No photo available yet

Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unistat T305 HT Dynamisches

Peter Huber Kältemaschinenbau Unistat T305 HT Dynamisches — Temperature control units.

The Unistat T305 HT Dynamisches is a high-temperature pressure temperature control unit with a variable-speed pump and Pilot ONE controller, covering 65–300 °C. It is the HT modification of the T305 series, distinguished by its high-temperature range and dynamic pressure pump. It serves closed-loop tempering of exothermic or high-temperature processes where sub-ambient cooling is not required.

Peter Huber Kältemaschinenbau Unistate Bis 425C Series — Model Comparison & Selection

Model Display resolutionPower supplyPump connectionPump typeTemperature range, °CExpansion volume, m³
Unistat T305 HT Dynamisches Current 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable 65–300—
Chili Dynamisches 0.01200-240V 1~/2~ 50/60HzM24x1,5 AG malePressure pump, variable 65–3000.002
Unistat T305 Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable 65–300—
Unistat T305w HT Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable 65–300—
Unistat T320 Dynamisches 0, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st65–3000.024
Unistat T320w HT Dynamisches 0, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st65–3000.024
Unistat T330 HT Dynamisches 0, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st65–3000.024
Unistat T330w HT Dynamisches 0, 01 K400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st65–3000.024
Unistat T340 Dynamisches 0, 01460V 3~ 60HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T340 HT Dynamisches 0, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T345 Dynamisches 0, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T345 HT Dynamisches 0, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T350 HT Dynamisches 0, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T402 Dynamisches 0, 01400V 3~N 50/60HzM24x1, 5 AG malePressure pump, variable 80–425—
Unistat TR401 Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 (Flow) M24x1, 5Pressure pump, variable 50–4000.005
Unistat TR402 Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 (Flow) M24x1, 5Pressure pump, variable 80–425—

Full specifications

Core Performance & Functional Specifications
Control output POKO ONE
Maximum delivery flow rate 45 l/min (0, 9 bar)
Pump connection M24x1, 5 AG male
Pump speed adjustable Yes
Pump type Pressure pump, variable speed
External sensor connection Pt100
Temperature range 65...300 °C
Dimensions, Weight & Mechanical Design
Weight 42 kg
Electrical, Power & Drive Specifications
Power supply 200-240V 1~/2~ 50/60Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

The Unistat T305 HT Dynamisches is a high-temperature variant within the Unistat T305 series. Unlike the standard T305 Dynamisches or the T305w HT Dynamisches, this model is configured for a 65–300 °C operating window and uses a variable-speed pressure pump (Pilot ONE controller, POKO ONE control output). It is not intended for sub-ambient work; the Chili Dynamisches and T320 Dynamisches siblings cover different temperature ranges. The unit weighs 42 kg and is rated IP20 for dry indoor installation.

Behavior under load and media

Cooling capacity is strongly temperature-dependent: 3.2 kW at 300 °C falling to 0.6 kW at 100 °C, so heat-removal sizing must use the actual process temperature. The pump delivers up to 45 l/min at 0.9 bar; external loop pressure drop must be checked against this curve. The maximum permissible kinematic viscosity is 50 mm²/s, so the thermofluid must be selected to stay below this limit at the lowest operating temperature. Minimum filling capacity is 3.5 l and the expansion tank holds 4.5 l; the expansion vessel must be unlocked for externally closed applications and locked for open-bath use.

Installation, control and operation

Electrical supply is 200–240 V, 1~ or 2~, 50/60 Hz, with a maximum current just under 16 A at 230 V; the unit can be limited to 13 A or 10 A at first start-up. External branch circuit protection to UL 489 is required, and hard-wired connection must be performed by an electrician. The pump connection is M24x1.5 male with Y-bore to DIN 3863; hoses must use ball sockets to DIN 3863 and sleeve nuts to DIN 3870. The Pilot ONE controller provides Ethernet, USB (host and device) and RS232 interfaces, with 0.01 K display resolution and Pt100 external sensor input. Process Safety mode requires the compressor set to Permanently ON; Power Failure Auto resumes control after an outage if it was active. The unit is classified III / FL to DIN 12876 and pressure equipment category Art. 4.3 PED.

