Peter Huber Kältemaschinenbau TC100
Peter Huber Kältemaschinenbau TC100 — Immersion coolers with Noise level 55 dB.
The TC100 is a two-stage cascade immersion cooler with a 48 mm × 155 mm probe on a 1150 mm flexible connection, covering -100 to +40 °C. Within the TC100 series it is the base model without the E-variant electronics; the -F variants differ in noise. It serves as a direct-contact cold trap or bath cooler in laboratory and pilot vacuum/process lines.
Peter Huber Kältemaschinenbau Eintauchkuehler Kuehlfallen Series — Model Comparison & Selection
| Model | Noise level, dB | Controller | Power supply | Refrigerant | Safety class | Temperature range, °C |
|---|---|---|---|---|---|---|
| TC100 Current | 55 | ON-OFF-Regler | 220-240V 1~/2~ 50/60Hz | R-1150 (A3, H220) | I / NFL | -100–40 |
| CT50 Single OLÉ | 59 | OLÉ | 220-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | III / FL | -50–50 |
| TC100E-F | 55 | MPC | 110-120V 1~ 60Hz | R-1150 (A3, H220) | I / NFL | -100–40 |
| TC100-F | 58 | ON-OFF-Regler | 220-240V 1~/2~ 50/60Hz | R-1150 (A3, H220) | I / NFL | -100–40 |
| TC45 | — | ON-OFF-Regler | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -45–100 |
| TC45E | — | MPC | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -45–100 |
| TC45E-F | 54 | MPC | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -45–100 |
| TC45-F | 59 | ON-OFF-Regler | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -45–100 |
| TC50 | — | ON-OFF-Regler | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -50–50 |
| TC50E | — | MPC | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -50–50 |
| TC50E-F | 60 | MPC | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -50–50 |
| TC50-F | 60 | ON-OFF-Regler | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -50–50 |
Full specifications
| Noise level | 55 dB(A) |
| Pump speed adjustable | no |
| Pump type | no pump |
| Temperature range | -100...40 °C |
| Dimensions (W x D x H) mfr | 295 x 500 x 570 mm |
| Refrigeration machine mfr | air-cooled, CFC- and HCFC-free |
| Weight | 61 kg |
| Power supply | 220-240V 1~/2~ 50/60Hz |
| Current consumption max. mfr | 6, 5 A (max) |
| Controller | ON-OFF-controller |
| Refrigerant | R-452A (A1, H280) (1st stage) 2nd stage: R-1150 (A3, H220) |
| Refrigerant quantity 1st stage mfr | 0, 26 kg (1st stage) |
| Refrigerant quantity 2nd stage mfr | 0, 056 kg (2nd stage) |
| Ambient temperature limit | 5...40 °C |
| Cooling capacity mfr | 0, 16 at 0 °C to 0, 01 at -100 °C kW (at °C bath temperature) |
| Safety class | I / NFL |
Description
Construction and series position
The TC100 is an air-cooled, CFC/HCFC-free cascade refrigeration unit built as a separate chassis (295 × 500 × 570 mm, 61 kg) connected by a 1150 mm flexible line to a 48 mm nominal diameter, 155 mm long immersion probe. Cooling is provided by a two-stage circuit: R-452A (A1, H280) in the first stage and R-1150 (A3, H220) in the second. Within the series, the TC100 is the base configuration; the TC100E and TC100E-F add electronics, and the TC100-F is the -F variant at 58 dB(A) versus 55 dB(A) for this model. The CT50 Single OLÉ is a smaller sibling at 59 dB(A).
Thermal behavior
Cooling capacity is quoted against bath temperature: 0.16 kW at 0 °C falling to 0.01 kW at -100 °C, so the unit is intended for low-load, low-temperature duty rather than bulk heat removal. Capacity also degrades as ambient temperature rises above the reference, so sizing must include margin for rooms warmer than 20 °C. Control is ON-OFF (no proportional or pump-assisted circulation; no pump is fitted).
Installation and operation
The unit is air-cooled and requires the specified clearance around the condenser for airflow and service. Electrical supply is 220–240 V, 1~ or 2~, 50/60 Hz, with a maximum current of 6.5 A; protection must be between 10 A and 16 A. Voltage and frequency tolerances are coupled (±5 % voltage only with ±2 % frequency, and opposite-sign deviations are not permitted). Single/two-phase units ship with a country-specific plug that must be specified at order. The equipment is EMC Class A, Group 1 (EN 55011), i.e. intended for industrial environments, and is pressure equipment under Art. 4.3 PED. The second stage uses a flammable A3 refrigerant, which governs handling and service procedures.
Integration & Operating Notes
Capacity is only 0.01 kW at -100 °C versus 0.16 kW at 0 °C, so the load at the probe must be kept very small at the low end; oversizing the thermal mass or heat leak will prevent reaching setpoint.
The probe is 48 mm nominal diameter and 155 mm long, and the flexible connection is 1150 mm; confirm the vessel port and the physical distance between the refrigeration chassis and the probe before ordering.
A -5 % voltage deviation combined with +2 % frequency is not allowed; supply quality must satisfy EN 60034-1 with matched deviation signs, otherwise the compressor is at risk.
Single/two-phase units are delivered with a power cable and a country-specific plug that must be specified at ordering; three-phase units below 63 A come with cable but no plug.
Maximum current is 6.5 A; the upstream fuse must be at least 10 A and no more than 16 A, so the supply circuit and breaker rating must be chosen within this window.
The unit is air-cooled; space requirements around the chassis must be respected for airflow and service, and ambient temperature must stay within 5–40 °C, with capacity derating above the reference ambient.
The second stage contains R-1150 (A3, H220); service and leak handling must follow flammable-refrigerant procedures, and only qualified personnel should open the circuit.
Selection & Integration Q&A
Pick the TC100 when basic ON-OFF control and 55 dB(A) are sufficient. The TC100E/TC100E-F add electronics; the TC100-F is the -F variant at 58 dB(A). Choose a sibling only if its extra function or noise level is required.
220–240 V, 1~ or 2~, 50/60 Hz, max 6.5 A, with a 10–16 A fuse. Voltage and frequency tolerances are coupled (±5 % V only with ±2 % Hz, no opposite-sign deviation). Order the country-specific plug; three-phase units below 63 A ship without plug.
The country-specific plug must be specified at order; no pump or external circulation is included. The second stage uses flammable R-1150 (A3), so service must follow flammable-refrigerant procedures. EMC Class A means industrial-environment use only.