Peter Huber Kältemaschinenbau Unistat T305 Dynamisches
Peter Huber Kältemaschinenbau Unistat T305 Dynamisches — Temperature control units.
The Unistat T305 Dynamisches is a non-refrigerated, dynamically controlled temperature unit with a 65–300 °C working range and 2.8 L expansion volume. It is distinguished from the T305 HT and T305w HT siblings by its dynamic pressure control and absence of active cooling, requiring application heat loss or water cooling to reach 65 °C. It serves as a precise, pressure-controlled circulator for closed or open external circuits in industrial processes.
Peter Huber Kältemaschinenbau Unistate Bis 425C Series — Model Comparison & Selection
| Model | Display resolution | Power supply | Pump connection | Pump type | Temperature range, °C | Expansion volume, m³ |
|---|---|---|---|---|---|---|
| Unistat T305 Dynamisches Current | 0, 01 | 200-240V 1~/2~ 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | 65–300 | — |
| Chili Dynamisches | 0.01 | 200-240V 1~/2~ 50/60Hz | M24x1,5 AG male | Pressure pump, variable | 65–300 | 0.002 |
| Unistat T305 HT Dynamisches | 0, 01 | 200-240V 1~/2~ 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | 65–300 | — |
| Unistat T305w HT Dynamisches | 0, 01 | 200-240V 1~/2~ 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | 65–300 | — |
| Unistat T320 Dynamisches | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st | 65–300 | 0.024 |
| Unistat T320w HT Dynamisches | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | 65–300 | 0.024 |
| Unistat T330 HT Dynamisches | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | 65–300 | 0.024 |
| Unistat T330w HT Dynamisches | 0, 01 K | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | 65–300 | 0.024 |
| Unistat T340 Dynamisches | 0, 01 | 460V 3~ 60Hz | M38x1, 5 AG male | magnetically coupled, st | 65–300 | 0.056 |
| Unistat T340 HT Dynamisches | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st | 65–300 | 0.056 |
| Unistat T345 Dynamisches | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st | 65–300 | 0.056 |
| Unistat T345 HT Dynamisches | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st | 65–300 | 0.056 |
| Unistat T350 HT Dynamisches | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st | 65–300 | 0.056 |
| Unistat T402 Dynamisches | 0, 01 | 400V 3~N 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | 80–425 | — |
| Unistat TR401 Dynamisches | 0, 01 | 200-240V 1~/2~ 50/60Hz | M24x1, 5 (Flow) M24x1, 5 | Pressure pump, variable | 50–400 | 0.005 |
| Unistat TR402 Dynamisches | 0, 01 | 200-240V 1~/2~ 50/60Hz | M24x1, 5 (Flow) M24x1, 5 | Pressure pump, variable | 80–425 | — |
Full specifications
| Control output | POKO ONE |
| Pump connection | M24x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | Pressure pump, variable speed |
| External sensor connection | Pt100 |
| Temperature range | 65...300 °C |
| Weight | 37 kg |
| Power supply | 200-240V 1~/2~ 50/60Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and Series Positioning
The Unistat T305 Dynamisches is a pressure-pump-based temperature control unit with a variable-speed, soft-start pump and integrated pressure control. Unlike the Unistat T305 HT and T305w HT siblings, this model lacks active refrigeration and relies on application heat loss or external water cooling to achieve its lower temperature limit of 65 °C. The expansion tank is not thermoregulated and is lockable for open bath configurations. The unit is rated IP20 and weighs 37 kg.
Behavior Under Load and Media
- Temperature range: 65...300 °C; ambient operating range: 5...40 °C.
- Maximum pump delivery: 45 l/min at 0.9 bar; pump speed is adjustable.
- Maximum permissible kinematic viscosity of the thermofluid: 50 mm²/s.
- For air-cooled operation, the application heat loss must fall between 0.1 kW and 2.8 kW depending on the unit to reach the minimum temperature without a chiller.
- The unit provides adjustable overtemperature protection per DIN 12876 and safety classification III / FL.
Installation, Control and Operation
- Power supply: 200–240 V, 1~ or 2~, 50/60 Hz; delivered with power cable and country-specific plug (must be specified at order). Maximum current at 230 V is just under 16 A, and the unit can be limited to 13 A or 10 A at first start-up.
- Pump connection: M24x1.5 male, bore shape Y (60°) per DIN 3863; mating hoses require ball socket per DIN 3863 and sleeve nut per DIN 3870.
- Controller: Pilot ONE with POKO ONE control output; display resolution 0.01 K; interfaces: Ethernet, USB (Host and Device), RS232; external sensor connection: Pt100.
- Expansion vessel must remain unlocked for externally closed applications and locked for externally open applications.
- Alarm signalling via optical, acoustic, and relay contacts.
- Process Safety mode requires the compressor to be set to Permanently ON (note: this unit has no active refrigeration, so this mode is not applicable).
- Power Failure Auto resumes temperature control automatically after a power outage if it was active.
- Cooling water supply should be shut off during downtimes, even overnight.
Integration & Operating Notes
Without active refrigeration, reaching 65 °C requires either sufficient heat loss from the application (0.1–2.8 kW depending on unit) or suitable water cooling. Verify the application's heat loss falls within this range before commissioning.
Transport and store the unit upright only; horizontal transport damages the compressor. Before commissioning, unscrew/activate leveling feet and engage parking brakes if casters are present. Casters are not for transport and carry only 25% of total mass symmetrically.
Supply voltage and frequency must match the rating plate. At 230 V the maximum current is just under 16 A; the unit can be limited to 13 A or 10 A at first start-up. Voltage tolerance ±10% per EN60034-1 applies only if frequency deviation does not run in the opposite direction.
For externally closed applications the expansion vessel must remain unlocked; locking it causes overpressure damage. For externally open applications the expansion vessel must be locked to prevent fluid loss.
Shut off the cooling water supply even for shorter downtimes such as overnight to prevent water damage and unnecessary load on the cooling water circuit. Periodically check hoses and connections for leaks and quality, and replace as needed.
If temperature control is active during a power outage, the process will automatically continue after the power outage. Assess whether unattended restart is acceptable for the process; if not, disable this function in the controller settings.
When moving the unit between cold and warm environments, allow about 2 hours acclimation before power-on to prevent condensation-induced electrical faults.
Selection & Integration Q&A
Choose the T305 Dynamisches for applications where heat loss or water cooling can handle the lower end (65 °C) and dynamic pressure control is needed. The HT variants offer active refrigeration for lower temperatures; select those if sub-ambient cooling is required.
The pump connection is M24x1.5 male with bore shape Y (60°) per DIN 3863. Use hoses with ball socket per DIN 3863 and sleeve nut per DIN 3870. Ensure the circuit pressure drop does not exceed the pump's 0.9 bar maximum delivery pressure at 45 l/min.
Install in ambient temperature 5–40 °C, IP20 environment, not in ATEX zones. Transport and store upright only. Allow 2 hours acclimation after temperature changes. For air-cooled operation, application heat loss must be 0.1–2.8 kW to reach 65 °C.
Periodically check hoses and connections for leaks and quality; replace as needed. Shut off cooling water during downtimes. Power cable and country-specific plug must be specified at order; external pressure sensor is optional. No active refrigeration means no compressor maintenance.