–300 | 0.024 |
| Unistat T330 HT Dynamisches | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | 65–300 | 0.024 |
| Unistat T330w HT Dynamisches | 0, 01 K | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | 65–300 | 0.024 |
| Unistat T340 Dynamisches | 0, 01 | 460V 3~ 60Hz | M38x1, 5 AG male | magnetically coupled, st | 65–300 | 0.056 |
| Unistat T340 HT Dynamisches | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st | 65–300 | 0.056 |
| Unistat T345 HT Dynamisches | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st | 65–300 | 0.056 |
| Unistat T350 HT Dynamisches | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st | 65–300 | 0.056 |
| Unistat T402 Dynamisches | 0, 01 | 400V 3~N 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | 80–425 | — |
| Unistat TR401 Dynamisches | 0, 01 | 200-240V 1~/2~ 50/60Hz | M24x1, 5 (Flow) M24x1, 5 | Pressure pump, variable | 50–400 | 0.005 |
| Unistat TR402 Dynamisches | 0, 01 | 200-240V 1~/2~ 50/60Hz | M24x1, 5 (Flow) M24x1, 5 | Pressure pump, variable | 80–425 | — |
Full specifications
| Control output | POKO ONE |
| Expansion volume | 56 l |
| Heating capacity | 72 kW |
| Pump connection | M38x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | 65...300 °C |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- The T345 Dynamisches is the higher-heating-capacity member of the Unistat T305/T345 dynamic series; the T305, T305 HT and T305w HT siblings share the same series architecture but differ in heating capacity and temperature-range configuration.
- Circulation is handled by a magnetically coupled stainless-steel pump with adjustable speed, rated up to 200 l/min and 5.5 bar, connected via M38x1.5 male threads.
- The 56 l expansion tank is not thermoregulated and is lockable for open-bath operation; it must remain unlocked for externally closed applications.
- Control is via the Pilot ONE controller with POKO ONE output, 0.01 K display resolution, and Ethernet, USB (host and device) and RS232 interfaces; an external Pt100 sensor connection is provided.
Behavior and process limits
- The unit has no active refrigeration; the 65 °C lower limit depends on application heat loss or suitable water cooling, so it is not a chiller and cannot be specified for sub-ambient cooling without an external heat sink.
- Maximum permissible kinematic viscosity of the thermofluid is 50 mm²/s, which constrains fluid selection at low operating temperatures.
- Adjustable overtemperature protection per DIN 12876 is included and must be set for the application during commissioning.
- Process Safety mode requires the compressor to be set to Permanently ON to keep cooling active at the OT trigger; Power Failure Auto resumes temperature control automatically after a power outage if it was active.
Installation and operation
- The unit requires a 400 V 3~ 50 Hz three-phase supply; three-phase devices drawing less than 63 A are supplied with cable but no plug, so the site must provide a suitable plug and connection.
- Leakage current exceeds 3.5 mA, requiring appropriate protective earthing and RCD coordination; EMC classification is Class A Group 1 for industrial environments.
- IP20 protection restricts installation to dry indoor locations; ambient temperature must stay between 5 °C and 40 °C.
- The unit must be transported and stored upright only, moved only when emptied, and allowed about 2 hours acclimation when changing between cold and warm environments before power-on.
Integration & Operating Notes
The unit cannot reach its 65 °C lower limit on its own; the minimum depends on heat loss in the application or on suitable water cooling. Specifying the T345 for cooling duties below ambient without an external heat sink or water cooling will not meet the process requirement.
Protective earthing and residual-current device selection must account for a leakage current exceeding 3.5 mA; type-B or higher RCD coordination may be required. The supply is 400 V 3~ 50 Hz, and voltage/frequency tolerances must be respected simultaneously, with combinations such as -5% voltage and +2% frequency not allowed.
The site must provide a suitable plug and connection for the supplied cable. Hard-wired power connection must be carried out by an electrician, and supply voltage and frequency must match the rating plate before commissioning.
For externally closed applications the expansion vessel must remain unlocked; locking it will damage the application due to overpressure. For externally open applications the expansion vessel must be locked to prevent fluid loss. The 56 l tank is not thermoregulated, which affects fluid volume and expansion allowance in closed systems.
Casters are not suitable for transport and carry only 25% of total mass symmetrically; use an industrial truck for relocation. Before commissioning, activate parking brakes and unscrew or activate leveling feet to ensure stable, level installation and prevent creeping during operation.
The maximum permissible kinematic viscosity of the thermofluid is 50 mm²/s, which limits fluid choice at low operating temperatures. Select a fluid whose viscosity at the lowest expected temperature remains within this limit to avoid pump overload.
When Process Safety mode is used for exothermic process protection, the compressor must be configured to Permanently ON so that cooling remains active at the overtemperature trigger. Power Failure Auto resumes temperature control automatically after a power outage if it was active, which affects unattended operation risk assessment.
The cooling water supply should be shut off even for short downtimes such as overnight to prevent water damage and unnecessary load on the cooling water circuit. Hoses and connections must be periodically checked for leaks and quality and replaced as needed.
Selection & Integration Q&A
Pick the T345 when the process needs the higher 72 kW heating capacity and a 65–300 °C range; the T305, T305 HT and T305w HT differ in heating capacity and temperature-range configuration. The T345 has no active refrigeration, so sub-ambient duties require a chiller or water cooling.
A 400 V 3~ 50 Hz supply with voltage and frequency tolerances respected simultaneously; three-phase units under 63 A come with cable but no plug, so the site supplies the plug. Leakage current exceeds 3.5 mA, so earthing and RCD selection must be coordinated, and connection must be done by an electrician.
Install indoors and dry (IP20) with ambient 5–40 °C; transport and store upright only and move only when emptied. Allow about 2 hours acclimation when changing between cold and warm environments before power-on, and set leveling feet and parking brakes before commissioning.
The base delivery excludes the power plug for three-phase units under 63 A and the mating M38x1.5 hoses/fittings. Periodically check hoses and connections for leaks and quality and replace as needed; shut off cooling water even overnight. Set the adjustable DIN 12876 overtemperature protection during commissioning.