Peter Huber Kältemaschinenbau Unistat T305w HT Dynamisches

Peter Huber Kältemaschinenbau Unistat T305w HT Dynamisches — Temperature control units.

The Unistat T305w HT is a water-cooled, dynamically temperature-controlled pressure-pump unit with a 65–300 °C range (15–300 °C with water cooling), positioned above the air-cooled T305 HT and below the T320 in the Unistat series. Its variable-speed pressure pump delivers up to 45 l/min at 0.9 bar, making it suited to closed-loop tempering of reactors and process equipment requiring high-temperature stability.

Peter Huber Kältemaschinenbau Unistate Bis 425C Series — Model Comparison & Selection

Model Display resolutionPower supplyPump connectionPump typeTemperature range, °CExpansion volume, m³
Unistat T305w HT Dynamisches Current 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable 65–300—
Chili Dynamisches 0.01200-240V 1~/2~ 50/60HzM24x1,5 AG malePressure pump, variable 65–3000.002
Unistat T305 Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable 65–300—
Unistat T305 HT Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable 65–300—
Unistat T320 Dynamisches 0, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st65–3000.024
Unistat T320w HT Dynamisches 0, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st65–3000.024
Unistat T330 HT Dynamisches 0, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st65–3000.024
Unistat T330w HT Dynamisches 0, 01 K400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st65–3000.024
Unistat T340 Dynamisches 0, 01460V 3~ 60HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T340 HT Dynamisches 0, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T345 Dynamisches 0, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T345 HT Dynamisches 0, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T350 HT Dynamisches 0, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T402 Dynamisches 0, 01400V 3~N 50/60HzM24x1, 5 AG malePressure pump, variable 80–425—
Unistat TR401 Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 (Flow) M24x1, 5Pressure pump, variable 50–4000.005
Unistat TR402 Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 (Flow) M24x1, 5Pressure pump, variable 80–425—

Full specifications

Core Performance & Functional Specifications
Control output POKO ONE (potential-free contact)
Maximum delivery flow rate 45 l/min (pressure pump max.; also 0, 9 bar)
Pump connection M24x1, 5 AG male
Pump speed adjustable Yes
Pump type Pressure pump, variable speed
External sensor connection Pt100
Temperature range 65...300 °C
with water cooling: 15...300 °C
Dimensions, Weight & Mechanical Design
Weight 42 kg
Electrical, Power & Drive Specifications
Current consumption 15 A (max.)
Power supply 200-240V 1~/2~ 50/60Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s (max. permissible)
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C (permissible)
Cooling water connection G1/2 male
Cooling water differential pressure 1 bar (min.)
Cooling Water Pressure 6 bar (max.)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Behavior and media limits

Installation, connection and operation

Integration & Operating Notes

Cooling & Media
Verify site cooling water before specifying the unit

The unit requires cooling water inlet temperature between 5 °C and 20 °C and at least 1 bar differential pressure. Outside this window cooling power drops and flow demand rises, so the site water supply must be confirmed during selection; if inlet temperature falls below the condensation point, the internal cooling-water pipes must be insulated.

Unistat T305w HT Datasheet, page 2
Electrical
Confirm supply quality and branch protection before hook-up

Voltage tolerance is ±10% only if the frequency tolerance does not run in the opposite direction, so supply quality must be verified. External branch circuit protection according to UL 489 is required. Single/two-phase units are delivered with a country-specific plug that must be specified at order; hard-wired connection must be performed by an electrician.

Unistat T305w HT Datasheet, page 2; Unistat Operation Manual, page 37
Mounting
Transport upright, empty, and allow 2 h acclimation

Only transport and store the unit upright and only when emptied; horizontal transport damages the compressor. Casters are not for transport and carry only 25% of total mass symmetrically. Before commissioning, activate parking brakes and unscrew leveling feet. After moving between cold and warm environments, allow about 2 hours acclimation before power-on to prevent condensation-induced faults.

Unistat Operation Manual, pages 22 and 29
Process Limits
Set expansion vessel lock according to application type

For externally closed applications the expansion vessel must remain unlocked; locking it causes overpressure damage to the application. For externally open applications the expansion vessel must be locked to prevent fluid loss. This setting must be confirmed at commissioning and whenever the application circuit type changes.

Unistat Operation Manual, page 33
Control & I/O
Process Safety mode needs compressor set to Permanently ON

If Process Safety mode is used for exothermic process protection, the compressor must be configured to Permanently ON so cooling remains active at the over-temperature trigger. Power Failure Auto resumes temperature control automatically after a power outage if it was active, which affects unattended operation risk assessment.

Unistat Operation Manual, pages 19–20
Maintenance
Shut off cooling water during short downtimes

Shut off the cooling water supply even for short downtimes such as overnight to prevent water damage and unnecessary load on the circuit. Periodically check hoses and hose connections for leaks and quality and replace as required.

Unistat Operation Manual, page 29

Selection & Integration Q&A

When should I choose the T305w HT over the air-cooled T305 HT or the larger T320?

Pick the T305w HT when you need the range extended down to 15 °C via water cooling and have a suitable 5–20 °C water supply; the air-cooled T305 HT stops at 65 °C. Choose the T320 only if its larger capacity is required.

What mating fittings and site services does the cooling-water and pump circuit require?

Cooling-water connection is G1/2 male with 5–20 °C inlet and ≥1 bar differential pressure; pump connection is M24x1.5 AG male with bore shape Y (60°) per DIN 3863, requiring ball socket per DIN 3863 and sleeve nut per DIN 3870 on hoses.

What installation environment and duty-cycle limits apply from the manual?

Install within 5–40 °C ambient, IP20, free of water ingress, and never inside an ATEX zone. Transport upright and empty; allow 2 h acclimation after temperature changes; activate leveling feet and brakes before commissioning.

What maintenance and accessories are required beyond base delivery?

Base delivery includes cable and country-specific plug for single/two-phase (specify at order); external UL 489 branch protection and mating hoses/fittings are customer-supplied. Shut off cooling water for short downtimes and periodically inspect/replace hoses and connections.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unistat T305w HT Temperature Control System Datasheet
Datasheet ·18,4 КБ ·2 pp.
View PDF
PDF
Unistat Operation Manual
Manual ·3,7 МБ ·102 pp.
View PDF