Peter Huber Kältemaschinenbau Unistat T340 HT Dynamisches
Peter Huber Kältemaschinenbau Unistat T340 HT Dynamisches — Temperature control units.
The Unistat T340 HT Dynamisches is a high-temperature dynamic temperature control unit with a 65–300 °C working range, 48 kW heating and 300 kW cooling at 30 °C. It is the largest high-temperature modification in the Unistat T305/T340 series, distinguished by its 56 l expansion volume and 200 l/min pump. It serves closed-loop process temperature control in industrial plants requiring three-phase 400 V supply.
Peter Huber Kältemaschinenbau Unistate Bis 425C Series — Model Comparison & Selection
| Model | Display resolution | Power supply | Pump connection | Pump type | Temperature range, °C | Expansion volume, m³ |
|---|---|---|---|---|---|---|
| Unistat T340 HT Dynamisches Current | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st | 65–300 | 0.056 |
| Chili Dynamisches | 0.01 | 200-240V 1~/2~ 50/60Hz | M24x1,5 AG male | Pressure pump, variable | 65–300 | 0.002 |
| Unistat T305 Dynamisches | 0, 01 | 200-240V 1~/2~ 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | 65–300 | — |
| Unistat T305 HT Dynamisches | 0, 01 | 200-240V 1~/2~ 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | 65–300 | — |
| Unistat T305w HT Dynamisches | 0, 01 | 200-240V 1~/2~ 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | 65–300 | — |
| Unistat T320 Dynamisches | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st | 65–300 | 0.024 |
| Unistat T320w HT Dynamisches | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | 65–300 | 0.024 |
| Unistat T330 HT Dynamisches | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | 65–300 | 0.024 |
| Unistat T330w HT Dynamisches | 0, 01 K | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | 65–300 | 0.024 |
| Unistat T340 Dynamisches | 0, 01 | 460V 3~ 60Hz | M38x1, 5 AG male | magnetically coupled, st | 65–300 | 0.056 |
| Unistat T345 Dynamisches | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st | 65–300 | 0.056 |
| Unistat T345 HT Dynamisches | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st | 65–300 | 0.056 |
| Unistat T350 HT Dynamisches | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st | 65–300 | 0.056 |
| Unistat T402 Dynamisches | 0, 01 | 400V 3~N 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | 80–425 | — |
| Unistat TR401 Dynamisches | 0, 01 | 200-240V 1~/2~ 50/60Hz | M24x1, 5 (Flow) M24x1, 5 | Pressure pump, variable | 50–400 | 0.005 |
| Unistat TR402 Dynamisches | 0, 01 | 200-240V 1~/2~ 50/60Hz | M24x1, 5 (Flow) M24x1, 5 | Pressure pump, variable | 80–425 | — |
Full specifications
| Alarm output | optic, acoustic, relay |
| Control output | POKO ONE |
| Expansion volume | 56 l |
| Heating capacity | 48 kW |
| Pump connection | M38x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | 65...300 °C |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling capacity | 300 kW (at 30 °C) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- The T340 HT is the high-temperature variant of the Unistat T340 platform, positioned above the T305 and T305w HT siblings in capacity and expansion volume.
- Heating capacity is 48 kW; cooling capacity is stated as 300 kW at 30 °C. The unit is a dynamic temperature control system with a magnetically coupled stainless steel pump, speed-adjustable, delivering up to 200 l/min at 5.5 bar.
- The expansion tank holds 56 l and is not thermoregulated; it is lockable for open-bath applications and must remain unlocked for externally closed circuits.
- Pump connections are M38x1.5 AG male. The unit is rated IP20 and safety class III / FL.
Control and process behavior
- Control is via the Pilot ONE controller with POKO ONE control output; display resolution is 0.01 K. An external Pt100 sensor connection is provided.
- The working range of 65–300 °C makes this a hot-side unit; it is not intended for near-ambient or sub-ambient work.
- Maximum permissible kinematic viscosity is 50 mm²/s, which constrains thermofluid selection at the low end of the range.
- Adjustable overtemperature protection per DIN 12876 is provided; the safety cut-out must be set for the process during commissioning.
- Alarm output is optic, acoustic and relay-based.
