Peter Huber Kältemaschinenbau KISS 104A
Peter Huber Kältemaschinenbau KISS 104A — Thermostatic baths.
The KISS 104A is a 3-litre polycarbonate-bath thermostatic circulator with a non-variable pressure/suction pump and KISS controller, rated 25–100 °C (20–100 °C with water cooling, 15–100 °C with refrigerator). Its 1.5–2.1 kW heating capacity and ±0.05 K stability suit precise temperature control of external circuits and bath applications up to 100 °C.
Direct drop-in replacements
Peter Huber Kältemaschinenbau Mit Polycarbonatbaedern Bis 100C Series — Model Comparison & Selection
| Model | Controller | Interface | Max delivery flow rate, liter/minute | Pump connection | Weight, kg | Ambient temperature limit, °C |
|---|---|---|---|---|---|---|
| KISS 104A Current | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male | 5 | 5–40 |
| CC-104A | Pilot ONE | Ethernet, USB (Host u. D | 22 | M16x1 AG male | 6 | 5–40 |
| CC-106A | Pilot ONE | Ethernet, USB (Host u. D | 22 | M16x1 AG male | 5 | 5–40 |
| CC-108A | Pilot ONE | Ethernet, USB (Host u. D | 27 | M16x1 AG male (Option, # | 6 | 5–40 |
| CC-110A | Pilot ONE | Ethernet, USB (Host u. D | 22 | M16x1 AG male | 6 | 5–40 |
| CC-112A | Pilot ONE | Ethernet, USB (Host u. D | 27 | M16x1 AG male (Option, # | 8 | 5–40 |
| CC-118A | Pilot ONE | Ethernet, USB (Host u. D | 22 | M16x1 AG male (Option, # | 8 | 5–40 |
| KISS 106A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 5 | 5–40 |
| KISS 108A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 6 | 5–40 |
| KISS 110A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 6 | 5–40 |
| KISS 112A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 8 | 5–40 |
| KISS 118A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 8 | 5–40 |
Full specifications
| Bath volume | 3 l |
| Filling volume | 2 l (min.) |
| Maximum delivery flow rate | 14 l/min (pressure pump max.; 0, 25 bar) |
| Pump connection | M16x1 AG male |
| Pump speed adjustable | no |
| Pump type | Pressure and suction pump, speed not variable |
| Temperature range | 25...100 °C with cooling unit: 15...100 °C · with water cooling: 20...100 °C |
| Temperature stability mfr | ±0, 05 K |
| Alarm message mfr | optic, acoustic |
| Dimensions (W x D x H) mfr | 147 x 234 x 329 mm |
| Heating capacity mfr | 1, 5 - 2, 1 kW |
| Delivery capacity suction pump max. mfr | 10, 5 l/min (suction pump max.; 0, 17 bar) |
| Bath opening WxD/BD mfr | ø = 25 / 150 |
| Weight | 5 kg |
| Current consumption | 10 A (max.) |
| Power supply | 200-240V 1~/2~ 50/60Hz |
| Controller | KISS |
| Digital interface | USB (Device), RS232 Interface |
| Temperature display | OLED Display |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- The KISS 104A is the 3-litre member of the KISS series; sibling models CC-104A through CC-110A share the 104/106/108/110 size codes but differ in configuration and are not interchangeable with this unit.
- Bath volume is 3 l with a minimum filling volume of 2 l; the bath opening is ø25 mm / 150 mm depth, and overall dimensions are 147 × 234 × 329 mm at 5 kg.
- The pump is a pressure and suction type with fixed speed (not adjustable), delivering up to 14 l/min at 0.25 bar pressure and 10.5 l/min at 0.17 bar suction; connection is M16×1 AG male.
- Control is via the KISS controller with OLED display; interfaces are USB (Device) and RS232.
Thermal behaviour and media
- The base temperature range is 25–100 °C; water cooling extends the low end to 20 °C and a refrigerator to 15 °C, so the cooling method must be matched to the required process temperature.
- Heating capacity is voltage-dependent: 2.1 kW at 240 V, 2.0 kW at 230 V, 1.8 kW at 220 V, 1.6 kW at 208 V and 1.5 kW at 200 V.
- Temperature stability is ±0.05 K; alarm messages are optic and acoustic.
- The plastic pump is not resistant to acetone or polyglycols (depending on manufacturer); for freeze protection, water should be mixed with glysantine or ethylene glycol.
Installation, control and operation
- Supply is 200–240 V 1~/2~ 50/60 Hz with maximum current 10 A; branch circuit protection must be 10–16 A, and external branch circuit protection according to UL 489 is required.
- Voltage and frequency tolerances are ±10% and must not run in opposite directions.
- Protection rating is IP20, so the unit must be installed in a dry indoor location; ambient temperature must remain 5–40 °C.
- Safety class is III / FL with adjustable overtemperature protection according to DIN 12876.
- An autostart function restarts the unit after power failure; this behaviour must be considered during commissioning.
- Single- and two-phase devices are supplied with a power cable and country-specific plug, which must be specified at ordering.
- An optional Pt100 sensor connection (#10688) displays (does not control) process temperature and is only available factory-fitted at additional charge.
Integration & Operating Notes
The base range starts at 25 °C; reaching 20 °C requires water cooling and 15 °C requires a refrigerator. Specify the cooling option against the actual process setpoint, because the unit cannot reach sub-25 °C temperatures without one of these cooling sources.
Heating capacity is 2.1 kW at 240 V but only 1.5 kW at 200 V. Verify the site voltage before sizing heat-up time or confirming the process power requirement; a 200 V supply delivers roughly 30% less heating power than 240 V.
Maximum current is 10 A; the branch circuit must be rated between 10 A and 16 A and include external branch circuit protection according to UL 489. Voltage and frequency tolerances are ±10% and must not combine in opposite directions.
IP20 protection means the unit must be kept dry and protected from water ingress. Ambient temperature at the installation site must remain between 5 °C and 40 °C, and space requirements around the unit must be respected for ventilation and access.
The autostart function automatically restarts the unit after a mains interruption. For processes where unattended restart is unsafe, this behaviour must be accounted for during commissioning and risk assessment.
The plastic pump is not resistant to acetone or polyglycols (depending on manufacturer). Select an alternative bath fluid or request a more resistant plastic; for freeze protection, mix water with glysantine or ethylene glycol rather than using pure water.
Single- and two-phase devices are supplied with a power cable and a country-specific plug. The plug type must be stated when ordering to avoid delays or rework at installation.
The optional Pt100 sensor connection (#10688) is for displaying, not controlling, process temperature and is only available factory-fitted at additional charge. It must be included in the factory order; it cannot be retrofitted as a field option.
Selection & Integration Q&A
Choose the KISS 104A for a 3-litre bath with 25–100 °C base range and fixed-speed pressure/suction pump. The CC-104A and larger CC-106A/108A/110A are separate configurations; select them only when their documented volume or range matches the process, since this card's facts cover the KISS 104A only.
200–240 V 1~/2~ 50/60 Hz, max. 10 A, with a branch circuit rated 10–16 A and external UL 489 protection. Voltage and frequency tolerances are ±10% and must not run in opposite directions. Specify the country-specific plug at ordering.
Install indoors in a dry location: IP20 protection prohibits water ingress, and ambient must stay between 5 °C and 40 °C. Respect space requirements around the unit for ventilation and access, and account for the autostart function after power failure.
The Pt100 sensor connection (#10688) is a factory-fitted option at additional charge and only displays, not controls, process temperature. The power cable and country-specific plug must be specified at ordering. Bath fluid is not supplied; avoid acetone and polyglycols with the plastic pump.