Peter Huber Kältemaschinenbau Unimotive GL 07w-XT
Peter Huber Kältemaschinenbau Unimotive GL 07w-XT — Temperature control units.
Water-cooled temperature control unit with magnetically coupled stainless-steel pump and 12 kW heating, covering -45 to 150 °C for externally closed circuits. The XT modification of the GL 07w is rated for water-ethylene glycol mixtures with corrosion protection down to -45 °C, distinguishing it from plain-water siblings. It serves automotive test benches and externally closed process loops requiring three-phase 400 V supply and site cooling water.
Peter Huber Kältemaschinenbau Unimotive Series — Model Comparison & Selection
| Model | Display resolution | Heating capacity, W | Power supply | Pump connection | Refrigerant | Safety class |
|---|---|---|---|---|---|---|
| Unimotive GL 07w-XT Current | 0, 01 | 12000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | I / NFL |
| Unimotive 10w | 0.01 | 12000 | 400V 3~ 50Hz | M38x1,5 AG male | R-452A (A1, H280) | III / FL |
| Unimotive 10w-XT | 0, 01 | 12000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | I / NFL |
| Unimotive 20w | 0, 01 | 12000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | III / FL |
| Unimotive 20w-XT | 0, 01 | 12000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | I / NFL |
| Unimotive 26w | 0, 01 K | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | III / FL |
| Unimotive 26w-XT | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | I / NFL |
| Unimotive 27w | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | III / FL |
| Unimotive GL 07w | 0, 01 | 12000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 10w | 0, 01 | 24000 | 460V 3~ 60Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 10w-XT | 0, 01 | 24000 | 460V 3~ 60Hz | M38x1, 5 AG male | R-744 (A1, H280) | I / NFL |
| Unimotive GL 20w | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 20w-XT | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | I / NFL |
| Unimotive GL 30w | 0, 01 | 24000 | 460V 3~ 60Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 50w | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 50w-XT | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | I / NFL |
| Unimotive GL 60w | 0, 01 | 48000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 60w-XT | 0, 01 | 48000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | I / NFL |
Full specifications
| Control output | POKO ONE (potential-free contact) |
| Heating capacity | 12 kW |
| Pump connection | M38x1, 5 AG male |
| Pump speed adjustable | no |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -45...150 °C |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Refrigerant | R-744 (A1, H280) (1st stage) |
| Maximum viscosity | 200 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G3/4 male |
| Cooling water differential pressure | 1 bar (min) |
| Cooling Water Pressure | 6 bar (max) |
| Protection rating | IP20 |
| Safety class | I / NFL |
Description
Construction and series positioning
- The Unimotive GL 07w-XT is a water-cooled unit with a first-stage R-744 (CO2) refrigerant circuit and a magnetically coupled stainless-steel pump with fixed speed (not adjustable).
- Pump connection is M38x1.5 AG male; cooling water connections are G3/4 male.
- Within the Unimotive series (siblings 10w, 10w-XT, 20w, 20w-XT), this GL 07w-XT is the variant specified for water-ethylene glycol mixtures with corrosion protection down to -45 °C, intended for externally closed applications rather than open baths.
- Heating capacity is 12 kW; temperature range -45...150 °C; controller is Pilot ONE with POKO ONE potential-free contact output.
Behavior and media constraints
- Cooling capacity data are valid at 20 °C room temperature, 15 °C cooling water inlet and 1 bar differential pressure; higher cooling water temperature reduces cooling power and may require increased cooling water flow.
- The circulating fluid must not exceed a kinematic viscosity of 200 mm²/s, which limits fluid selection at low temperatures.
- The unit is designed for water-ethylene glycol mixtures with corrosion protection (e.g. Glysantin®) down to -45 °C; it is not intended for plain water at low temperatures.
- The unit is for externally closed applications; pressure-sensitive loads such as glass reactors must account for the unit's maximum flow pressure, possibly with an external bypass.
Installation, connection and control
- Power supply is 400 V 3~ 50 Hz; three-phase devices below 63 A are delivered with cable but without a plug, so the customer must provide and fit the plug.
- Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%; certain combinations (e.g. -5% voltage with +2% frequency) are not allowed.
- Leakage current exceeds 3.5 mA; connection must be to a socket with protective earth, and hardwiring must be done by a certified electrician. Replacement power cables are limited to 3 m for 1–2 phase and 4 m for 3-phase connections.
- Cooling water must be supplied at 1–4 bar differential pressure and no more than 6 bar static pressure; only drinking water or filtered river water may be used (sea water prohibited), and a building-side backflow preventer is required.
- Cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM; if cooling water temperature falls below the condensation point, internal cooling water pipes must be insulated.
- Interfaces: Ethernet, USB (Host and Device), RS232; external sensor connection Pt100; display resolution 0.01 K.
- Protection rating IP20; ambient temperature limit 5...40 °C; EMC Class A Group 1 (EN55011), so it may cause radio interference in residential environments.
- The unit must not be installed or started within an ATEX zone; refrigerant charge and room size must be checked during installation.
Integration & Operating Notes
The unit requires 400 V 3~ 50 Hz and cannot run on a single-phase supply. Three-phase devices below 63 A are delivered with cable but without a plug, so the customer must provide and fit the appropriate plug; hardwiring must be done by a certified electrician.
Protective earth and residual current protective device selection must account for a leakage current exceeding 3.5 mA. Always connect to a power socket with protective earth (PE).
Cooling water must be compatible with copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Sea water must not be used. A building-side backflow preventer is required to protect the drinking water supply.
When cooling water temperatures fall below the condensation point, the cooling water pipes inside the unit must be insulated to prevent condensation and corrosion. Outdoor installation also requires frost-protected lines and a minimum cooling water temperature of 5 °C.
The responsible body must install a condensation water / heat transfer fluid drip tray below the temperature control unit. Minimum wall clearances depend on air- or water-cooling and on whether the unit runs with or without a bath.
The unit must not be installed or started inside an ATEX zone. Refrigerant charge (R-744, H280) and room size must be checked during installation to avoid dangerous refrigerant concentration.
The unit is intended for externally closed applications, not open baths. The circulating fluid's kinematic viscosity must not exceed 200 mm²/s, which constrains fluid choice at low temperatures. Pressure-sensitive loads such as glass reactors must account for the unit's maximum flow pressure, possibly with an external bypass.
Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%; certain combinations such as -5% voltage with +2% frequency are not allowed. Verify supply quality at the installation site against this envelope.
Selection & Integration Q&A
Pick the GL 07w-XT when the process needs -45...150 °C with 12 kW heating and a water-ethylene glycol mixture with corrosion protection down to -45 °C in an externally closed circuit. Larger siblings 10w-XT/20w-XT apply when higher capacity is required; the non-XT 10w/20w variants are not specified for the glycol mixture.
A 400 V 3~ 50 Hz supply with protective earth (plug not included; cable supplied for <63 A), leakage current >3.5 mA accounted for in RCD design, and cooling water at 1–4 bar differential and ≤6 bar static, compatible with copper/1.4401/EPDM, with a building-side backflow preventer.
Ambient 5...40 °C, IP20, EMC Class A Group 1 (may interfere in residential areas), no ATEX zone, refrigerant charge vs room size checked, drip tray below the unit, wall clearances per the manual table, and frost-protected cooling water lines ≥5 °C for outdoor siting.
The three-phase plug is not included (cable is supplied below 63 A); the customer provides the plug and the building-side backflow preventer and drip tray. Replacement power cables are limited to 3 m (1–2 phase) or 4 m (3-phase). Shut off cooling water during downtimes and warm the fluid to room temperature before switch-off.