The KISS 104A is a 3-litre polycarbonate-bath thermostatic circulator with a non-variable pressure/suction pump and KISS controller, rated 25–100 °C (20–100 °C with water cooling, 15–100 °C with refrigerator). Its 1.5–2.1 kW heating capacity and ±0.05 K stability suit precise temperature control of external circuits and bath applications up to 100 °C.

Direct drop-in replacements

Eppendorf
MixMate ®
53%
match
⚖️ Compare

Peter Huber Kältemaschinenbau Mit Polycarbonatbaedern Bis 100C Series — Model Comparison & Selection

Model ControllerInterfaceMax delivery flow rate, liter/minutePump connectionWeight, kgAmbient temperature limit, °C
KISS 104A Current KISSUSB (Device), RS232 Inte14M16x1 AG male55–40
CC-104A Pilot ONEEthernet, USB (Host u. D22M16x1 AG male65–40
CC-106A Pilot ONEEthernet, USB (Host u. D22M16x1 AG male55–40
CC-108A Pilot ONEEthernet, USB (Host u. D27M16x1 AG male (Option, #65–40
CC-110A Pilot ONEEthernet, USB (Host u. D22M16x1 AG male65–40
CC-112A Pilot ONEEthernet, USB (Host u. D27M16x1 AG male (Option, #85–40
CC-118A Pilot ONEEthernet, USB (Host u. D22M16x1 AG male (Option, #85–40
KISS 106A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #55–40
KISS 108A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #65–40
KISS 110A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #65–40
KISS 112A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #85–40
KISS 118A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #85–40

Full specifications

Core Performance & Functional Specifications
Bath volume 3 l
Filling volume 2 l (min.)
Maximum delivery flow rate 14 l/min (pressure pump max.; 0, 25 bar)
Pump connection M16x1 AG male
Pump speed adjustable no
Pump type Pressure and suction pump, speed not variable
Temperature range 25...100 °C
with cooling unit: 15...100 °C · with water cooling: 20...100 °C
Temperature stability mfr ±0, 05 K
Alarm message mfr optic, acoustic
Dimensions (W x D x H) mfr 147 x 234 x 329 mm
Heating capacity mfr 1, 5 - 2, 1 kW
Delivery capacity suction pump max. mfr 10, 5 l/min (suction pump max.; 0, 17 bar)
Bath opening WxD/BD mfr ø = 25 / 150
Dimensions, Weight & Mechanical Design
Weight 5 kg
Electrical, Power & Drive Specifications
Current consumption 10 A (max.)
Power supply 200-240V 1~/2~ 50/60Hz
Interfaces, Connections & Control
Controller KISS
Digital interface USB (Device), RS232 Interface
Temperature display OLED Display
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Thermal behaviour and media

Installation, control and operation

Integration & Operating Notes

Process Limits
Cooling method sets the achievable low temperature

The base range starts at 25 °C; reaching 20 °C requires water cooling and 15 °C requires a refrigerator. Specify the cooling option against the actual process setpoint, because the unit cannot reach sub-25 °C temperatures without one of these cooling sources.

KISS 104A Immersion Cooler Datasheet, Technical Data
Electrical
Heating power falls with supply voltage

Heating capacity is 2.1 kW at 240 V but only 1.5 kW at 200 V. Verify the site voltage before sizing heat-up time or confirming the process power requirement; a 200 V supply delivers roughly 30% less heating power than 240 V.

KISS 104A Immersion Cooler Datasheet, Technical Data
Electrical
Branch circuit must be 10–16 A with UL 489 protection

Maximum current is 10 A; the branch circuit must be rated between 10 A and 16 A and include external branch circuit protection according to UL 489. Voltage and frequency tolerances are ±10% and must not combine in opposite directions.

KISS 104A Immersion Cooler Datasheet, Technical Data
Mounting
Dry indoor location with 5–40 °C ambient

IP20 protection means the unit must be kept dry and protected from water ingress. Ambient temperature at the installation site must remain between 5 °C and 40 °C, and space requirements around the unit must be respected for ventilation and access.

KISS 104A Immersion Cooler Datasheet, Technical Data
Control & I/O
Autostart restarts the unit after power failure

The autostart function automatically restarts the unit after a mains interruption. For processes where unattended restart is unsafe, this behaviour must be accounted for during commissioning and risk assessment.

KISS 104A Immersion Cooler Datasheet, Overview
Process Limits
Plastic pump incompatible with acetone and polyglycols

The plastic pump is not resistant to acetone or polyglycols (depending on manufacturer). Select an alternative bath fluid or request a more resistant plastic; for freeze protection, mix water with glysantine or ethylene glycol rather than using pure water.

KISS 104A Immersion Cooler Datasheet, page 2
Electrical
Country-specific plug must be specified at ordering

Single- and two-phase devices are supplied with a power cable and a country-specific plug. The plug type must be stated when ordering to avoid delays or rework at installation.

KISS 104A Immersion Cooler Datasheet, page 2
Control & I/O
Pt100 option displays but does not control temperature

The optional Pt100 sensor connection (#10688) is for displaying, not controlling, process temperature and is only available factory-fitted at additional charge. It must be included in the factory order; it cannot be retrofitted as a field option.

KISS 104A Immersion Cooler Datasheet, Overview

Selection & Integration Q&A

When should I choose the KISS 104A over the CC-104A or larger CC-106A/108A/110A siblings?

Choose the KISS 104A for a 3-litre bath with 25–100 °C base range and fixed-speed pressure/suction pump. The CC-104A and larger CC-106A/108A/110A are separate configurations; select them only when their documented volume or range matches the process, since this card's facts cover the KISS 104A only.

What supply and circuit protection does the KISS 104A require at the installation site?

200–240 V 1~/2~ 50/60 Hz, max. 10 A, with a branch circuit rated 10–16 A and external UL 489 protection. Voltage and frequency tolerances are ±10% and must not run in opposite directions. Specify the country-specific plug at ordering.

What are the mounting and ambient limits for the KISS 104A?

Install indoors in a dry location: IP20 protection prohibits water ingress, and ambient must stay between 5 °C and 40 °C. Respect space requirements around the unit for ventilation and access, and account for the autostart function after power failure.

What maintenance and accessory items are not included in the base delivery?

The Pt100 sensor connection (#10688) is a factory-fitted option at additional charge and only displays, not controls, process temperature. The power cable and country-specific plug must be specified at ordering. Bath fluid is not supplied; avoid acetone and polyglycols with the plastic pump.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
KISS 104A Immersion Cooler Datasheet
Datasheet ·135,7 КБ ·2 pp.