Peter Huber Kältemaschinenbau Unimotive 10w
Peter Huber Kältemaschinenbau Unimotive 10w — Temperature control units.
Water-cooled Unimotive 10w immersion circulator with magnetically coupled stainless-steel pressure pump, covering -45 to 95 °C with 12 kW heating and 14 kW cooling at 20 °C. It is the base 10-size modification of the Unimotive series, distinguished by its water-cooled refrigeration circuit and 400 V three-phase supply. Intended for temperature control of water-glycol circuits in automotive test stands and similar industrial processes.
Peter Huber Kältemaschinenbau Unimotive Series — Model Comparison & Selection
| Model | Display resolution | Heating capacity, W | Power supply | Pump connection | Refrigerant | Safety class |
|---|---|---|---|---|---|---|
| Unimotive 10w Current | 0.01 | 12000 | 400V 3~ 50Hz | M38x1,5 AG male | R-452A (A1, H280) | III / FL |
| Unimotive 10w-XT | 0, 01 | 12000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | I / NFL |
| Unimotive 20w | 0, 01 | 12000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | III / FL |
| Unimotive 20w-XT | 0, 01 | 12000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | I / NFL |
| Unimotive 26w | 0, 01 K | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | III / FL |
| Unimotive 26w-XT | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | I / NFL |
| Unimotive 27w | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | III / FL |
| Unimotive GL 07w | 0, 01 | 12000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 07w-XT | 0, 01 | 12000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | I / NFL |
| Unimotive GL 10w | 0, 01 | 24000 | 460V 3~ 60Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 10w-XT | 0, 01 | 24000 | 460V 3~ 60Hz | M38x1, 5 AG male | R-744 (A1, H280) | I / NFL |
| Unimotive GL 20w | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 20w-XT | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | I / NFL |
| Unimotive GL 30w | 0, 01 | 24000 | 460V 3~ 60Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 50w | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 50w-XT | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | I / NFL |
| Unimotive GL 60w | 0, 01 | 48000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 60w-XT | 0, 01 | 48000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | I / NFL |
Full specifications
| Control output | POKO ONE |
| Expansion volume | 19.9 l |
| Filling volume | 8.2 l (min) |
| Heating capacity | 12 kW |
| Pump connection | M38x1,5 AG male |
| Pump speed adjustable | no |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -45...95 °C |
| Temperature stability | 0.05 K |
| Alarm message mfr | optic, acoustic, relay |
| Dimensions (W x D x H) mfr | 730 x 804 x 1738 mm |
| Refrigeration machine mfr | water-cooled, CFC- and HCFC-free |
| Delivery capacity pressure pump max. mfr | 201 l/min; 5,3 bar |
| Pump characteristics mfr | Pressure pump, speed not variable, magnetically |
| Control input mfr | ECS ONE |
| Weight | 440 kg |
| Current consumption | 31.5 A (max) |
| Power supply | 400V 3~ 50Hz |
| Fuse mfr | 3x32A |
| Controller | Pilot ONE |
| Display resolution | 0.01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Refrigerant | R-452A (A1, H280) (1st stage) |
| Refrigerant quantity | 2.9 kg (1st stage) |
| Maximum viscosity | 200 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G3/4 male |
| Cooling water consumption | 620 l/h (cooling water 15°C, flow 0°C) |
| Cooling water differential pressure | 1 bar (min) |
| Cooling Water Pressure | 6 bar (max) |
| Protection rating | IP20 |
| Cooling capacity mfr | 20°C:14 kW; 0°C:10 kW; -20°C:5 kW; -30°C:2,1 kW; (at °C flow temperature) |
| Safety class | III / FL |
Description
Construction and series positioning
- The Unimotive 10w is a water-cooled temperature control unit with a CFC/HCFC-free refrigeration machine using R-452A (2.9 kg, 1st stage) and a magnetically coupled stainless-steel pressure pump (speed not variable).
- Within the Unimotive series it is the base 10-size water-cooled variant; the sibling 10w-XT, 20w, 20w-XT and 26w differ in size or configuration and are not interchangeable with this model's ratings.
- Cabinet dimensions 730 x 804 x 1738 mm, weight 440 kg, protection IP20, safety class III / FL.
Working range and load behavior
- Temperature range -45 to 95 °C with 0.01 K display resolution and 0.05 K stability; heating capacity 12 kW.
- Cooling capacity falls with flow temperature: 14 kW at 20 °C, 10 kW at 0 °C, 5 kW at -20 °C, 2.1 kW at -30 °C and 0.8 kW at -40 °C, so low-temperature duty must be checked against the process load.
- Pump delivery up to 201 l/min at 5.3 bar maximum; flow decreases as external pressure drop rises. Pressure-sensitive applications (e.g. glass reactors) must account for the maximum flow pressure.
- The unit is built for water-ethylene glycol mixtures with corrosion protection (e.g. Glysantin®) down to -45 °C; filling volume minimum 8.2 l, expansion volume 19.9 l.
Installation, connection and control
- Power supply 400 V 3~ 50 Hz, maximum current 31.5 A, fusing 3x32 A; delivered with cable but without plug (three-phase devices below 63 A). Voltage/frequency tolerance ±5 %/±2 %, not to be combined in the unfavorable direction. Leakage current exceeds 3.5 mA.
- Cooling water connections G3/4 male; cooling water intake 5–20 °C, differential pressure 1–4 bar, static pressure max 6 bar, consumption 620 l/h at 15 °C cooling water and 0 °C flow. A building-side backflow preventer is required.
- Pump connections M38x1.5 male; hoses must have inside diameters at least equal to the pump connections, larger for long lines.
- Control via Pilot ONE controller with POKO ONE control output and ECS ONE control input; interfaces Ethernet, USB (host and device) and RS232; external Pt100 sensor connection; alarm via optic, acoustic and relay signals.
Integration & Operating Notes
The responsible body must install a condensation water / heat transfer fluid drip tray below the unit. Minimum wall clearances depend on the cooling type and whether the unit runs with or without a bath. Refrigerant charge (2.9 kg R-452A) and room size must be checked to avoid dangerous refrigerant concentration. Installation inside an ATEX zone is prohibited.
Selection & Integration Q&A
Pick the 10w when the required cooling load fits its curve (14 kW at 20 °C down to 0.8 kW at -40 °C) and a water-cooled 10-size unit is sufficient. The 10w-XT and 20w/20w-XT/26w offer different sizes or configurations; select them only when their documented capacity or configuration is needed, as ratings are not interchangeable.
Provide cooling water at 5–20 °C with 1–4 bar differential pressure and max 6 bar static pressure, using G3/4 male connections. Consumption is 620 l/h at 15 °C water and 0 °C flow. Install a building-side backflow preventer, use only drinking or filtered river water (no sea water), and insulate lines where condensation can occur.
Ambient temperature must stay 5–40 °C; installation in an ATEX zone is prohibited. A drip tray is required below the unit, and wall clearances depend on cooling type and bath configuration. Check refrigerant charge (2.9 kg R-452A) against room volume. Shut off cooling water during downtimes, and bring fluid to room temperature before switch-off.
The unit ships with a power cable but without a plug for the 400 V three-phase connection; a suitable plug and 3x32 A fusing must be provided on site. Pump hoses with M38x1.5 fittings and cooling water hoses with G3/4 fittings are not included. A drip tray and backflow preventer are site responsibilities. Regular maintenance per the manual includes cooling water circuit checks.