Peter Huber Kältemaschinenbau Unimotive 20w
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unimotive 20w

Peter Huber Kältemaschinenbau Unimotive 20w — Temperature control units.

The Unimotive 20w is a water-cooled temperature control unit with a magnetically coupled stainless steel pump, rated for -45 to 95 °C in externally closed circuits. Its 12 kW heating capacity and 920 l/h cooling-water demand at 15 °C distinguish it from smaller Unimotive 10w siblings. It serves as a central temperature source for closed process loops requiring sub-ambient to near-boiling control.

Peter Huber Kältemaschinenbau Unimotive Series — Model Comparison & Selection

Model Display resolutionHeating capacity, WPower supplyPump connectionRefrigerantSafety class
Unimotive 20w Current 0, 0112000400V 3~ 50HzM38x1, 5 AG maleR-452A (A1, H280)III / FL
Unimotive 10w 0.0112000400V 3~ 50HzM38x1,5 AG maleR-452A (A1, H280)III / FL
Unimotive 10w-XT 0, 0112000400V 3~ 50HzM38x1, 5 AG maleR-452A (A1, H280)I / NFL
Unimotive 20w-XT 0, 0112000400V 3~ 50HzM38x1, 5 AG maleR-452A (A1, H280)I / NFL
Unimotive 26w 0, 01 K24000400V 3~ 50HzM38x1, 5 AG maleR-452A (A1, H280)III / FL
Unimotive 26w-XT 0, 0124000400V 3~ 50HzM38x1, 5 AG maleR-452A (A1, H280)I / NFL
Unimotive 27w 0, 0124000400V 3~ 50HzM38x1, 5 AG maleR-452A (A1, H280)III / FL
Unimotive GL 07w 0, 0112000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)III / FL
Unimotive GL 07w-XT 0, 0112000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)I / NFL
Unimotive GL 10w 0, 0124000460V 3~ 60HzM38x1, 5 AG maleR-744 (A1, H280)III / FL
Unimotive GL 10w-XT 0, 0124000460V 3~ 60HzM38x1, 5 AG maleR-744 (A1, H280)I / NFL
Unimotive GL 20w 0, 0124000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)III / FL
Unimotive GL 20w-XT 0, 0124000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)I / NFL
Unimotive GL 30w 0, 0124000460V 3~ 60HzM38x1, 5 AG maleR-744 (A1, H280)III / FL
Unimotive GL 50w 0, 0124000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)III / FL
Unimotive GL 50w-XT 0, 0124000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)I / NFL
Unimotive GL 60w 0, 0148000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)III / FL
Unimotive GL 60w-XT 0, 0148000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)I / NFL

Full specifications

Core Performance & Functional Specifications
Control output POKO ONE
Heating capacity 12 kW
Pump connection M38x1, 5 AG male
Pump speed adjustable no
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -45...95 °C
Dimensions, Weight & Mechanical Design
Weight 446 kg
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Refrigerant R-452A (A1, H280) (1st stage)
Maximum viscosity 200 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G3/4 male
Cooling water consumption 920 l/h (at 15°C, flow 0°C)
Cooling water differential pressure 1 bar (min)
Cooling Water Pressure 6 bar (max)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

The Unimotive 20w is a water-cooled temperature control unit built for externally closed applications, using a magnetically coupled stainless steel pump with fixed speed and M38x1.5 AG male pump connections. Within the Unimotive series it sits above the 10w and 10w-XT in heating capacity (12 kW) and below the 26w; the -XT siblings are separate variants not to be conflated with this model. The refrigerant circuit uses R-452A in the first stage, and the unit carries safety class III / FL with adjustable overtemperature protection per DIN 12876. The controller is Pilot ONE with POKO ONE control output, a display resolution of 0.01 K, and interfaces including Ethernet, USB (host and device) and RS232.

Working range and media behavior

The temperature range spans -45 to 95 °C. The unit is constructed for water-ethylene glycol mixtures with corrosion protection (e.g. Glysantin®) down to -45 °C, so fluid selection must follow this constraint. Cooling water intake must be between 5 and 20 °C; as cooling water temperature rises, cooling power drops and cooling water flow rate may increase, so site performance must be derated accordingly. The cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, limiting acceptable cooling water quality. Only drinking water or filtered river/sea water may be used; sea water is prohibited.

