Peter Huber Kältemaschinenbau KISS 106A
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau KISS 106A

Peter Huber Kältemaschinenbau KISS 106A — Thermostatic baths.

The KISS 106A is an immersion circulator with a 4.4 L polycarbonate bath opening of 130 x 110 mm and a pressure/suction pump, rated for 25...100 °C without auxiliary cooling. It is the mid-size 106 variant in the KISS series, distinguished from the 104/108/110 siblings by its 4.4 L bath volume. It serves as a stand-alone thermostatic bath for sample tempering in the 25–100 °C range.

Direct drop-in replacements

Eppendorf
MixMate ®
63%
match
⚖️ Compare

Peter Huber Kältemaschinenbau Mit Polycarbonatbaedern Bis 100C Series — Model Comparison & Selection

Model ControllerInterfaceMax delivery flow rate, liter/minutePump connectionWeight, kgAmbient temperature limit, °C
KISS 106A Current KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #55–40
CC-104A Pilot ONEEthernet, USB (Host u. D22M16x1 AG male65–40
CC-106A Pilot ONEEthernet, USB (Host u. D22M16x1 AG male55–40
CC-108A Pilot ONEEthernet, USB (Host u. D27M16x1 AG male (Option, #65–40
CC-110A Pilot ONEEthernet, USB (Host u. D22M16x1 AG male65–40
CC-112A Pilot ONEEthernet, USB (Host u. D27M16x1 AG male (Option, #85–40
CC-118A Pilot ONEEthernet, USB (Host u. D22M16x1 AG male (Option, #85–40
KISS 104A KISSUSB (Device), RS232 Inte14M16x1 AG male55–40
KISS 108A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #65–40
KISS 110A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #65–40
KISS 112A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #85–40
KISS 118A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #85–40

Full specifications

Core Performance & Functional Specifications
Alarm output optic, acoustic
Dimensions (w × D × H) 147 x 307 x 330 mm
Maximum delivery flow rate 14 l/min (pressure pump; 0, 25 bar)
Pump connection M16x1 AG male (Option, #19606) (Option, #19606)
Pump speed adjustable no
Pump type Pressure and suction pump, speed not variable
Temperature range 25...100 °C
with cooling unit: 15...100 °C · with water cooling: 20...100 °C
Temperature stability mfr ±0, 05 K
Heating capacity mfr 1, 5 - 2, 1 kW
Delivery capacity suction pump max. mfr 10, 5 l/min (suction pump; 0, 17 bar)
Bath opening WxD/BD mfr 130 x 110 / 150 mm (WxD/depth)
Bath volume mfr 4, 4 l
Filling volume min. mfr 2, 5 l (min)
Dimensions, Weight & Mechanical Design
Weight 5 kg
Electrical, Power & Drive Specifications
Current consumption 10 A (max)
Power supply 200-240V 1~/2~ 50/60Hz
Interfaces, Connections & Control
Controller KISS
Digital interface USB (Device), RS232 Interface
Temperature display OLED Display
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Thermal behavior and media

Control, installation and operation

Integration & Operating Notes

Electrical
Provide UL 489 branch protection and 10–16 A fuse

The datasheet requires external branch circuit protection according to UL 489 and specifies a minimum fuse of 10 A and maximum of 16 A on a 200-240 V 1~/2~ 50/60 Hz supply. Size the circuit and protective device to these limits; the unit draws up to 10 A.

Datasheet, Technical Data / Installation
Process Limits
Sub-ambient operation needs water cooling or a refrigerator

The base range starts at 25 °C. Reaching 20...100 °C requires water cooling and 15...100 °C requires a refrigerator; without auxiliary cooling the bath cannot regulate below ambient, so the cooling method must be selected with the required low temperature.

Datasheet, Technical Data
Cooling & Media
Avoid acetone and polyglycols with the standard plastic pump

The plastic pump is not resistant to acetone or polyglycols (depending on manufacturer). For these media an alternative pump material is required; for freeze protection below ambient, mix water with glysantine or ethylene glycol rather than using pure water.

Datasheet, Technical Data / Operation
Control & I/O
Pt100 option is display-only and factory-fitted

The optional Pt100 sensor connection #10688 is available only factory-fitted and is used to display (not control) e.g. the process temperature. It cannot be added in the field and cannot close a control loop.

Datasheet, Technical Data
Operation
Autostart restarts after power failure without operator action

The autostart function restarts the unit automatically after a power failure. For unattended processes, confirm this behavior is acceptable or provide external sequencing, since the bath will resume heating without a manual restart.

Datasheet, Technical Data
Mounting
Install indoors at 5–40 °C ambient, IP20 only

The unit is IP20 and must be installed in an environment between 5 °C and 40 °C. Respect the space requirements at the installation site for access and ventilation; the IP20 rating excludes wet or dusty locations.

Datasheet, Technical Data / Installation

Selection & Integration Q&A

When should I choose the KISS 106A rather than a sibling such as the 104A or 108A?

Pick the 106A when the sample load and immersion depth fit its 4.4 L bath and 130 x 110 mm opening. Step to a smaller sibling for less bath volume or a larger one when more volume or opening area is needed; all share the 25...100 °C base range.

What auxiliary cooling and fluid do I need to reach 15 °C with this unit?

The base range is 25...100 °C; 20...100 °C requires water cooling and 15...100 °C requires a refrigerator. For freeze protection, mix water with glysantine or ethylene glycol rather than using pure water.

Which accessories or options are not included in the base delivery and must be ordered?

The Pt100 process sensor connection #10688 is factory-fitted only and display-only; the M16x1 male pump connection #19606 is an option. Water cooling or a refrigerator is a separate selection needed for sub-ambient operation.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
KISS 106A Immersion Circulator
Datasheet ·126,6 КБ ·2 pp.
View PDF