Peter Huber Kältemaschinenbau Grande Fleur w-eo Dynamisches
Peter Huber Kältemaschinenbau Grande Fleur w-eo Dynamisches — Temperature control units.
The Grande Fleur w-eo Dynamisches is a water-cooled temperature control unit for external open loops only, with no expansion tank of its own. It covers -40 to 200 °C with 1.5 kW heating and 0.6 kW cooling at 20–200 °C, falling to 0.04 kW at -40 °C. The variable-speed pressure pump delivers up to 47 l/min at 0.9 bar.
Peter Huber Kältemaschinenbau Unistate Bis 45C Series — Model Comparison & Selection
| Model | Noise level, dB | Display resolution | Power supply | Pump connection | Pump type | Temperature range, °C |
|---|---|---|---|---|---|---|
| Grande Fleur w-eo Dynamisches Current | — | — | — | — | — | — |
| Grande Fleur Dynamisches | — | — | — | — | — | — |
| Grande Fleur w Dynamisches | — | 0, 01 | 208-240V 1~/2~ 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | -40–200 |
| Petite Fleur-eo Dynamisches | 58 | 0, 01 K | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | Pressure pump, variable | -40–200 |
| Petite Fleur w Dynamisches | — | 0, 01 | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | Pressure pump, variable | -40–200 |
| Unistat 410w Dynamisches | 61 | 0, 01 K | 220-240V 1~/2~ 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | -45–250 |
| Unistat 425w Dynamisches | 61 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
| Unistat 430w Dynamisches | 61 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
| Unistat P405w Dynamisches | — | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st | -45–250 |
| Unistat P425 Dynamisches | 65 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
| Unistat P425w Dynamisches | 61 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
| Unistat P430 Dynamisches | 65 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
| Unistat P430w Dynamisches | 65 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
Full specifications
Description
Construction and series positioning
- This modification is the water-cooled (w) external-open (eo) member of the Grande Fleur Dynamisches series; it has no expansion tank, so the connected application must absorb the thermal fluid's volume change.
- Siblings differ in cooling medium and acoustics: Grande Fleur Dynamisches is rated 61 dB(A), Petite Fleur Dynamisches 58 dB(A), and Grande Fleur-eo Dynamisches shares the 47 l/min pump flow.
- Cabinet is 295 x 530 x 570 mm at 55 kg, IP20, for dry indoor locations.
Thermal behavior
- Heating capacity is 1.5 kW across the -40...200 °C range.
- Cooling capacity is 0.6 kW from 0 °C through 200 °C, 0.35 kW at -20 °C, 0.2 kW at -30 °C and 0.04 kW at -40 °C; published output assumes 20 °C room temperature and 2 bar cooling-water differential pressure.
- Temperature stability is ±0.01 K with 0.01 K display resolution.
- Refrigerant is R-290 (A3, H220), GWP 0.02; flammable-refrigerant measures apply.
Control, connection and operation
- Pilot ONE controller with POKO ONE control output and ECS ONE control input; interfaces are Ethernet, USB host and device, and RS232; external Pt100 sensor input.
- Alarm is optic, acoustic and relay.
- Variable-speed pressure pump, max 47 l/min at 0.9 bar, M24x1.5 male pump connection; cooling water G1/2 male at 2–6 bar and 5–20 °C.
- Adjustable overtemperature protection per DIN 12876, safety class III/FL.
Integration & Operating Notes
Supply cooling water within 5–20 °C and 2–6 bar differential pressure; lower pressure or warmer water reduces the published cooling capacity, which is rated at 20 °C room temperature and 2 bar differential. Insulate pipes where water can fall below dew point.
The cooling-water circuit uses copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only cooling water compatible with these materials may be connected; aggressive or untreated water can corrode the circuit.
Supply is 110–120 V 1~ 60 Hz with ±5% voltage and ±2% frequency tolerance; combinations such as -5% voltage with +2% frequency are not permitted, so verify the site supply before commissioning.
Single-phase devices are delivered with a power cable and a country-specific plug that must be stated on the order; confirm the plug type against the site outlet to avoid a rework at installation.
The IP20 enclosure requires a dry indoor location with ambient temperature between 5 and 40 °C; the unit is EMC Class A, Group 1 to EN55011, so additional EMC precautions may be needed in residential or sensitive environments.
The maximum permissible kinematic viscosity of the heat-transfer fluid is 50 mm²/s; select a fluid that remains below this limit across the full -40...200 °C operating range, otherwise pump delivery and temperature control degrade.
Adjustable overtemperature protection per DIN 12876 with safety class III/FL must be set for the specific thermal fluid and process; the alarm reports optically, acoustically and via relay, so wire the relay into the plant interlock.
Pump connections are M24x1.5 male with bore shape Y (60°) per DIN 3863; procure hoses with ball socket per DIN 3863 and sleeve nut per DIN 3870 so the circuit seals correctly and can be serviced without adapters.
Selection & Integration Q&A
Pick the w-eo when the loop is external and open and site cooling water at 5–20 °C, 2–6 bar is available; it has no expansion tank, so the application must absorb volume change. The Grande Fleur Dynamisches is rated 61 dB(A) and the -eo variant shares the 47 l/min pump flow.
An open external loop of at least 1.5 l filling capacity that handles thermal expansion, plus cooling water at 5–20 °C and 2–6 bar differential through G1/2 male ports. The cooling-water circuit contains copper, 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water may be used.
Install indoors at 5–40 °C ambient in an IP20-protected dry location. Supply is 110–120 V 1~ 60 Hz with ±5% voltage and ±2% frequency tolerance applied simultaneously; combinations such as -5% voltage with +2% frequency are not allowed. EMC is Class A, Group 1 to EN55011.
The base unit ships with a power cable and country-specific plug only if specified at ordering. Pump ports are M24x1.5 male and require hoses with ball socket per DIN 3863 and sleeve nut per DIN 3870; the external open loop and its expansion volume are not part of delivery.