Installation and operation specifics

The unit requires a 400 V 3~ 50 Hz supply with voltage tolerance ±5% only with simultaneous frequency tolerance ±2%, and leakage current exceeds 3.5 mA, so protective earthing and RCD selection must account for this. Three-phase devices below 63 A are delivered with cable but without plug. Cooling water connections are G3/4 male, with a minimum differential pressure of 1 bar and maximum pressure of 6 bar. A building-side backflow preventer is mandatory, and cooling water supply should be shut off during downtimes because flow can occur even when the unit is switched off. The unit must not be installed in an ATEX zone, and a drip tray must be installed below it by the responsible body. Outdoor installation requires frost-protected cooling water lines and a minimum cooling water temperature of 5 °C.

Integration & Operating Notes

Process Limits
Cooling water intake must stay between 5 and 20 °C

Operating outside this intake temperature window invalidates the published cooling performance and may stress the refrigerant circuit. Site cooling water supply must be verified before commissioning, and performance derating must be applied as intake temperature rises toward 20 °C.

Datasheet, Technical Data
Electrical
Leakage current above 3.5 mA demands RCD coordination

The unit's leakage current exceeds 3.5 mA, so protective earthing must be verified and the RCD type/rating selected accordingly. Three-phase devices below 63 A arrive with cable but no plug, so a suitable plug must be provided on site by the installer.

Datasheet, Technical Data; Operation Manual, Electrical Connection
Mounting
Drip tray and backflow preventer are mandatory site works

The responsible body must install a condensation water / heat transfer fluid drip tray below the unit, and a building-side backflow preventer must protect the drinking water supply. Neither is part of the unit delivery, so both must be planned before commissioning.

Operation Manual, Installation
Cooling & Media
Sea water is prohibited; only drinking or filtered water allowed

The cooling water circuit uses copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only drinking water or filtered river/sea water may be used. Sea water must not be used. When cooling water temperatures fall below the condensation point, the cooling water pipes in the unit must be insulated.

Operation Manual, Cooling Water Connection; Datasheet, Technical Data
Operation
Shut off cooling water during overnight downtimes

Cooling water flow can occur even when the unit is switched off, so the supply should be shut off for shorter downtimes such as overnight. Before switching off, the heat transfer fluid must be brought to room temperature to avoid overpressure in the system.

Operation Manual, Normal Operation / Shutting Down
Process Limits
Pressure-sensitive glass reactors need flow-pressure review

The maximum flow pressure of the unit must be taken into account for pressure-sensitive applications such as glass reactors; an external bypass may be required. The pump is fixed-speed with no adjustment, so the hydraulic circuit must be designed around its fixed flow-pressure characteristic.

Operation Manual, Function Description; Datasheet, Technical Data

Selection & Integration Q&A

When should I choose the Unimotive 20w over the 10w or 26w siblings?

Pick the 20w when the process needs 12 kW heating and the -45 to 95 °C range in an externally closed loop. The 10w offers less heating capacity; the 26w is larger. The -XT variants are separate modifications and should not be specified interchangeably with the 20w.

What cooling water supply and fittings must be provided for the 20w?

Provide G3/4 male cooling water connections with 5–20 °C intake, minimum 1 bar differential and maximum 6 bar pressure; expect about 920 l/h at 15 °C intake with 0 °C flow. A building-side backflow preventer and a drip tray below the unit are mandatory.

What site and duty-cycle limits apply to installing the 20w?

Ambient must be 5–40 °C, IP20, no ATEX zone, and refrigerant charge versus room size must be checked. Outdoor siting requires frost-protected cooling water lines and intake never below 5 °C. Shut off cooling water during overnight downtimes since flow can occur even when switched off.

What is not included in base delivery and must be procured separately?

Three-phase units below 63 A ship with cable but without plug, so a plug must be supplied. Pump connections are M38x1.5 male requiring DIN 3863 ball socket and DIN 3870 sleeve nut fittings; tempering hoses with inside diameter at least matching the pump connections are not included.