Properties (other) Change heat transfer fluid (heat transfer fluid, liter indicator, bypass usage); cooling bath/cooling (CC-E only)
X
X
X
–
–
–
Sensor calibration
–
–
–
–
–
–
Automated compressor
X
X
X
–
–
–
Enter program
X
–
–
X
–
–
Pump settings Pump speed setpoint; pump pressure setpoint; control mode
X
X
X
–
–
–
Fluid properties Heat transfer fluid; liter indication; pressurisation (only with XT models)
X
X
X
–
–
–
Configuration manual parameters
Safety function High limit alarm internal; low limit alarm internal; high limit alarm process; low limit alarm process; hydrostatic correction; warning time limit (CC-E only); min. level (for temperature control units with analogue level sensor); max. level (for temperature controllers with analogue level sensor); optional freeze protection
X
X
X
–
–
–
setpoint
X
X
X
–
–
–
Setpoint limitation Min. setpoint; max. setpoint
X
X
X
–
–
–
Language
X
X
X
–
–
–
Temperature format
X
X
X
–
–
–
Temperature control mode
X
X
X
–
–
–
Favorites menu
X
–
–
–
X
–
Pilot ONE Ethernet
X
X
X
–
–
–
“X” = Value is reset, “–” = Value is not reset; (A) = All together; (B) = Device parameters without OT; (C) = Device Parameter; (D) Program Controller; (E) = Menu; (F) = Com.G@te
Overview of the E-grade variants
Temperature control units / E-grade
E-grade Basic
E-grade Exclusive
E-grade Professional
Beer force-aging thermostats
–
X
O
Unistat temperature control units
–
–
X
UniCAL
–
–
X
Other temperature control units
X
O
O
X = standard equipment, O = optional, – = not possible
Selecting the heat transfer fluid
Freezing point °C
Ethylene glycol
Water
1. Setpoint °C
1. Setpoint bar abs.
2. Setpoint °C
2. Setpoint bar abs.
3. Setpoint °C
3. Setpoint bar abs.
Achievable end temperature in °C
-50 °C
60.00
40.00
95
1.5
120
2.2
150
4.2
150
-45 °C
55.00
45.00
95
1.5
120
2.4
145
4.2
145
-40 °C
52.50
47.50
95
1.5
120
2.4
145
4.2
145
-35 °C
47.50
52.50
95
1.5
120
2.6
140
4.2
140
-30 °C
45.00
55.00
95
1.5
120
2.6
140
4.2
140
-25 °C
40.00
60.00
95
1.5
120
2.6
140
4.2
140
-20 °C
35.00
65.00
95
1.5
120
2.6
135
4.2
135
Mixing ratio in percent by volume: Ethylene glycol / Water. Exemplary representation of the system pressure. Factory settings for models with pressure blanketing.
Table
Description
(A)
(B)
(C)
(D)
(E)
(F)
IP address; subnet mask; remote access
Time unit
X
X
X
–
–
–
2nd setpoint
X
X
X
–
–
–
Representation of textual emphases
Emphasis
Description
Abc
Step-by-step explanation of the procedure.
→
Reference to information or procedures.
»Abc«
Reference to a paragraph in the document.
>Abc< [123]
Reference to the wiring diagram in the annex, specifying the designation and search string (number).
>Abc< [ABC]
Reference to a drawing in the same paragraph with specification of designation and search string (character).
▪
List, first level
‒
List, second level
Overview of transportation routes
Refrigerants
Refrigerant filling ex factory
Air freight
Road transportation
Sea freight
UN 3358
≥ 0.1 kg to ≤ 11.99 kg
not possible a)
possible
possible
UN 3358
≥ 12 kg
possible b)
possible b)
UN 3358 SP A103
≤ 99.9 g
possible
possible
possible
UN 2857
≤ 11.99 kg
possible
possible
possible
UN 2857
≥ 12 kg
not possible a)
possible b)
possible b)
a) Filling for transportation: Nitrogen. On-site filling with refrigerants: Only permitted by partners certified by Huber (branch, service partner, dealer or end customer). b) Transportation as hazardous goods.
