Peter Huber Kältemaschinenbau Unimotive GL 10w
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unimotive GL 10w

Peter Huber Kältemaschinenbau Unimotive GL 10w — Temperature control units.

The Unimotive GL 10w is a water-cooled, R-744 (CO2) refrigeration temperature control unit built for water-ethylene glycol mixtures from -45 to 95 °C, with 24 kW heating and a magnetically coupled stainless-steel pump. It is the GL variant of the 10w series, adding a water-cooled condenser for facilities with a cooling-water loop instead of the air-cooled Unimotive 10w. It serves automotive test benches and process circuits requiring deep sub-ambient temperature control.

Peter Huber Kältemaschinenbau Unimotive Series — Model Comparison & Selection

Model Display resolutionHeating capacity, WPower supplyPump connectionRefrigerantSafety class
Unimotive GL 10w Current 0, 0124000460V 3~ 60HzM38x1, 5 AG maleR-744 (A1, H280)III / FL
Unimotive 10w 0.0112000400V 3~ 50HzM38x1,5 AG maleR-452A (A1, H280)III / FL
Unimotive 10w-XT 0, 0112000400V 3~ 50HzM38x1, 5 AG maleR-452A (A1, H280)I / NFL
Unimotive 20w 0, 0112000400V 3~ 50HzM38x1, 5 AG maleR-452A (A1, H280)III / FL
Unimotive 20w-XT 0, 0112000400V 3~ 50HzM38x1, 5 AG maleR-452A (A1, H280)I / NFL
Unimotive 26w 0, 01 K24000400V 3~ 50HzM38x1, 5 AG maleR-452A (A1, H280)III / FL
Unimotive 26w-XT 0, 0124000400V 3~ 50HzM38x1, 5 AG maleR-452A (A1, H280)I / NFL
Unimotive 27w 0, 0124000400V 3~ 50HzM38x1, 5 AG maleR-452A (A1, H280)III / FL
Unimotive GL 07w 0, 0112000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)III / FL
Unimotive GL 07w-XT 0, 0112000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)I / NFL
Unimotive GL 10w-XT 0, 0124000460V 3~ 60HzM38x1, 5 AG maleR-744 (A1, H280)I / NFL
Unimotive GL 20w 0, 0124000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)III / FL
Unimotive GL 20w-XT 0, 0124000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)I / NFL
Unimotive GL 30w 0, 0124000460V 3~ 60HzM38x1, 5 AG maleR-744 (A1, H280)III / FL
Unimotive GL 50w 0, 0124000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)III / FL
Unimotive GL 50w-XT 0, 0124000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)I / NFL
Unimotive GL 60w 0, 0148000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)III / FL
Unimotive GL 60w-XT 0, 0148000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)I / NFL

Full specifications

Core Performance & Functional Specifications
Control output POKO ONE
Heating capacity 24 kW
Pump connection M38x1, 5 AG male
Pump speed adjustable no
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -45...95 °C
Dimensions, Weight & Mechanical Design
Weight 600 kg
Electrical, Power & Drive Specifications
Current consumption 47 A (max)
Power supply 460V 3~ 60Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Refrigerant R-744 (A1, H280) (1st stage)
Maximum viscosity 200 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G3/4 male
Cooling water consumption 2100 l/h (at 15°C, flow 0°C)
Cooling water differential pressure 1 bar (min)
Cooling Water Pressure 6 bar (max)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Working range and media behavior

Control, installation and operation

Integration & Operating Notes

Electrical
Provide UL 489 branch protection and a site-made plug connection

The unit draws 47 A at 460 V 3~ 60 Hz and requires 3x63 A fuses plus external branch circuit protection according to UL 489. Three-phase devices below 63 A are delivered with cable but without plug, so the site must provide the mating connection and protective earth.

Datasheet, Specifications (installation
Cooling & Media
Cooling water must stay within 5–20 °C and 1–4 bar differential

The cooling-water circuit requires a differential pressure between 1 and 4 bar and must never exceed 6 bar absolute. Intake temperature is limited to 5–20 °C; below the condensation point the internal cooling-water pipes must be insulated. Only drinking water or filtered river water is permitted; sea water is prohibited.

Datasheet, Specifications (installation); Operation Manual, Cooling Water Connection
Cooling & Media
Cooling capacity derates with warmer site water

Nominal cooling-water consumption is 2100 l/h at 15 °C water and 0 °C flow. As cooling-water temperature increases, cooling power drops and the required flow rate may rise, so capacity must be derated for the actual site water temperature rather than assumed from the nominal point.

Datasheet, Specifications (operation
Process Limits
Fluid must be glycol-based and within 200 mm²/s viscosity

The unit is constructed for water-ethylene glycol mixtures with corrosion protection (e.g. Glysantin®) down to -45 °C. The maximum permissible kinematic viscosity is 200 mm²/s, which limits fluid choice at the lowest operating temperatures; other fluid families are outside the intended envelope.

Datasheet, Overview and Specifications (selection

Selection & Integration Q&A

When should I choose the GL 10w over the air-cooled Unimotive 10w or the larger 20w variants?

Pick the GL 10w when a facility cooling-water loop (5–20 °C, 1–4 bar differential) is available and heat must be rejected to water rather than room air. Choose the air-cooled 10w where no water loop exists; move to the 20w only if the required cooling or heating capacity exceeds this unit's 24 kW heating and its pump curve.

What must the site provide electrically and hydraulically to commission this unit?

A 460 V 3~ 60 Hz supply within the combined ±5%/±2% envelope, 3x63 A fuses and UL 489 branch protection, plus a mating plug since the cable is supplied without one. Hydraulically: G3/4 male cooling-water connection with 1–4 bar differential, a building-side backflow preventer, and hoses sized at least to the M38x1.5 pump connections.

What environment and duty-cycle limits apply to installation and operation?

Ambient 5–40 °C, IP20, not inside an ATEX zone, EMC class A group 1 (industrial only). Cooling water must not fall below 5 °C and lines need frost protection outdoors. Bring fluid to room temperature before switch-off, and shut off cooling water during downtimes since flow can occur with the unit off.

What maintenance and consumables are needed, and what is not included in base delivery?

Base delivery excludes the power plug, the drip tray (responsible body must install), the building-side backflow preventer and the heat transfer fluid itself. Cooling-water circuit materials (copper, 1.4401, MS, PA, PPE, PTFE, EPDM) dictate water quality; only drinking or filtered river water is allowed. Replacement power cables are limited to 4 m for three-phase connections.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unimotive GL 10w
Datasheet ·97,0 КБ ·2 pp.
View PDF
PDF
Unimotive Operation Manual
Manual ·2,8 МБ ·82 pp.
View PDF