Peter Huber Kältemaschinenbau Unistat T350 HT Dynamisches

Peter Huber Kältemaschinenbau Unistat T350 HT Dynamisches — Temperature control units.

The Unistat T350 HT Dynamisches is a high-temperature dynamic temperature control unit with a 65–300 °C working range, 96 kW heating and 300 kW cooling at 30 °C. It is the larger sibling of the T305/T305 HT models, distinguished by its 300 kW cooling capacity and 56 l expansion volume. It serves high-temperature process loops requiring rapid, precise temperature control via the Pilot ONE controller.

Peter Huber Kältemaschinenbau Unistate Bis 425C Series — Model Comparison & Selection

Model Display resolutionPower supplyPump connectionPump typeTemperature range, °CExpansion volume, m³
Unistat T350 HT Dynamisches Current 0, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Chili Dynamisches 0.01200-240V 1~/2~ 50/60HzM24x1,5 AG malePressure pump, variable 65–3000.002
Unistat T305 Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable 65–300—
Unistat T305 HT Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable 65–300—
Unistat T305w HT Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable 65–300—
Unistat T320 Dynamisches 0, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st65–3000.024
Unistat T320w HT Dynamisches 0, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st65–3000.024
Unistat T330 HT Dynamisches 0, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st65–3000.024
Unistat T330w HT Dynamisches 0, 01 K400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st65–3000.024
Unistat T340 Dynamisches 0, 01460V 3~ 60HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T340 HT Dynamisches 0, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T345 Dynamisches 0, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T345 HT Dynamisches 0, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T402 Dynamisches 0, 01400V 3~N 50/60HzM24x1, 5 AG malePressure pump, variable 80–425—
Unistat TR401 Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 (Flow) M24x1, 5Pressure pump, variable 50–4000.005
Unistat TR402 Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 (Flow) M24x1, 5Pressure pump, variable 80–425—

Full specifications

Core Performance & Functional Specifications
Control output POKO ONE
Expansion volume 56 l
Heating capacity 96 kW
Pump connection M38x1, 5 AG male
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range 65...300 °C
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling capacity 300 kW (at 30 °C)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Behavior and process window

Installation, control and operation

Integration & Operating Notes

Electrical
Three-phase supply with coupled voltage/frequency tolerance

The unit requires 400 V 3~ 50 Hz. Voltage and frequency tolerances are coupled: ±5 % voltage only with ±2 % frequency, and opposite-sign deviations (e.g. -5 % voltage with +2 % frequency) are not allowed. Verify the site supply against this envelope before connection.

Unistat T350 HT Technical Data, page 2
Electrical
Plug not included for three-phase devices under 63 A

Three-phase devices drawing less than 63 A are delivered with cable but without a plug. The customer must provide and fit a suitable plug, and the hard-wired connection must be carried out by an electrician.

Unistat T350 HT Technical Data, page 2; Unistat Operation Manual, page 37
Electrical
High leakage current requires earth and RCD review

Leakage current exceeds 3.5 mA. Protective earthing must be designed accordingly, and residual current device selection must account for high leakage to avoid nuisance tripping or inadequate protection.

Unistat T350 HT Technical Data, page 2
Mounting
Transport upright and empty; casters not for transport

Only transport and store the unit upright and emptied; horizontal transport damages the compressor and filled transport risks thermal fluid overflow. Casters are not suitable for transport and carry only 25 % of total mass symmetrically; use an industrial truck. Activate parking brakes and leveling feet before commissioning.

Unistat Operation Manual, page 22
Mounting
Expansion vessel lock depends on open or closed application

For externally closed applications the expansion vessel must remain unlocked; locking it causes overpressure damage. For externally open applications the expansion vessel must be locked to prevent fluid loss. Set this correctly before operation.

Unistat Operation Manual, page 33
Process Limits
Thermofluid viscosity must stay below 50 mm²/s

Maximum permissible kinematic viscosity is 50 mm²/s. Select a thermofluid whose viscosity at the lowest operating temperature remains within this limit to avoid pump overload and poor heat transfer.

Unistat T350 HT Technical Data, page 1
Process Limits
External circuit pressure drop must match pump curve

The pump delivers up to 200 l/min at 5.5 bar. Check the pressure drop of the external circuit (hoses, heat exchanger, application) against the pump curve to ensure the required flow is achieved; pump speed is adjustable.

Unistat T350 HT Technical Data, page 1
Control & I/O
Process Safety mode requires compressor permanently ON

When using Process Safety mode, the compressor must be set to Permanently ON so that cooling remains active at the overtemperature trigger. Configure this in the controller before relying on the safety function.

Unistat Operation Manual, page 19
Control & I/O
Power Failure Auto resumes control automatically

If temperature control is active during a power outage, the process will automatically continue after power is restored when Power Failure Auto is enabled. Assess whether unattended restart is acceptable for the process.

Unistat Operation Manual, page 20
Maintenance
Shut off cooling water even for overnight downtime

Shut off the cooling water supply to the unit even for shorter downtimes such as overnight to prevent water damage and unnecessary load on the cooling water circuit.

Unistat Operation Manual, page 29
Maintenance
Periodically inspect hoses and connections

Check periodically for leaks and the quality of the hose and hose connections and replace as required. This is part of preventive maintenance for the fluid circuit.

Unistat Operation Manual, page 29
Environment
Not permitted inside ATEX zone without Ex px cabinet

Do not install or start up the temperature control unit within an ATEX zone. Explosion protection requires an Ex px cabinet. Verify the area classification before installation.

Unistat Operation Manual, page 15
Environment
Allow 2 hours acclimation after temperature change

When moving the unit between cold and warm environments, allow about 2 hours acclimation before power-on to prevent condensation-induced electrical faults.

Unistat Operation Manual, page 29

Selection & Integration Q&A

When should I choose the Unistat T350 HT Dynamisches over the T305 HT or T305w HT?

Choose the T350 HT when your process needs the higher cooling capacity (300 kW at 30 °C) and 96 kW heating in the 65–300 °C range. The T305/T305 HT are smaller siblings; the T305w HT is a different configuration. Select based on required heat removal and flow (up to 200 l/min).

What electrical supply and protection must be provided for this unit?

Provide a 400 V 3~ 50 Hz supply with voltage/frequency tolerances coupled (±5 % V only with ±2 % F, no opposite-sign deviations). Leakage current >3.5 mA requires earth and RCD review. Devices under 63 A come with cable but no plug; hard-wiring by an electrician.

What are the mounting and environment limits for installation?

Install upright and level on a dry, protected location (IP20), ambient 5–40 °C, not in an ATEX zone. Transport only upright and emptied; casters not for transport. Allow 2 h acclimation after temperature changes. Activate parking brakes and leveling feet before commissioning.

What maintenance is required and what is not included in base delivery?

Periodically check hoses and connections for leaks and quality, replace as needed; shut off cooling water even overnight. The base delivery includes cable but no plug for three-phase devices under 63 A. Thermofluid and external circuit components are not included and must be selected within 50 mm²/s viscosity.