Function description
Function
E-grade Basic
E-grade Exclusive
E-grade Professional
E-grade evaluation versions available (valid for 30 days)
X
X
X
Storage/loading temperature control programs onto USB stick
–
X
X
Copying settings
–
–
X
Service data recorders (flight recorder)
X
X
X
PB commands1
X
X
X
Communication watchdog
–
–
X
Process data recording directly to a USB stick: Setpoint, Actual Value Internal and Actual Value Process / Heating Capacity %, Cooling Capacity %, and Pump Pressure / Pump Speed, and VPC Pressure
–/–/–
X/X/–
X/X/X
Calendar start
–
–
X
1 See Data Communication Manual. Anything controllable via the graphical user interface of the Pilot ONE can be controlled by PB commands.
Engineering Facts
14 facts · every claim linked to a page
Installation
The unit must not be installed or started inside an ATEX zone, so hazardous-area installations are excluded.
Determines whether the unit can be placed in an explosive atmosphere at all.
Source
“Do NOT install or start up the temperature control unit within an ATEX zone.”
Installation
Refrigerant charge and room size must be checked during installation to avoid dangerous refrigerant concentration.
Room volume and charge size govern safe placement of refrigerant-based units.
Source
“Observe the amount of refrigerant (see data sheet/rating plate) and the room size when installing the temperature control unit.”
Installation
A drip tray for condensate/heat transfer fluid must be installed below the unit by the responsible body.
Affects site preparation and containment design before commissioning.
Source
“The responsible body must install a condensation water / heat transfer fluid drip tray below the temperature control unit.”
Installation
Minimum wall clearances depend on air- or water-cooling and on whether the unit runs with or without a bath.
Clearances must be respected to avoid inadequate energy dissipation and overheating.
Source
“Values in the table: without air outlet or connections / with air outlet or connections”
Installation
The unit is a Group 1 Class A device for industrial EMC environments, so it may not be EMC-compatible elsewhere.
Determines EMC suitability of the installation environment and wiring measures.
Source
“It is a Group 1 and Class A device according to the currently applicable EN55011.”
Installation
Only drinking water or filtered river/sea water may be used as cooling water, and sea water is prohibited.
Cooling water quality directly affects warranty and heat exchanger fouling.
Source
“Sea water must not be used for water cooling.”
Operation
Cooling water supply should be shut off during downtimes because flow can occur even when the unit is switched off.
Prevents water damage and unnecessary consumption between operating periods.
Source
“Shut off the cooling water supply to the temperature control unit even for shorter downtimes (e.g. overnight).”
Installation
A backflow preventer on the building side is required to protect the drinking water supply.
Mandatory plumbing measure for water-cooled units connected to mains water.
Source
“A backflow from the cooling water line into the drinking water supply must be prevented on the building side.”
Installation
Outdoor installation requires frost-protected cooling water lines and a minimum cooling water temperature of 5 °C.
Defines winterization and heating requirements for outdoor siting.
Source
“The cooling water temperature must not fall below 5 °C.”
Safety
The unit must always be connected to a socket with protective earth, and hardwiring must be done by a certified electrician.
Electrical safety requirement governing the power connection method.
Source
“Always connect the temperature control unit to a power socket with protective earth (PE).”
Installation
Replacement power cables are length-limited to 3 m for 1–2 phase and 4 m for 3-phase connections.
Constrains cable routing and replacement during installation.
Source
“Connections with 1 or 2 phases have a maximum length of 3 meters; connections with 3 phases a maximum length of 4 meters.”
Performance
Pressure-sensitive applications such as glass reactors require accounting for the unit's maximum flow pressure and possibly an external bypass.
Prevents overpressure damage to glass apparatus when selecting the unit.
Source
“The maximum flow pressure of a temperature control unit must be taken into account in case of pressure-sensitive applications, such as glass reactors.”
Operation
Heat transfer fluid must be brought to room temperature before switching off to avoid overpressure in the system.
Shutdown procedure that protects unit and application from overpressure.
Source
“the heat transfer fluid must always be adjusted to room temperature before switching off”
Selection
Hose inside diameters must at least match the pump connections, with larger diameters for longer lines.
Sizing rule for hoses that affects pressure drop and achievable temperature control.
Source
“The inside diameters of the temperature control hoses must correspond at least to the pump connections.”