Integration & Operating Notes

Mounting
Transport and store only upright; drain before moving

Horizontal transport damages the compressor, and moving a filled unit risks thermal fluid overflow. Plan rigging and floor layout so the unit stays upright from delivery to commissioning, and empty the fluid circuit before any relocation.

Operation Manual, Section Commissioning (page 22
Electrical
Provide UL 489 branch protection and state plug type on order

External branch circuit protection according to UL 489 is mandatory ahead of the unit. Single- and two-phase units are delivered with a country-specific plug only if specified at order, so the purchase order must state the plug type to avoid site rework.

Datasheet, page 2
Process Limits
Expansion vessel lock state depends on open vs closed application

For externally closed applications the expansion vessel must remain unlocked; locking it causes overpressure damage. For externally open applications it must be locked to prevent fluid loss. Confirm the application type before commissioning and set the vessel accordingly.

Operation Manual, Section Commissioning (page 33
Control & I/O
Process Safety mode requires compressor set to Permanently ON

When Process Safety mode is enabled, the compressor must be configured to Permanently ON so that cooling remains active at the over-temperature trigger. Verify this setting in the Pilot ONE controller before relying on the safety function.

Operation Manual, Section Normal Operation (page 19
Maintenance
Shut off cooling water even for overnight downtime

The cooling water supply should be shut off even for short downtimes such as overnight to prevent water damage and unnecessary load on the cooling water circuit. Include this step in the daily shutdown procedure.

Operation Manual, Section Commissioning (page 29
Maintenance
Periodically inspect hoses and connections for leaks and quality

Hoses and hose connections must be checked periodically for leaks and quality and replaced as required. Establish a preventive maintenance interval based on duty cycle and fluid compatibility to avoid sudden failure in high-temperature service.

Operation Manual, Section Commissioning (page 29
Environment
Install indoors only; IP20 and 5–40 °C ambient limit

The IP20 rating excludes washdown and outdoor siting. Ambient temperature must stay between 5 and 40 °C, so room HVAC and placement away from heat sources must be considered. Allow about 2 hours acclimation when moving between cold and warm environments before power-on.

Datasheet, page 1; Operation Manual, page 29

Selection & Integration Q&A

When should I choose the Unistat T305 HT Dynamisches instead of the standard T305 Dynamisches or the T305w HT Dynamisches?

Choose the T305 HT Dynamisches for closed-loop processes requiring 65–300 °C with a variable-speed pressure pump. The standard T305 Dynamisches covers a different temperature range; the T305w HT is a water-cooled variant. Select the HT only when sub-ambient cooling is not needed.

What are the requirements for hoses and fittings mating to the pump connection?

The pump connection is M24x1.5 male with Y-bore to DIN 3863. Use flexible tempering hoses with ball sockets to DIN 3863 and sleeve nuts to DIN 3870. Non-matching fittings will not seal, so verify the standard before procuring hoses.

What are the duty-cycle and environment limits for installing this unit?

Install indoors in a dry location (IP20) with ambient 5–40 °C. Transport and store upright only; drain before moving. Allow 2 h acclimation after temperature changes. Do not install in an ATEX zone. Leveling feet or parking brakes must be engaged before operation.

What maintenance and accessories are required beyond the base delivery?

The base delivery does not include hoses, fittings, or a country-specific plug unless specified at order. Periodically inspect hoses and connections for leaks and replace as needed. Shut off cooling water during downtimes. External UL 489 branch protection must be provided by the installer.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unistat T305 HT Technical Data
Datasheet ·17,8 КБ ·2 pp.
View PDF
PDF
Unistat Operation Manual
Manual ·3,7 МБ ·102 pp.
View PDF