Peter Huber Kältemaschinenbau TC45-F
Peter Huber Kältemaschinenbau TC45-F — Immersion coolers with Noise level 59 dB.
The TC45-F is an air-cooled immersion cooler with a 13 mm × 600 mm probe and 1050 mm flexible connection, rated for -45 to 100 °C using R-290 refrigerant. Its 0.24 kW at 0 °C falls to 0.05 kW at -40 °C, so it suits low-load setpoint holding rather than high-heat-load pull-down. It is the compact low-capacity member of the series, intended for direct immersion into vessels or baths.
Peter Huber Kältemaschinenbau Eintauchkuehler Kuehlfallen Series — Model Comparison & Selection
| Model | Noise level, dB | Controller | Power supply | Refrigerant | Safety class | Temperature range, °C |
|---|---|---|---|---|---|---|
| TC45-F Current | 59 | ON-OFF-Regler | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -45–100 |
| CT50 Single OLÉ | 59 | OLÉ | 220-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | III / FL | -50–50 |
| TC100 | 55 | ON-OFF-Regler | 220-240V 1~/2~ 50/60Hz | R-1150 (A3, H220) | I / NFL | -100–40 |
| TC100E-F | 55 | MPC | 110-120V 1~ 60Hz | R-1150 (A3, H220) | I / NFL | -100–40 |
| TC100-F | 58 | ON-OFF-Regler | 220-240V 1~/2~ 50/60Hz | R-1150 (A3, H220) | I / NFL | -100–40 |
| TC45 | — | ON-OFF-Regler | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -45–100 |
| TC45E | — | MPC | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -45–100 |
| TC45E-F | 54 | MPC | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -45–100 |
| TC50 | — | ON-OFF-Regler | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -50–50 |
| TC50E | — | MPC | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -50–50 |
| TC50E-F | 60 | MPC | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -50–50 |
| TC50-F | 60 | ON-OFF-Regler | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -50–50 |
Full specifications
| Noise level | 59 dB(A) |
| Pump speed adjustable | no |
| Pump type | no pump |
| Temperature range | -45...100 °C |
| Dimensions (W x D x H) mfr | 190 x 295 x 360 mm |
| Refrigeration machine mfr | air-cooled, natural refrigerant |
| -20 mfr | 0, 18 kW (at -20 °C) |
| -30 mfr | 0, 1 kW (at -30 °C) |
| -40 mfr | 0, 05 kW (at -40 °C) |
| Weight | 16 kg |
| Power supply | 208-240V 1~/2~ 50/60Hz |
| Current consumption max. mfr | 1, 4 A (max) |
| Controller | ON-OFF-controller |
| Refrigerant | R-290 (A3, H220) (1st stage) |
| Refrigerant quantity 1st stage mfr | 0, 041 kg (1st stage) |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Cooling capacity mfr | 0, 24 kW (at 0 °C) |
| Safety class | I / NFL |
Description
Construction and series positioning
- Single-stage air-cooled refrigeration circuit using R-290 (A3, H220), 0.041 kg charge; no pump and no adjustable pump speed.
- ON-OFF controller; probe nominal diameter 13 mm, length 600 mm, flexible connection 1050 mm.
- Footprint 190 × 295 × 360 mm, 16 kg. Sibling models TC100/TC100E/TC100E-F are quieter (55 dB(A)) and larger; the TC45-F is the compact low-capacity variant at 59 dB(A).
Behavior under load
- Cooling capacity is 0.24 kW at 0 °C, 0.18 kW at -20 °C, 0.10 kW at -30 °C and 0.05 kW at -40 °C, all valid at 20 °C room temperature; capacity may drop at higher ambient.
- The refrigeration unit runs continuously, so heat rejection to the room is continuous and must be planned.
Installation and operation
- Supply 208–240 V, 1~/2~, 50/60 Hz; max current 1.4 A; IP20; EMC Class A, Group 1; safety class I/NFL.
- Ambient 5–40 °C; dry indoor location only. Upstream fuse 10–16 A. Country-specific plug must be specified at order.
- Voltage tolerance ±5% only with frequency tolerance ±2%; -5% voltage with +2% frequency is not allowed.
Integration & Operating Notes
The ±5% voltage tolerance is only valid together with ±2% frequency tolerance, and the combination of -5% voltage with +2% frequency is explicitly not allowed. Verify the site supply against this combined window before commissioning, because a supply that passes each limit separately may still be outside the permitted combination.
Capacity falls from 0.24 kW at 0 °C to 0.05 kW at -40 °C. The required heat load at the lowest operating temperature must be checked against this curve; sizing on the 0 °C figure will under-cool at deep setpoints.
The probe is 13 mm in diameter and 600 mm long with a 1050 mm flexible connection. Confirm the vessel port accepts the 13 mm probe and that 600 mm insertion reaches the required liquid depth without the flexible connection being strained.
IP20 requires a dry indoor location protected from water and dust. R-290 is a flammable A3 refrigerant, so the installation area must be assessed for ignition sources and ventilation before the unit is placed.
The datasheet specifies a minimum fuse of 10 A and a maximum of 16 A. Size the upstream protection inside this window; a breaker outside it is not covered by the manufacturer's installation requirements.
Single- and two-phase devices are delivered with a power cable and a country-specific plug that must be specified at ordering. Confirm the plug standard with the customer before release to avoid rework at commissioning.
The refrigeration unit works continuously rather than cycling, so the room receives a continuous heat rejection. Account for this in room ventilation or air-conditioning sizing, especially in small or closed laboratories.
Selection & Integration Q&A
Pick the TC45-F for low heat loads at deep setpoints down to -45 °C where its 0.05 kW at -40 °C suffices. Choose a TC100-series sibling when higher capacity or lower noise (55 dB(A) vs 59 dB(A)) is needed.
208–240 V, 1~/2~, 50/60 Hz, max 1.4 A, with a 10–16 A upstream fuse. Voltage tolerance ±5% applies only with ±2% frequency tolerance; -5% voltage with +2% frequency is not allowed. Specify the country plug at order.
Dry indoor location, ambient 5–40 °C, IP20, EMC Class A industrial environment. R-290 flammable refrigerant requires ignition-source and ventilation assessment. ON-OFF control, no pump; refrigeration runs continuously.
The country-specific plug must be specified at order; no pump is fitted. The 600 mm probe and 1050 mm flexible connection define the immersion interface. Continuous operation means ongoing room heat rejection must be planned.