Electrical and installation
- Power supply is 400 V 3~ 50 Hz. Three-phase devices below 63 A are supplied with cable but without a plug; final connection must be made by an electrician.
- Leakage current exceeds 3.5 mA, which affects protective earthing and RCD selection.
- EMC classification is Class A, Group 1 to EN55011, intended for industrial environments.
- Ambient temperature must remain between 5 and 40 °C. The unit must be transported and stored upright and moved only when emptied.
- Interfaces include Ethernet, USB (host and device) and RS232.
Integration & Operating Notes
Voltage and frequency tolerances must be evaluated together; combinations such as -5% voltage with +2% frequency are explicitly not allowed. Verify the actual site supply against the combined envelope before commissioning, because a supply that looks acceptable on each parameter alone can still be outside the permitted operating window.
The unit is delivered with cable but without a plug for three-phase supplies below 63 A. The customer must provide the final connection, and the hard-wired power connection must be carried out by an electrician with supply voltage and frequency matched to the rating plate.
Protective earthing and residual current device selection must account for a leakage current exceeding 3.5 mA. Standard 30 mA RCDs may not be appropriate; the earthing design and RCD rating must be engineered for this leakage level.
The unit must be transported and stored upright only, because horizontal transport damages the compressor. It may only be moved when emptied of thermal fluid. Casters, if present, are not suitable for transport and carry only 25% of total mass symmetrically; use an industrial truck. Leveling feet must be activated and parking brakes engaged before commissioning.
For externally closed applications the expansion vessel must remain unlocked; locking it causes overpressure damage. For externally open applications it must be locked to prevent fluid loss. Confirm the circuit type before setting the vessel lock, as the wrong state damages the application or loses fluid.
The maximum permissible kinematic viscosity is 50 mm²/s. Thermofluid selection must be checked against this limit at the lowest operating temperature, because a fluid too viscous for the pump at 65 °C will not circulate correctly.
The pump delivers up to 200 l/min at 5.5 bar. The pressure drop of the customer's external circuit must be checked against this head at the required flow; a loop with excessive pressure drop will not achieve the design flow rate.
When Process Safety mode is used for exothermic process protection, the compressor must be set to Permanently ON so that cooling remains active at the overtemperature trigger. Without this setting the safety function will not have cooling available when it trips.
If temperature control is active during a power outage, Power Failure Auto resumes the process automatically after power returns. Assess whether unattended automatic restart is acceptable for the process before enabling this behavior.
The cooling water supply should be shut off even for short downtimes such as overnight. This prevents water damage and unnecessary load on the cooling water circuit during idle periods.
Hoses and hose connections must be periodically checked for leaks and quality and replaced as required. This is the preventive maintenance scope for the fluid circuit; degraded hoses are a leakage risk at up to 300 °C.
The unit must not be installed or started inside an ATEX zone; explosion protection would require an Ex px cabinet. With IP20 protection it must be placed in a dry, protected location away from water ingress, and ambient temperature must stay between 5 and 40 °C.
When moving the unit between cold and warm environments, allow about 2 hours acclimation before turning it on. Starting immediately risks condensation-induced electrical faults.
Selection & Integration Q&A
Choose the T340 HT when the process needs the larger 56 l expansion volume and the higher 48 kW heating / 300 kW cooling capacity at 30 °C. The T305 and T305w HT are smaller high-temperature variants; select them when the T340's capacity and expansion volume exceed the duty. All share the 65–300 °C range.
The pump connections are M38x1.5 AG male, and the external loop pressure drop must be checked against the pump's 5.5 bar head at up to 200 l/min. The thermofluid must stay below 50 mm²/s kinematic viscosity at the lowest operating temperature.
Install in a dry, protected location (IP20, no ATEX zone) with ambient 5–40 °C. Transport and store upright only, move only when emptied, and allow 2 hours acclimation before first power-on. The 400 V 3~ 50 Hz supply must meet the linked voltage/frequency tolerance envelope.
Periodically check hoses and connections for leaks and quality and replace as needed; shut off cooling water even overnight. The unit ships with cable but without a plug for three-phase connection below 63 A, so the final connection hardware and an electrician are site-supplied.